Technical Publication

LOGIQ V3/V5/V5 Expert

Basic Service Manual

Direction Number:

 

5726264-100 English

Rev.8

All Material Copyright © 2016-2017 by General Electric Company Inc. 
All Rights Reserved.

Содержание LOGIQ V3

Страница 1: ...GE Technical Publication LOGIQ V3 V5 V5 Expert Basic Service Manual Direction Number 5726264 100 English Rev 8 All Material Copyright 2016 2017 by General Electric Company Inc All Rights Reserved...

Страница 2: ...ound system Copyright 2016 2017 General Electric Company GE Medical Systems a General Electric Company going to market as GE Manufacturer GE Medical Systems China Co Ltd No 19 Changjiang Road WuXi Nat...

Страница 3: ...27 or 1 262 524 5698 Revision Date YYYY MM DD Reason for change Rev 1 2016 01 21 Initial Release Rev 2 2016 06 12 Update control panel figure and book format Rev 3 2016 08 10 Update warning label Rev...

Страница 4: ...i 2 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 Important precautions Translation policy...

Страница 5: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 3 5726264 100 English Rev 8...

Страница 6: ...i 4 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 7: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 5 5726264 100 English Rev 8...

Страница 8: ...i 6 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 9: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 7 5726264 100 English Rev 8...

Страница 10: ...i 8 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 11: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 9 5726264 100 English Rev 8...

Страница 12: ...i 10 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 13: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 11 5726264 100 English Rev 8...

Страница 14: ...within this 14 day period Certified electrical contractor statement For USA Only All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipmen...

Страница 15: ...listing the system type manual title part number revision number page number and suggestion details Mail the information to GE Medical Systems China Co Ltd No 19 Changiang Road Wuxi National Hi Tech...

Страница 16: ...fety requirements refer to Chapter 1 in the Service Manual DANGER DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARN...

Страница 17: ...sic service software does not extend to or cover this software or related documentation If you are a GE employee or a customer who has entered into such a license agreement with GE to use this proprie...

Страница 18: ...i 16 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 19: ...cal users of the Proprietary Service Manual 1 4 Purpose of Operator Manual s 1 4 LOGIQ V3 V5 V5 Expert models covered by this manual 1 5 Important conventions Conventions used in book 1 6 Standard haz...

Страница 20: ...15 Minimal floor plan suggestion 2 16 Recommended floor plan suggestion 2 17 Networking setup requirements 2 18 Chapter 3 System Setup Overview Contents in this chapter 3 2 Setup reminders Average se...

Страница 21: ...of this chapter 4 2 Contents in this chapter 4 2 Special Equipment required 4 2 General procedures Overview 4 4 Power ON Boot Up 4 5 Power off 4 8 Check System Date and Time 4 10 Manufacturer 4 11 Rem...

Страница 22: ...Mechanical Functions Checks 4 43 Application Turnover Check List Software Configuration Checks 4 45 Site Log My Trainer Overview 4 47 Chapter 5 Components and Functions Theory Overview Contents in thi...

Страница 23: ...tions 8 6 Recording important settings and parameters 8 7 Loading the System Software 8 8 Rewrite the Serial Number 8 22 Chapter 9 Renewal Parts Overview Contents in this chapter 9 2 List of Abbreviat...

Страница 24: ...imits 10 21 Outlet test wiring arrangement 10 23 Grounding continuity 10 24 Chassis leakage current test 10 25 Probe leakage current test 10 27 When there s too much leakage current AC DC Fails 10 30...

Страница 25: ...100 English Rev 8 Chapter 1 Introduction This chapter describes important issues related to safely servicing the Ultrasound system The service provider must read and understand all the information pr...

Страница 26: ...on page 1 3 1 3 Important conventions on page 1 6 1 4 Product icons on page 1 10 1 5 Labels locations on page 1 11 1 6 Safety considerations on page 1 16 1 7 Dangerous procedure warnings on page 1 24...

Страница 27: ...s and Legal notes and the Table of Contents TOC An Index has been included after Chapter 10 Table 1 1 Contents in this manual Chapter number Chapter title Description 1 Introduction Contains a content...

Страница 28: ...not necessarily reflect the probes available on your ultrasound system Please refer to the probe list for available probes and features 7 Diagnostics Troubleshooting Provides procedures for running d...

Страница 29: ...Expert models covered by this manual NOTE When not otherwise specified the contents of this manual applies to all LOGIQ V3 V5 V5 Expert models Table 1 2 LOGIQ V3 V5 V5 Expert Model Designations Model...

Страница 30: ...described next 1 3 1 1 Model designations This manual covers the LOGIQ V3 V5 V5 Expert Ultrasound systems listed in 1 2 4 LOGIQ V3 V5 V5 Expert models covered by this manual on page 1 5 1 3 1 2 Icons...

Страница 31: ...tes are used to provide important information about an item or a procedure NOTE Be sure to read the notes the information contained in a note can often save you time or effort DANGER Danger is used to...

Страница 32: ...le or the symbols for Danger Warning or Caution as seen throughout this chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards...

Страница 33: ...Some others icons make you aware of specific procedures that should be followed Be sure to read the notes the information contained in a note can often save you time or effort Table 1 4 Standard Icon...

Страница 34: ...5 Expert Basic Service Manual 5726264 100 English Rev 8 1 4 Product icons It is important to refer to the current revision of the Ultrasound system s Basic User Manual for a full list of product label...

Страница 35: ...ion of the Ultrasound system s Basic User Manual for a full list of product labels prior to servicing the system 1 5 1 Rating Plate Location LOGIQ V3 V5 V5 Expert labels are provided in English The la...

Страница 36: ...Introduction 1 12 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 1 5 1 Rating Plate Location continued Figure 1 1 LOGIQ V5 Expert Label Location...

Страница 37: ...Labels locations LOGIQ V3 V5 V5 Expert Basic Service Manual 1 13 5726264 100 English Rev 8 1 5 1 Rating Plate Location continued Figure 1 2 LOGIQ V5 Label Location...

Страница 38: ...Introduction 1 14 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 1 5 1 Rating Plate Location continued Figure 1 3 LOGIQ V3 Label Location...

Страница 39: ...up_products asp prod_ id 44177 Table 1 5 Label Icons Label Icon Purpose Meaning Location Every system has a unique marking for identification the Unique Device Identification UDI Label The UDI label c...

Страница 40: ...recautions must be observed during all phases of operation service and repair of this equipment Failure to comply with these precautions or with specific warnings elsewhere in this manual violates saf...

Страница 41: ...people available to deliver and unpack the LOGIQ V3 V5 V5 Expert Attempts to move the Ultrasound system considerable distances or on an incline by one person could result in injury or damage or both W...

Страница 42: ...mber light on the OP panel on off button will turn off Ultrasound System components may be energized Always refer to the Ultrasound system s Proprietary Service Manual for LOTO warnings and cautions W...

Страница 43: ...NING Never use a probe that has fallen to the floor Even if it looks OK it may be damaged WARNING When the Ultrasound system is raised for a repair or moved along any incline use extreme caution since...

Страница 44: ...hat nobody touches the console arm when moving the operator panel CAUTION Do not move the Ultrasound system if the Operator Panel is in unlocked position CAUTION Do not transport LOGIQ V3 V5 V5 Expert...

Страница 45: ...age case Lock the wheels brake Ensure that the system is firmly secured while inside the vehicle Secure system with straps or as directed otherwise to prevent motion during transport Prevent vibration...

