Disinfecting probes (continued)
CAUTION
CREUTZFELDT-JACOB DISEASE
This device is not indicated for neurological use. Neurological
contact on patients with this disease must be avoided. If a
device/probe becomes contaminated, there is no adequate
means to disinfect it. In this case, the contaminated device/
probe must be discarded in accordance with local biologic
waste hazard procedures.
Biological
Hazard
WARNING
Ultrasound transducers can easily be damaged by improper
handling and by contact with certain chemicals. Failure to
follow these precautions can result in serious injury and
equipment damage.
•
Do not immerse the probe into any liquid beyond the level
specified for that probe. Never immerse the transducer
connector or probe adapters into any liquid.
•
Avoid mechanical shock or impact to the transducer and do
not apply excessive bending or pulling force to the cable.
•
Transducer damage can result from contact with
inappropriate coupling or cleaning agents:
•
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride
compounds or hydrogen peroxide
•
Avoid contact with solutions or coupling gels containing
mineral oil or lanolin
•
Avoid temperatures above 50
°
C.
•
Inspect the probe prior to use and after use for damage or
degeneration to the housing, strain relief, lens and seal.
Do not use a damaged or defective probe.
Содержание LOGIQ V2
Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 This page intentionally left blank...
Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 170: ...After the Exam is Over 3 8 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 Figure 3 10 Formats selection...
Страница 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 381: ......