LOGIQ
LOGIQ E10 Series: Battery Options
Refresh Procedure
:
1.
Plug the AC cord to the wall outlet and turn on the circuit
breaker.
2.
Wait until the battery is fully charged. This takes about 2.5 hours
to fully charge the battery (depending on the remaining battery
capacity).
3.
Wait at least 1 hour.
4.
Remove all probes.
5.
Turn on the system.
6.
Unplug the AC cable, letting the system run on battery until it
automatically shuts down. It may take 30 minutes (Power
Assistant Battery)/ 50 minutes (Scan on Battery) to complete
shut down.
7.
Wait at least 5 hours.
8.
Plug the AC cable to the wall outlet and turn on the circuit
breaker.
9.
Wait until the battery is fully charged. It takes about 2.5 hours to
fully charge the battery.
Refreshing the Battery –
helps maintain the battery life and accuracy of the fuel gauge and
helps to avoid unexpected shutdowns. Recommended every 6 months.
Battery Deterioration
When the system detects battery deterioration
the following message appears.
If this Message appears, the LOGIQ E10s disables both
the Power Assistant and Scan on Battery.
Contact the technical service department or your
local GE representative.
LOGIQ is a trademark of the General Electric Company. DICOM is a registered trademark of the National Electrical
Manufacturers Association. Third party trademarks are the property of their respective owners. April 2020 JB79491XX