Страница 46: ...he Ultrasound system is equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with safety ground The power outlet used for this equipment should not be...

Страница 47: ...egradation to the housing cable strain relief lens seal connector pins locking mechanism Do not use a damaged or defective probe Never immerse the probe connector or adapter into any liquid The system...

Страница 48: ...es may be warm enough to pose a potential heat hazard if touched even while in shutdown mode WARNING Explosion Warning DO NOT operate the equipment in an explosive atmosphere Operation of any electric...

Страница 49: ...esidual energy 6 Verify isolation All potentially hazardous stored or residual energy is relieved WARNING Energy Control and Power Lockout for LOGIQ V3 V5 V5 Expert When servicing parts of the Ultraso...

Страница 50: ...are regulated medical waste for transportation purposes and must be transported as a hazardous material NOTE The USER SERVICE staff should dispose of all the waste properly per federal state and loca...

Страница 51: ...nterference RFI EMI and can be radiated through space or conducted over interconnecting power of signal cables In addition to electromagnetic energy EMC also includes possible effects from electrical...

Страница 52: ...nnection point located on the rear of the Ultrasound system near the power connector Follow general guidelines for handling of electrostatic sensitive equipment WARNING Risk of electrical shock Ultras...

Страница 53: ...t the local distributor or appropriate support resource as listed below Before you call identify the following information and acquire log Alt D to send to the Customer Care team 1 System ID serial nu...

Страница 54: ...1 800 668 0732 Latin America Service Application Support 1 800 321 7937 1 262 524 5698 Europe OLC EMEA GE Ultraschall Deutschland Gmbh Co KG Beethovenstra e 239 Postfach 11 05 60 D 42655 Solingen Germ...

Страница 55: ...in the Safety chapter of the Basic User Manual Authorized EU Representative European registered place of business GE Medical Systems Information Technologies GmbH GEMS IT GmbH Munzinger Strasse 5 D 7...

Страница 56: ...Introduction 1 32 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 57: ...100 English Rev 8 Chapter 2 Site Preparations This chapter provides the information required to plan and prepare for the setup of an Ultrasound system Included are descriptions of the facility and el...

Страница 58: ...OGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 2 1 Overview 2 1 1 Contents in this chapter 2 1 Overview on page 2 2 2 2 General Ultrasound system requirements on page 2 3 2 3 Faci...

Страница 59: ...d system environmental requirements 2 2 2 1 If the Ultrasound system is very cold or hot When unpacking the Ultrasound system allow the temperature of the Ultrasound system to stabilize before powerin...

Страница 60: ...e in degrees C Conversion to Degrees F Degrees C 9 5 32 Table 2 2 System Environmental Requirements Operational Storage Transport 16hrs Temperature 10 40 C 50 104 F 5 50 C 23 122 F 5 50 C 23 122 F Hum...

Страница 61: ...he room NOTE Each person in the room places an additional 300 BTU h demand on the cooling system 2 2 2 4 Lighting Bright light is needed for Ultrasound system installation updates and repairs However...

Страница 62: ...rom the distribution panel to the Ultrasound outlet NOTE Please note that image artifacts can occur if at any time within the facility the ground from the main facility s incoming power source to the...

Страница 63: ...t needed to support this Ultrasound system must also be present within 1 m 3 2 ft of the Ultrasound system Electrical installation must meet all current local state and national electrical codes 2 2 3...

Страница 64: ...IEC 61000 4 11 Voltage dips short interruptions and voltage variations on mains supply 5 T 95 dip for 0 5 cycle 5 T 95 dip for 0 5 cycle Mains power quality should be that of a typical commercial or...

Страница 65: ...ion Possible EMI sources should be identified before the Ultrasound system is installed Electrical and electronic equipment may produce EMI unintentionally as the result of a defect Some of these sour...

Страница 66: ...s to interfere with the ultrasound signals Replace broken Radio Frequency gaskets If more than 20 or a pair of the fingers on an Radio Frequency gasket are broken replace the gasket Do not turn on the...

Страница 67: ...Begin site preparation checks as soon as possible if possible six weeks before delivery to allow enough time to make any changes CAUTION Ensure that the probe face temperature does not exceed the norm...

Страница 68: ...2 3 1 Contents in this section 2 3 2 Purchaser responsibilities on page 2 13 2 3 3 Required facility needs on page 2 14 2 3 4 Desirable features on page 2 15 2 3 5 Minimal floor plan suggestion on pa...

Страница 69: ...e products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on these products must comply with the req...

Страница 70: ...efore the Ultrasound system Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one phase a neutral not shared with any other circuit and a full size grou...

Страница 71: ...are within 2 m 6 5 ft of each other with peripheral within 1 m of the unit to connect cables Power outlets for other medical equipment Power outlets for test equipment within 1 m 3 2 ft of Ultrasound...

Страница 72: ...0 inches 2 Film Viewer 3 Counter Top Sink with hot and cold water and Supplies Storage 4 Linen Supply 5 Probes Supplies 6 Examination Table 1930 x 610 mm 76 x 24 inches 7 Footswitch 8 Stool 9 Ultrasou...

Страница 73: ...and Sink with hot and cold water 6 Overhead Lights Dimmer Dual Level Lighting bright and dim 7 Emergency Oxygen 8 Suction Line 9 Ultrasound system 10 Dedicated Power Outlet Circuit Breaker protected...

Страница 74: ...ia 100 1000 Mbit Interface 2 3 7 4 Purpose of the DICOM network function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital n...

Страница 75: ...at there are no spaces in any field of the form See Entries must include A host name local port number AE Title IP address and Net Mask for the Ultrasound system The IP addresses for the default gatew...

Страница 76: ...Site Preparations 2 20 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 2 3 7 5 DICOM option setup requirements continued Figure 2 3 Worksheet for DICOM Network Information...

Страница 77: ...stall LOGIQ V3 V5 V5 Expert system Included is a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim How to prepare the facility and unit of the act...

Страница 78: ...pter 3 1 Overview on page 3 2 3 2 Setup reminders on page 3 3 3 3 Receiving and unpacking the equipment on page 3 6 3 4 Preparing for setup on page 3 13 3 5 Completing the setup on page 3 14 3 6 Syste...

Страница 79: ...xpert consoles with optional equipment may take slightly longer 3 2 2 Setup warnings DANGER WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE I E METER S GROUND SWITCH IS OP...

Страница 80: ...the unit to acclimate before you turn it on It requires one hour for each 2 5 C increment it s temperature is below 3 C or above 40 C The following table describes guidelines for reaching operational...

Страница 81: ...s should be fully read and understood before operating the LOGIQ V3 V5 V5 Expert and kept near the Ultrasound system for quick reference CAUTION Acoustic Output Hazard Although the ultrasound energy t...

Страница 82: ...king the LOGIQ V3 V5 V5 Expert on page 3 8 3 3 3 Warnings for receiving and unpacking CAUTION Two people are needed to unpack the Ultrasound system because of its weight Attempts to move the Ultrasoun...

Страница 83: ...100 English Rev 8 3 3 4 LOGIQ V3 V5 V5 Expert transportation box label The LOGIQ V3 V5 V5 Expert transportation box label is located at the front of the transportation box Figure 3 1 LOGIQ V3 V5 V5 E...

Страница 84: ...Expert console is well assembly after unpacking the system CAUTION Please carefully unpack the system and do not dispose the package of LOGIQ V3 V5 V5 Expert so that it can be reused for service CAUTI...

