Maintenance
• Clorox
®
Wipes (they do not contain bleach)
• Over-the-counter detergents (such as Fantastic
®
, Tilex
®
, and so on)
3.
Wipe the cables and leadwires with a dry, clean cloth or paper towel and let
them air dry.
Disinfecting the Cables and Leadwires
Use the following procedure to disinfect the cables and leadwires:
1.
Clean and dry the cables and leadwires before disinfecting them.
See
“Cleaning the Cables and Leadwires” on page 18
for instructions.
2.
Wipe the cables and leadwires with a soft, lint-free cloth that is moistened with
an appropriate disinfectant.
Use the following solutions, as recommended in
APIC Guidelines for Selection
and Use of Disinfectants (1996)
:
• Sodium hypochlorite (5.2% household bleach) minimum 1:500 dilution
(minimum 100 ppm free chlorine) and maximum 1:10 dilution.
• Any sodium hypochlorite wipe product that meets the previous guidelines.
3.
Dry the cables and leadwires with a clean cloth or paper towel and let them
air dry for at least 30 minutes.
NOTE:
• Drying times vary based on the environmental conditions.
• DO NOT use excessive drying techniques, such as ovens, forced heat, or
sun drying.
Storing Cables and Leadwires
To ensure that the cables and leadwires are in proper working order, use the following
guidelines to store them when not in use:
• Store cables and leadwires in a dry, well-ventilated area.
• Hang cables and leadwires vertically
• Do NOT coil cables or leadwires around the device.
Cleaning, Disinfecting, and Sterilizing Reusable
Electrodes
Clean reusable electrodes immediately after using them on a patient.
2053535-004D
MAC™2000 ECG Analysis System
19