Страница 85: ...ate Note To avoid injury hold the strap clasp with one hand when cutting the strap 3 Remove the top cover 4 Remove the three plastic locks Note Rotate the inside plastic lock counterclockwise to remov...

Страница 86: ...it Note There is no dust bag if the system is transported by sea Ignore this step if there is no dust bag 7 Remove the clear plastic wrapped around the LOGIQ V3 V5 V5 Expert from the unit 8 Cut the pa...

Страница 87: ...and remove the aluminum foil bag from the unit Note To avoid damaging the unit please use a pair of scissors instead of the knife 10 Remove the foams beside the LCD monitor 11 Unlock the wheels and th...

Страница 88: ...System Setup 3 12 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 3 3 6 Packing the Equipment Please pack LOGIQ V3 V5 V5 Expert in the reverse order of unpacking...

Страница 89: ...e system has been damaged please refer to Damage in transportation on page i 12 in the beginning of this manual 3 4 3 EMI protection The LOGIQ V3 V5 V5 Expert has been designed to minimize the effects...

Страница 90: ...System specifications on page 3 14 3 5 4 Electrical specifications on page 3 16 3 5 7 Connections on the I O Rear Panel on page 3 20 3 5 8 Connecting probes on page 3 20 3 5 9 Powering the system on...

Страница 91: ...system with 17 LCD Monitor approx 45 kg 99 lbs Weight system with 19 LCD Monitor approx 48 kg 106 lbs Table 3 3 Physical dimensions of LOGIQ V3 V5 V5 Expert Height Width Depth Unit 1344 system with 1...

Страница 92: ...oduct labels prior to serving the system 3 5 4 2 Electrical specifications for LOGIQ V3 V5 V5 Expert In the table below the electrical specifications for LOGIQ V3 V5 V5 Expert includes monitor and on...

Страница 93: ...Service Manual 3 17 5726264 100 English Rev 8 3 5 5 Connecting Cables WARNING For software Option Installation the operator must login as Administrator Table 3 5 List of Connecting Cables Name Part No...

Страница 94: ...e connector panel Figure 3 3 Rear Connector Panel NOTE The AC printer can be connected to the isolated Printer USB port on the rear panel only 1 Printer Dedicated external USB printer connection 2 Dua...

Страница 95: ...Y 6 GND 14 VSY 7 GND 15 NC 8 GND Table 3 7 Pin Assignments of USB Pin No Signal Pin No Signal 1 5 VDC 5 5 VDC 2 DATA 6 DATA 3 DATA 7 DATA 4 GND 8 GND Table 3 8 Pin Assignments of Audio Pin No Signal P...

Страница 96: ...ersion of IEC60601 1 If in doubt consult the technical service department or your local representative for GE 3 5 7 1 Connect Ethernet Connect the network cable to the Ethernet connector on the I O Re...

Страница 97: ...of the LOGIQ V3 V5 V5 Expert Basic User Manual Chapter 16 Information includes Entering Location Adjusting Date and Time Selecting User interface Language Selecting Online Manual Language Selecting U...

Страница 98: ...Key installation continued TVI LOGIQ View or 3D 4D If any of them is installed please replace the blank key cap according to the assigned software function Figure 3 5 Replace Key Cap There is a key r...

Страница 99: ...2 1 User Configurable Key installation continued After the key cap is replaced Enter Utility System User Configurable Key and assign the user defined key Figure 3 7 Assign User Configurable Key 3 6 2...

Страница 100: ...language Keyboard Film installation continued 2 The paster is divided into four parts peel off two parts of the paster in the middle of the keyboard film Figure 3 9 Peel Off the Paster 3 Make sure the...

Страница 101: ...5 Expert Basic Service Manual 3 25 5726264 100 English Rev 8 3 6 2 2 Multi language Keyboard Film installation continued 4 Peel off the other part of the paster and align the whole film on the keyboar...

Страница 102: ...LOGIQ V3 V5 V5 Expert Peripherals Description Weight kg Power Control Model B W USB Printer USB port Sony UP D898MD Printer B W USB Printer USB port Sony UP D711MD Printer Color USB Printer USB port S...

Страница 103: ...le 3 12 Materials furnished with B W Printer Item Description Quantity Note 1 Sony UP D711MD Printer 1 2 Sony UP D898MD Printer 1 3 Paper Roll 1 4 USB cable 1 Table 3 13 Materials furnished with Color...

Страница 104: ...Materials furnished with the Footswitch Item Description Quantity Note 1 1 Pedal Footswitch 1 2 3 Pedal Footswitch 1 Table 3 17 Materials furnished with the Footswitch Item Description Quantity Note...

Страница 105: ...5 min 3 7 3 1 3 Preparations 1 Unpack B W Printer 2 Ensure no physical damage 3 7 3 2 Installation Procedure NOTE The printer driver is customized for the LOGIQ V3 V5 V5 Expert at the factory you do n...

Страница 106: ...inued NOTE The DC printer only can be connected to the system via the DC power port and USB port on the neck of the system as shown below Figure 3 13 Power and USB Ports for DC printer a DC Power Port...

Страница 107: ...in the Name Filed This name is used on the Button screen After you select the printer from the Printer pull down Properties menu again it turns white Press Save NOTE After selecting the printer the fi...

Страница 108: ...ize of the printer take Sony UP D898 as an example Press Utility System Peripherals Select the UP D898 from the pull down menu under Then Click Properties Figure 3 17 Properties Click Print Printing P...

Страница 109: ...Sony UP D25MD Printer 2 Ensure no physical damage 3 7 3 3 4 Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC o...

Страница 110: ...ter 2 Ensure no physical damage 3 7 3 4 4 Installation Procedure 1 Place the device on printer shelf 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC output in the wall outl...

Страница 111: ...needed 3 7 3 5 2 Manpower One person 5 min 3 7 3 5 3 Preparations 1 Unpack the Printer Shelf 2 Ensure no physical damage 3 7 3 5 4 Installation Procedure 1 Screw 2 screws on the rear side of the keybo...

Страница 112: ...6 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 3 7 3 5 4 Installation Procedure continued 3 Fix 2 screws on both left and right sides to fix the printer shelf Figure 3 24 Fix t...

Страница 113: ...Installation 3 7 3 6 1 Tools No special tools needed 3 7 3 6 2 Manpower One person 2 min 3 7 3 6 3 Preparations 1 Unpack the Footswitch 2 Ensure no physical damage 3 7 3 6 4 Installation Procedure 1 C...

Страница 114: ...5 Configuring Footswitch Footswitch supports these configurations No Function Freeze Next Heartcycle Prev Heartcycle Print1 Print2 Update Next Step Scan Assistant Previous Step Scan Assistant Scan Ass...

Страница 115: ...HDD Installation 3 7 3 7 1 Tools No special tools needed 3 7 3 7 2 Manpower One person 1 min 3 7 3 7 3 Preparations 1 Unpack the USB HDD 2 Ensure no physical damage 3 7 3 7 4 Installation Procedure 1...

Страница 116: ...Tools No special tools needed 3 7 3 8 2 Manpower One person 1 min 3 7 3 8 3 Preparations 1 Unpack the ECG 2 Ensure no physical damage 3 7 3 8 4 Installation Procedure 1 Connect the ECG to the USB por...

Страница 117: ...tools needed 3 7 3 9 2 Manpower One person 1 min 3 7 3 9 3 Preparations 1 Unpack the DVDRW 2 Ensure no physical damage 3 7 3 9 4 Installation Procedure 1 Take out the DVDRW and USB Y cable and connec...

Страница 118: ...LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 3 7 3 9 4 Installation Procedure continued 3 Connect the two connectors on the Y cable to the USB ports on the system Figure 3 31...

Страница 119: ...ptor 3 7 3 10 1 Tools No special tools needed 3 7 3 10 2 Manpower One person 1 min 3 7 3 10 3 Preparations 1 Unpack the wireless adaptor 2 Ensure no physical damage 3 7 3 10 4 Installation Procedure 1...

Страница 120: ...ts Please contact with your local GE Ultrasound representative to get the latest information 1 Power on the system NOTE Keep the power cord connection during the installation 2 After the power up sequ...

Страница 121: ...re is no need to restart the system after each installation if several option keys are installed at one time Select Cancel for the first several times and select OK after the last installation to acti...

Страница 122: ...user defined key caps with the corresponding option caps on the control panel Figure 3 36 Replace Caps 7 This completes the option installation of LOGIQ V3 V5 V5 Expert NOTE Installing corresponding h...

Страница 123: ...operation and service of the complete Ultrasound system 3 9 1 Contents in this Section 3 9 2 User s Manual s on page 3 47 3 9 3 Product Locator Installation Card on page 3 47 3 9 2 User s Manual s Use...

Страница 124: ...System Setup 3 48 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 125: ...1 5726264 100 English Rev 8 Chapter 4 General Procedures and Functional Checks This chapter provides procedures for quickly checking major functions of the LOGIQ V3 V5 V5 Expert and diagnostics instr...

Страница 126: ...checking major functions of the scanner and diagnostics instructions using the built in service software 4 1 2 Contents in this chapter 4 1 Overview on page 4 2 4 2 General procedures on page 4 3 4 3...

Страница 127: ...und system performance and cooling purposes WARNING Energy Control and Power Lockout for LOGIQ V3 V5 V5 Expert When servicing parts of the Ultrasound system where there is exposure to voltage greater...

Страница 128: ...procedures are used more often than other The intention with this section is to keep the most used procedures in one place 4 2 1 1 Contents in this section 4 2 2 Power ON Boot Up on page 4 5 4 2 3 Pow...

Страница 129: ...POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER USE A THREE TO TWO PRONG ADAPTER THIS DEFEATS THE SAFETY GROUND DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT T...

Страница 130: ...g sufficient slack so that the unit can be moved slightly 3 Verify that the power cable is without any visible scratches or any sign of damage 4 Verify that the on site mains voltage is within the lim...

Страница 131: ...Back End Processor c Back End Processor and rest of scanner starts with the sequence listed in the next steps d Back End Processor is turned ON and starts to load the software e The Start Screen is d...

Страница 132: ...oices Figure 4 1 System Exit Window Logoff Use this button to log off the current user The system remains ON and ready for a new user to log on If the Logoff button is dimmed it indicates that no user...

Страница 133: ...d be connected to the system Cancel Use this button to exit from the System Exit menu and return to the previous operation 4 2 3 1 System shutdown Disconnect the Mains Power Cable is necessary For exa...

Страница 134: ...system time of last boot up The BIOS time is changed by 24 hours earlier than the current system by resetting BIOS time replacing BIOS module or changing BIOS time This warning message is to remind th...

Страница 135: ...gure 4 3 Manufacturer Or via System About Additional About Information to check the CPU manufacturer information Figure 4 4 Additional About Information 4 2 6 Removable media Refer to the latest revis...

Страница 136: ...This ensures the presets loaded after the software reload are as up to date as possible All user presets except changes to Summary Anatomy and Biometry pages can be saved on an DVD R disk or USB memo...

Страница 137: ...menu 3 Enter the label for the media It is best to use all capital letters with no spaces or punctuation marks Select Format Figure 4 5 Format and Verify Media 4 The system displays a pop up menu as...

Страница 138: ...prior service 1 Insert a formatted media into the drive 2 Enter enter Utility System Backup Restore NOTE If you are not logged in as GE Service or with administrator privileges the Operator Login win...

Страница 139: ...Login window is displayed Log on with administrator privileges 3 In the Restore list select Patient Archive Report Archive User Defined Configuration and Service 4 In the Media field select the Backup...

Страница 140: ...moved from the hard drive by date Therefore the best way is to label media by date NOTE Images will be moved from the hard drive by date Therefore the best way to label media is by date When images a...

Страница 141: ...ckup Move 7 Press EZBackup EZMove Figure 4 11 EZBackup Wizard 1 8 Verify the information on the first page of EZBack EZMove wizard then select Next If you want to backup all of the exams in the range...

Страница 142: ...d 1 9 It indicates the size of the data and the storage Select Next to continue NOTE The calculation for the number of backup CD is only an estimate Allow for one additional CD when performing an EZBa...

Страница 143: ...tinued Figure 4 14 Insert Media Message 10 The status menu appears When the backup move has been complete press Next Figure 4 15 EZBackup Wizard 3 NOTE If you need to insert the next media a message a...

Страница 144: ...nal Checks 4 20 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 4 2 7 4 Archiving Images continued Figure 4 16 EZBackup Completion Window All databases presets and images should n...

Страница 145: ...onal checks for LOGIQ V3 V5 V5 Expert are described Functional checks are used to verify that the product works as intended Functional checks may also be used during troubleshooting 4 3 2 Contents in...

Страница 146: ...8 Active Images key 9 Worksheet key 10 Report key 11 End Exam key 12 Alphanumeric Keyboard 13 Scan Coach 14 PW Mode PW Gain CW Gain 15 B Mode B Mode Gain 16 Depth B Steer 17 Trackball 18 Comment key...

Страница 147: ...V5 V5 Expert Basic Service Manual 4 23 5726264 100 English Rev 8 4 3 6 LOGIQ V5 LOGIQ V3 SoftMenu Key Tour Rotary key Rotate to adjust the menu Page Up Down key To turn the menu page up and down Figu...

Страница 148: ...reen 8 Delete Image 9 Save As Menu 10 Number of Images in Exam 11 Page Indication 12 Trackball Functionality Status 13 Current date and time Caps Lock lit when on network connection indicator PC conne...

Страница 149: ...LOGIQ V3 V5 V5 Expert Basic Service Manual 4 25 5726264 100 English Rev 8 4 3 8 Performance Tests 4 3 8 1 Test Phantoms The use of test phantoms is only recommended if required by your facility s cust...

Страница 150: ...n the Operator Panel to access 2D mode These Image Controls are used to optimize the 2D picture Verify that all the listed controls are working as intended Use Gain and TGC controls to optimize the ov...

Страница 151: ...nsures that the data collection consistent independently of who performs the test This preset only includes fundamental settings for 2D mode Processing modes like SRI Harmonics etc are turned off To d...

Страница 152: ...V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 4 3 13 Image Quality Check IQC preset for service continued Figure 4 21 Image Quality Check 3 Press Probe Preset on the control panel Sel...

Страница 153: ...icture on the screen Use the trackball to position the cursor over the required area of the image 4 3 17 Adjust the M Mode controls These Image Controls are used to optimize the M mode picture Verify...

Страница 154: ...om 2D mode or from M mode or a combination of these 4 3 18 2 Preparations Connect one of the probes to the scanner s left most probe connector Turn ON the scanner The 2D Mode window is displayed defau...

Страница 155: ...prove depth penetration and sensitivity This does not affect the frequency used for 2D and M Mode NOTE Frequency setting may affect FPS SV and Low Velocity Reject Adjust PowerOutput to obtain an accep...

Страница 156: ...l 5726264 100 English Rev 8 4 3 21 Select Color M Mode 1 Select M Mode 2 Use the trackball to position the ROI frame over the area to be examined 3 Press Scan Area The instruction Size should be highl...

Страница 157: ...cate disturbed flows Adjust Frequency to optimize the color flow display Higher settings improve resolution Lower settings improve depth penetration and sensitivity This does not affect the frequency...

Страница 158: ...y probe you can first bring up a 2D picture for navigation purpose and then add PW Doppler 4 3 22 2 Preparations Connect one of the probes to the scanner See 4 3 27 Probe Connectors Check on page 4 39...

Страница 159: ...setting will increase the depth penetration Adjust Frame rate to a higher setting to improve motion detection or to a lower setting to improve resolution NOTE Frequency and Frame rate settings may af...

Страница 160: ...ons Connect one of the probes to the scanner s left most probe connector See 4 3 27 Probe Connectors Check on page 4 39 for info about connecting the probes For available probes see 9 3 2 Probe on pag...

Страница 161: ...ade off between the frame rate and quantification noise TVI provides velocity information only in the beam direction The apical view typically provides the best window since the beams are then approxi...

Страница 162: ...er and displays a second active caliper 5 Move the trackball to position the second active caliper at the end point distance or the most posterior point tissue depth 6 Press Set to complete the measur...

Страница 163: ...nnector 8 Carefully position the probe cable in the probe cord holder spot so it is free to move but not resting on the floor CAUTION DO NOT allow the probe head to hang free Impact to the probe head...

Страница 164: ...obe Probes can be disconnected at any time However the probe should not be selected as the active 1 Unlock the probe 2 Pull the probe and connector straight out of the probe port 3 Carefully slide the...

Страница 165: ...Chinese Label H41852LM 4 3 28 3 Preparations None 4 3 28 4 ECG Check Table 4 2 ECG Checks Step Task Expected Result s 1 Connect the ECG harness to the connector under the control panel The unit displ...

Страница 166: ...ode is displayed default mode 4 3 31 Adjust the Cineloop controls Press Freeze The left and right markers are displayed on either side of the last detected heart cycle on the ECG trace Press Freeze Th...

Страница 167: ...s on the LOGIQ V3 V5 V5 Expert s Operator Panel The factory default is Print for the standard printer Store for the screen capture to clipboard 4 3 35 Turn OFF Power to LOGIQ V3 V5 V5 Expert See 4 2 3...

Страница 168: ...V5 Expert Basic Service Manual 5726264 100 English Rev 8 4 3 36 2 Casters Wheels and Brakes Checks Examine the wheels frequently for defects to avoid breaking or jamming Table 4 4 Wheel Characteristi...

Страница 169: ...ing the scanner to customer for use 4 4 1 Software Configuration Checks Table 4 5 Software Configuration Checks Step Task to do Notes 1 Verify Date and Time is correct 2 Verify that Location Hospital...

Страница 170: ...General Procedures and Functional Checks 4 46 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 4 5 Site Log Table 4 6 Site Log DATE SRVICE PERSON PROBLEM COMMENTS...

Страница 171: ...system To access My Trainer 1 Press Alt H to enter My Trainer Or Press Utility System User Configurable Key Check Enable then set My Trainer in the Keyboard Key or User Defined Key Figure 4 23 Set My...

Страница 172: ...Circuit breaker Probe Connection 2 System Setting Facility name set up Date time set up Enable options check Language setting User conf User preset 3 Peripheral Connection Back panel ports Network Co...

Страница 173: ...elect the button to expand the section to show the list of subsections 3 Subsections 4 The highlighted subsection shows this subsection is currently opened and displayed on the right side 5 Exit Press...

Страница 174: ...IQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 9 Text instruction about how to operate Select the arrow to move to the next page Select the arrow to go to the home page Select the a...

Страница 175: ...64 100 English Rev 8 Chapter 5 Components and Functions Theory This chapter explains LOGIQ V3 V5 V5 Expert s system concepts component arrangement and subsystem functions It also describes the power d...

Страница 176: ...eory 5 2 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 5 1 Overview 5 1 1 Contents in this chapter 5 1 Overview on page 5 2 5 2 Block Diagram and Theory on page 5 3 5 3 Common S...

Страница 177: ...m and Theory LOGIQ V3 V5 V5 Expert Basic Service Manual 5 3 5726264 100 English Rev 8 5 2 Block Diagram and Theory 5 2 1 Block Diagram 5 2 1 1 System Diagram Figure 5 1 LOGIQ V3 V5 V5 Expert System Bl...

Страница 178: ...hernet It also interfaces with SDKG board to provide Keyboard KBD LCD Speaker and HDD connection SDKG PWA is backplane board it has a backplane connector which interfaces with SMIO board and provides...

Страница 179: ...nglish Rev 8 5 3 Common Service Platform 5 3 1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems The Common Servi...

Страница 180: ...Components and Functions Theory 5 6 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 181: ...ert Basic Service Manual 6 1 5726264 100 English Rev 8 Chapter 6 Service Adjustments This chapter describes how to test and make adjustments to the LOGIQ V3 V5 V5 Expert You can use these to test the...

Страница 182: ...ervice Adjustments 6 2 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 6 1 Overview 6 1 1 Contents in this chapter 6 1 Overview on page 6 2 6 2 LCD Monitor Adjustments on page 6 3...

Страница 183: ...e defect pixels appearing This section describes how to discriminate the defect pixels with the number of defect pixels allowed for the LOGIQ V3 V5 V5 Expert to support LCD monitor replacement decisio...

Страница 184: ...right side bottom of LCD Figure 6 2 LCD Brightness and Contrast adjust buttons 1 Left key adjustment button 2 Middle key press to display Brightness or contrast indicators 3 Right key adjustment butto...

Страница 185: ...ed To exit the setting menu press the Middle key for the third time The setting menu disappears if there is no operation in 3 to 5 seconds Keep pressing the middle key for 3 to 5 seconds the Brightnes...

Страница 186: ...position the cursor over the adjustment icon then press Set B Pause to display the setting menu Figure 6 7 Setting Menu Icon 2 Follow the procedure below to adjust the Color Temperature Contrast and B...

Страница 187: ...to choose the color temperature The chosen color temperature displays in the Color Temp area Figure 6 9 Color Temperature To adjust the Contrast Press Set B Pause to choose the adjustment button hold...

Страница 188: ...e Set B Pause key and move the trackball to move the adjustment button Move the adjustment button to the left to decrease the brightness or move the adjustment button to the right to increase the brig...

Страница 189: ...chapter describes LOGIQ V3 V5 V5 Expert Series how to setup and run the tools and software that help maintain image quality and system operation Very basic host system and board levels are run whenev...

Страница 190: ...al 5726264 100 English Rev 8 7 1 Overview 7 1 1 Contents in this chapter 7 1 Overview on page 7 2 7 2 Gathering Trouble Data on page 7 3 7 3 Screen Capture on page 7 5 7 4 Global Service User Interfac...

Страница 191: ...lysis There are different options to acquire this data that would give different results 7 2 2 Collect Vital System Information The following information is necessary in order to properly analyze data...

Страница 192: ...ble at all times When Alt D is pressed a menu box appears that allows for a place to enter a description od the issue a checkbox to indicate a System lockup a choice to Export a pre formatted CD R DVD...

Страница 193: ...ce Manual 7 5 5726264 100 English Rev 8 7 3 Screen Capture There may be times when the customer or field engineer will want to capture a presentation on the screen This is accomplished by first saving...

Страница 194: ...Capturing a Screen The following is a generic process to capture any screen from the scanner 1 Navigate to and display the image screen to be captured 2 Press Store This will place a snapshot of the s...

Страница 195: ...napshot to be stored 4 Select Save As Icon on the right side of the image screen Figure 7 3 Menu Save As 5 A Save dialog box will be opened Choose the archive location to save image on the USB Drive o...

Страница 196: ...ting 7 8 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 7 3 1 Capturing a Screen continued NOTE It is better to save the image in Jpeg format Image of this format can be easily r...

Страница 197: ...pture the whole screen from the scanner 1 Insert the Service Dongle in the system s USB port 2 Press NL SL key on the control panel the whole screen is captured 3 Press Ctrl Alt Delete the Windows Sec...

Страница 198: ...ert Basic Service Manual 5726264 100 English Rev 8 7 3 2 Capturing a Screen with Service Dongle continued 4 In Windows Task Manager select File New Task Run Figure 7 6 Windows Task Manager 5 Type in t...

Страница 199: ...ice Manual 7 11 5726264 100 English Rev 8 7 3 2 Capturing a Screen with Service Dongle continued 6 The paint screen is displayed press Ctrl V to paste the captured screen And then save the image in th...

Страница 200: ...100 English Rev 8 7 4 Global Service User Interface GSUI 7 4 1 Enter global service user interface 1 Select Utility Service then the service browser will start Figure 7 9 2 Select GE Service in option...

Страница 201: ...vice Manual 7 13 5726264 100 English Rev 8 7 4 1 Enter global service user interface continued Figure 7 10 Service Login Page 3 Choose Diagnostic in Global Service Interface to active diagnostic funct...

Страница 202: ...Diagnostics Troubleshooting 7 14 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 7 4 1 Enter global service user interface continued Figure 7 11 Active Diagnostic Function...

Страница 203: ...t numeric values with 4 digits or less When the button is configured as an execute button and pressed the loop count field will be queried to determine the number of times to execute the diagnostic 7...

Страница 204: ...e Mode or Not After submitting to restart or shutdown a confirmation screen gives one last chance to confirm or cancel the request 7 4 3 1 1 Remote access Remote access is a feature designed to enable...

Страница 205: ...erface GSUI LOGIQ V3 V5 V5 Expert Basic Service Manual 7 17 5726264 100 English Rev 8 7 4 3 1 27 4 3 1 1 Remote access continued Figure 7 13 Remotely Connect 3 Enter Utility Common Utilities Disruptiv...

Страница 206: ...Remote access continued Figure 7 14 Enable Virtual Dongle 4 If we want to see the Scan screen or have control of the unit we have to go to Virtual Console Observation and Start it Then connect using...

Страница 207: ...7 19 5726264 100 English Rev 8 7 4 3 1 47 4 3 1 1 Remote access continued Figure 7 16 Virtual Console Observation Connect 7 4 3 2 Service Diagnostics BF COMEXP and GFS modules are on the LOGIQ V3 V5 V...

Страница 208: ...the GFS GFS Front End Interface Test tests that the GFS can access Front End boards GFS Analog Test GFS Memory Access Test tests that GFS can access to the internal external external cache memory spa...

Страница 209: ...P Test tests HV STOP mechanism and checks each board is able to assert HV_STOP using its HV_STOP source 7 4 3 2 3 Memory Callisto Memory Test NOTE The Callisto Memory test may fail if it is performed...

Страница 210: ...is successful you will see a brief video clipo with audio For more information about the test click More Information Click Pass if the test successfully reproduces the video clip Click Fail if the tes...

Страница 211: ...visually verifying the monitor color depth capacity If the colors in the color spectrum do no blend smoothly together a problem may exist with the monitor color depth Use the graduated horizontal and...

Страница 212: ...the direction the wheel is being rotated Clicking the wheel will flash the picture of the mouse in the Mouse Status Test area The Drag and Drop Test verifies a mouse can successfully perform drag and...

Страница 213: ...nwrap the probe cord 3 Put the probe in the probe holder 4 Align the connector with the probe port and carefully push into the left most probe port Figure 7 17 Left most probe port 5 Turn the connecto...

Страница 214: ...Rev 8 7 4 3 2 147 4 3 2 13 Assessment Utility continued 9 In the Diagnostic Service Diagnostics Assessment Utility Probe Assessment to check the probe If there is no broken element and the probe is a...

Страница 215: ...glish Rev 8 7 4 3 2 157 4 3 2 13 Assessment Utility continued If there are broken elements and the probe is not acceptable for diagnostic The assessment is fail and the failed elements are shown in th...

Страница 216: ...Probe Pass Fail Criteria Pass No dead element Dead elements on the edge 0 1 or end 1 end Fail Max 2 weak elements on the scan line 3dB elements 2dB Any one of the elements in one scan line element 3d...

Страница 217: ...00 English Rev 8 7 5 Network and Insite II Configuration 7 5 1 Network Configuration 7 5 1 1 Wire LAN Network 1 Connect system with network 2 Enter Utility Connectivity TCP IP in IP settings window ch...

Страница 218: ...n input static address in IP Address box Subnet Mask and Default Gateway box In Network Speed choose the proper speed available Figure 7 21 Input static address 3 Select Save and a popup window displa...

Страница 219: ...Network and Insite II Configuration LOGIQ V3 V5 V5 Expert Basic Service Manual 7 31 5726264 100 English Rev 8 7 5 1 Network Configuration continued Figure 7 23 Network icon...

Страница 220: ...configure the Wireless LAN network the operator must login as administrator 1 Connect the wireless adapter in the USB port 2 Enter Utility Connectivity TCP IP press Configuration Figure 7 24 Network S...

Страница 221: ...726264 100 English Rev 8 7 5 1 Network Configuration continued 4 Input Network Key and Confirm Network Key then click Connect Figure 7 26 Connection Status 5 The network icon at the left bottom of scr...

Страница 222: ...Diagnostics Troubleshooting 7 34 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 223: ...5726264 100 English Rev 8 Chapter 8 Replacement Procedures This chapter describes how to remove and install or replace modules and subsystems in the LOGIQ V3 V5 V5 Expert It also includes instructions...

Страница 224: ...LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 8 1 Overview 8 1 1 Contents in this chapter 8 1 Overview on page 8 2 8 2 Warnings and important information on page 8 3 8 3 Loadin...

Страница 225: ...rs to discharge as there are no test points to verify isolation 6 Remove disconnect the battery if present Ultrasound System components may be energized WARNING Because of the limited access to cabine...

Страница 226: ...uipment with visible body fluids be taken or shipped from a clinic or site for example body coils or an ultrasound probe The purpose of the regulation is to protect employees in the transportation ind...

Страница 227: ...ove the images to a remote server or to removable media For instructions please see Disk management in the Basic User Manual User Guide CAUTION An error or a power loss may occur Always backup the Pat...

Страница 228: ...dunn userdefs CAUTION An error or a power loss may occur Always backup the Patient Archive and the Presets System Configurations before loading the software In order to complete a successful restore o...

Страница 229: ...3 V5 V5 Expert on paper Verify if it is current before starting a software loading If needed record the settings This subsection includes descriptions for recording data from the following screens CAU...

Страница 230: ...is believed to be unnecessary it would not hurt to disconnect the system from the network and remove all transducer NOTE Please ensure AC adapter is connected during system upgrade 1 Insert the USB me...

Страница 231: ...mpacted To select 2 wipe tool This option is to wipe tool all the patient data NOTE All the patient data cannot be recovered after wipe tool To select 3 Service tool The complete disk will be loaded T...

Страница 232: ...m software the system will display a screen to restore Computer name before entering the scanning screen 5 System USB memory stick will be loading Wait for the software installation to complete Typica...

Страница 233: ...continued Figure 8 2 Loading status 6 After finish updating system remove the USB memory stick and press any key to shut down the system NOTE If you do not remove the USB memory stick the software sys...

Страница 234: ...While it is believed to be unnecessary it would not hurt to disconnect the system from the network and remove all transducer NOTE Please ensure AC adapter is connected during system upgrade 1 Insert t...

Страница 235: ...d 3 Power on the system The system will detect the USB memory stick automatically 4 Select 2 Wipe Tool and then select Next to continue Figure 8 4 Upgrade message 5 The system indicates all data will...

Страница 236: ...Expert Basic Service Manual 5726264 100 English Rev 8 8 3 7 2 Wipe Tool continued 6 The process may take twenty minutes or more Figure 8 6 Wipe Tool Process 7 Select OK when the process is complete a...

Страница 237: ...tick NOTE While it is believed to be unnecessary it would not hurt to disconnect the system from the network and remove all transducer NOTE Please ensure AC adapter is connected during system upgrade...

Страница 238: ...m will not shutdown normally hold down the Power On Off Switch until the light turns off 3 Power on the system The system will detect the USB memory stick automatically 4 Select one of the options for...

Страница 239: ...oading the system with USB memory stick continued 5 This option is used by service engineer only It needs password to continue the password is GEWuXiChina and it is case sensitive Figure 8 9 System up...

Страница 240: ...lish Rev 8 8 3 7 3 Loading the system with USB memory stick continued Figure 8 11 System upgrade complete 8 System USB memory stick will be loading Wait for the software installation to complete NOTE...

Страница 241: ...remove the USB memory stick the software system loading process repeats when the system boots up Do not remove the USB stick after the software reload if you want to adjust the screen size Figure 8 13...

Страница 242: ...C adapter is connected during system upgrade 1 Do not remove the USB stick after the software reload if you want to adjust the screen size NOTE Do not remove the USB stick after the software reload if...

Страница 243: ...e 5 Select the screen size according to the monitor size which you can get from the rear cover of monitor Figure 8 16 Screen Size 6 After finish updating system remove the USB memory stick and press a...

Страница 244: ...the Serial Number If selected to load the complete disk during the software loading process when powering the system the system will indicate to rewrite the serial number 1 If the serial number on VPD...

Страница 245: ...on page 8 23 input the password and select OK If you will input the same serial number as the original in the step 2 on page 8 23 you can select Ignore without input of the password Then the step 2 o...

Страница 246: ...xpert Basic Service Manual 5726264 100 English Rev 8 8 3 8 Rewrite the Serial Number continued 3 The serial number is reset reboot the system Figure 8 21 Reboot the System NOTE Make sure the system is...

Страница 247: ...V3 V5 V5 Expert system and wait until system booting to scanning screen NOTE If selected to load C Disk only when loading the system software the system will display a screen to restore Computer name...

Страница 248: ...rrect Preset Region NOTE After the system software loading completion please reload the correct preset region 1 Reboot the system 2 Press Utility on the Control Pane then select System General 3 In th...

Страница 249: ...and working 1 Reboot the system 2 Press Utility on the control panel 3 Select Admin System Admin 4 Ensure that all the installed option keys are displayed and the status of Options are valid The stat...

Страница 250: ...is displayed on the right top location of the screen About the probe specification for intended use on LOGIQ V3 V5 V5 Expert Plug in at least one of each type of the probes and check if each of the pr...

Страница 251: ...all any DICOM devices used by the customers and check to ensure these DICOM devices work properly The instruction about installing DICOM devices is not incorporated in this manual To access the instru...

Страница 252: ...Replacement Procedures 8 30 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 253: ...LOGIQ V3 V5 V5 Expert Basic Service Manual 9 1 5726264 100 English Rev 8 Chapter 9 Renewal Parts This chapter lists the renewal parts available for the LOGIQ V3 V5 V5 Expert...

Страница 254: ...Renewal Parts 9 2 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8 9 1 Overview 9 1 1 Contents in this chapter List of Abbreviations Renewal Parts Lists...

Страница 255: ...bbreviations Table 9 1 List of Abbreviations ABBREVIATION DESCRIPTION 3D THREE DIMENSIONAL Assy ASSEMBLY FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufact...

Страница 256: ...ic Service Manual 5726264 100 English Rev 8 9 3 Renewal Parts Lists 9 3 1 Operator Console Assy The following figure illustrates what is the front side 1 left side 2 back rear side 3 right side 4 of t...

Страница 257: ...5 20 1 1 706 E8Cs RS 5670375 6 5 20 1 1 707 3Sc RS 47237516 2 75 20 1 1 708 6S RS 5499316 4 5 20 1 1 709 RAB2 6 RS KTZ303982 3 3 10 1 1 710 L6 12 RS 5454332 7 75 20 1 1 711 12L RS 5499501 7 75 20 1 1...

Страница 258: ...53 SONY UPD25 Color Printer BRAZIL kit 1 818 5151259 2 BW PRINTER UP D898MD USA 1 819 5151261 2 BW PRINTER UP D898MD EU 1 820 5151262 2 BW PRINTER UP D898MD CHN 1 821 5151263 2 BW PRINTER UP D898MD JP...

Страница 259: ...R2 R2 0 4 System and Application Soft ware for Service 1 838 5730360 5S Ayan R2 R2 0 5 System and Application Soft ware for Service 1 839 5730360 6S Ayan R2 R2 0 6 System and Application Soft ware for...

Страница 260: ...1 5176304 2 AC Power Cord China 1 1 902 5176773 2 AC Power Cord India 1 1 903 5177195 2 AC Power Cord Argentina 1 1 904 5176907 2 AC Power Cord UK 1 1 905 5177153 2 AC Power Cord Denmark 1 1 906 51771...

Страница 261: ...de Brazilian Portuguese version 1 N 1012 5719188 129 LOGIQ V3 V5 V5 Expert User Guide Estonian version 1 N 1013 5719188 131 LOGIQ V3 V5 V5 Expert User Guide Slovenian version 1 N 1014 5719188 140 LOGI...

Страница 262: ...V5 V5 Expert Release Notes French version 1 N 1042 5726260 106 LOGIQ V3 V5 V5 Expert Release Notes Spanish version 1 N 1043 5726260 108 LOGIQ V3 V5 V5 Expert Release Notes German version 1 N 1044 572...

Страница 263: ...tes Romanian version 1 N 1064 5726260 168 LOGIQ V3 V5 V5 Expert Release Notes Croatian version 1 N 1065 5726260 174 LOGIQ V3 V5 V5 Expert Release Notes Lithuanian version 1 N 1066 5726260 175 LOGIQ V3...

Страница 264: ...Renewal Parts 9 12 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 265: ...nd Maintenance This chapter describes Care and Maintenance on the Ultrasound system and peripherals These procedures are intended to maintain the quality of the Ultrasound system s performance Read th...

Страница 266: ...10 2 Warnings on page 10 3 10 3 Why do maintenance on page 10 4 10 4 Maintenance task schedule on page 10 6 10 5 Tools required on page 10 8 10 6 System maintenance on page 10 12 10 7 Electrical safe...

Страница 267: ...BE CAUTIOUS WHENEVER POWER IS STILL ON AND COVERS ARE REMOVED CAUTION Practice good ESD prevention Wear an anti static strap when handling electronic parts and even when disconnecting connecting cable...

Страница 268: ...stomers Quality Assurance Programs may require additional tasks and or inspections at a different frequency than listed in this manual 10 3 2 Keeping records It is good business practice that ultrasou...

Страница 269: ...cal physicists the supervising radiologist physician or appropriate designee Routine quality control testing must occur regularly The same tests are performed during each period so that changes can be...

Страница 270: ...of your LOGIQ V3 V5 V5 Expert and can best provide competent efficient service Contact GE for coverage information and or price for service The service procedures and recommended intervals shown in t...

Страница 271: ...ice Annually Peripheral Leakage Current Checks See Notes Twice Annually Surface Probe Leakage Current Checks See Notes Twice Annually Endocavity Probe Leakage Current Checks See Notes Quarterly Annual...

Страница 272: ...rench Open End 1 2 9 16 9 XL9974MM Xcelite hex Blade 2 5mm 9 44579 Wrench Open End 1 2 9 16 9 XL9975MM Xcelite hex Blade 3mm 9 45385 Pliers Arc Joint 7 inch 9 XL9976MM Xcelite hex Blade 4mm 9 45378 Pl...

Страница 273: ...Blade 5 32 inch 9 43499 1 2 inch Socket 1 4 inch drive 9 XL9926 Xcelite hex Blade 3 16 inch 9 4355 Flex Spinner 9 XL99764 Xcelite hex Blade 7 64 9 43523 Breaker 9 XL99964 Xcelite hex Blade 9 64 9 4353...

Страница 274: ...2886 Punch Pin 1 4 inch 9 34374 3 8 inch Metric Socket Set 12 PT 9 42973 Cold Chisel 1 2 inch 9 44311 16mm Socket 12 pt 9 GH77 Center Punch Automatic 9 33485 Metal Socket Tray 9 GH890 File Handle Adj...

Страница 275: ...ist strap 3M 2214 Small adjustable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 120V 230V Safety Analyzer The safety Analyzer tool should be calibrated and compliant with AAMI...

Страница 276: ...nd Ultrasound system options 3 Power up Turn the Ultrasound system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messa...

Страница 277: ...robe Elements Perform an Element Test on each probe to verify that all the probe elements and system channels are functional 6 Applicable Software Options Verify the basic operation of all optional mo...

Страница 278: ...rinter Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 ECG Verify basic o...

Страница 279: ...terior for dents scratches or cracks 3 Input Power Refer to 10 6 2 3 Mains cable inspection on page 10 14 4 Cables Connectors Check all internal cable harnesses and connectors for wear and secure conn...

Страница 280: ...ck the temperature log to see if there are any trends that could cause problems in the future 3 Check the Configuration Log update if needed 10 Power and System Status Indicators Check for proper oper...

Страница 281: ...ipment damage Failure to properly handle or maintain a probe may also void its warranty Any evidence of wear indicates the probe cannot be used Do a visual check of the probe pins and Ultrasound syste...

Страница 282: ...are made of nitrile derived from vegetable starch to prevent allergic latex reactions CAUTION Failure to follow the prescribed cleaning or disinfection procedures will void the probe s warranty DO NOT...

Страница 283: ...RY 12 MONTHS ACCORDING TO THE HISTORICAL DATA ONLY TRAINED PERSONS ARE ALLOWED TO PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE DANGER TO MINIMIZE RISK OF ELECTRICAL SHOCK ONLY TRAINED PERSONS ARE AL...

Страница 284: ...t the cables and strain relief at each end Inspect the transducer enclosure and lens for cracks holes and similar defects Test the system peripherals and probes for leakage current Excessive leakage c...

Страница 285: ...seconds for capacitors to discharge as there are no test points to verify isolation 6 Remove disconnect the battery if present Ultrasound System components may be energized CAUTION Compare all safety...

Страница 286: ...1 Table 10 14 Type CF Applied Part Leakage Current Limits ECG Connections Country Normal Condition Open Ground Reverse Polarity Open Neutral Mains Applied All 0 1 mA 0 5 mA 0 5 mA 0 5 mA 5 0 mA Table...

Страница 287: ...should be reported to the hospital immediately and the receptacle should not be used Figure 10 1 Typical alternate outlet tester NOTE No outlet tester can detect the condition where the Neutral ground...

Страница 288: ...exposed metal parts of the case The ground wire resistance should be less than 0 2 ohms Reference the procedure in the IEC60601 1 Figure 10 2 Ground continuity test DANGER ELECTRIC SHOCK HAZARD THE PA...

Страница 289: ...d or a similar test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the gr...

Страница 290: ...12 on page 10 21 Record all data on the PM Inspection Certificate Table 10 16 Typical Data Sheet for enclosure Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground Switch...

Страница 291: ...ltage and test lead placement It is abnormal if no leakage current is measured If no leakage current is detected check the configuration of the test equipment 10 7 6 2 Tools For needed tools see 10 5...

Страница 292: ...Meter Procedure Using Probe Adapter Follow the Safety Analyzer tool instruction to test each transducer for leakage current The electrical Safety Analyzer tool should be calibrated and compliant with...

Страница 293: ...ver switch the Polarity and the status of Neutral when the Ultrasound system is powered ON Be sure to turn the Ultrasound system power OFF before switching them using the POLARITY switch and or the NE...

Страница 294: ...rt Where applicable tighten all grounds Ensure star washers are under all ground studs Inspect wiring for bad crimps poor connections or damage Test the wall outlet verify it is grounded and is free o...

Страница 295: ...obe type in the appropriate space on the check list If excessive leakage current is slot dependent inspect the system connector for bent pins poor connections and ground continuity If the problem rema...

Страница 296: ...S e m a N r e m o t s u C Date Performed Warranty C ontract HBS System Type e t a D e r u t c a f u n a M r e b m u N l a i r e S r e b m u N l e d o M r e b m u N l a i r e S r e b m u N l e d o M t...

Страница 297: ...p p o D O I l a n r e t x E n o i t c n u F e d o M F C s r o t c e n n o C d n a s e l b a C n o i t c n u F e d o M M Applicable Software Opti l a r e h p i r e P d e v o r p p A E G s n o s DVD RW...

Страница 298: ...alue measured OK Comments Outlet correct ground and wiring config System ground continuity Chassis source leakage current probe Chassis source leakage current wheel Chassis source leakage current moni...

Страница 299: ...heral 2 ground continuity Peripheral 3 leakage current Peripheral 3 ground continuity Table 10 18 Electrical safety tests log Continued Electrical test performed Max value allowed Value measured OK Co...

Страница 300: ...Care and Maintenance 10 36 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 301: ...2 5 copyrights i 15 customer assistance 1 29 phone numbers 1 30 customer order verification 3 13 customer provided prerequisite 8 5 D damage in transport i 12 dangerous procedure warnings 1 24 data n...

Страница 302: ...l notes i 15 lighting 2 5 list of abbrevations 9 3 loading software 8 5 lockout tagout LOTO requirements 1 25 LOTO lockout tagout requirements 1 25 M mains cable inspection 10 14 maintenance physical...

Страница 303: ...lectrical specifications 3 16 physical dimensions 3 15 standard GE tool kit 10 8 standard hazard icons 1 8 statement certified electrical contractor statement i 12 system specifications 3 14 system ch...

Страница 304: ...Index 4 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...

Страница 305: ...GE...