background image

LOGIQ™ E10 Series 

Power Assistant and Scan on Battery Guide

www.gehealthcare.com

The  battery protects system data for a limited amount of time 
in the event of inadvertent AC power loss. 

Options: 

Power Assistant-

system power is maintained by the battery 

when AC power is lost or the power cord is unplugged. 
Reduces the start up time for portable exams. The system will 
run up to 20 minutes after AC power is disconnected. 

Scan on Battery 

(Option on LOGIQ E10s only)

- allows the 

user to scan using battery power (AC unplugged). With a fresh 
battery that is fully charged, the LOGIQ E10s stays powered 
for approximately 50 minutes.

Note: Detailed battery function and service information is outlined 
in the 

Basic User Manual

.

Содержание LOGIQ E10 Series

Страница 1: ...C power is lost or the power cord is unplugged Reduces the start up time for portable exams The system will run up to 20 minutes after AC power is disconnected Scan on Battery Option on LOGIQ E10s only allows the user to scan using battery power AC unplugged With a fresh battery that is fully charged the LOGIQ E10s stays powered for approximately 50 minutes Note Detailed battery function and servi...

Страница 2: ... following actions ensure that they are complete Any export input or Save As to external media such as CD DVD USB DICOMⓇ Retrieve from Query Retrieve DVR recording should be stopped Exit Volume Navigation by Exit and Clear if you are using Volume Navigation PA Flashing lights indicate the battery level is LOW Flashing operator panel lights are meant to notify the user that the battery charge level...

Страница 3: ... will run for up to 20 minutes and a Running on battery messaging will temporarily appear at the bottom of the monitor 5 When in PA mode most of the console devices such as the keyboard and printers are turned off to minimize battery usage 6 When the system is not in use plug it into a power source to recharge the battery 7 If an operation was interrupted the system displays a notification letting...

Страница 4: ...is not available on your system Detailed battery information will display in the pop up which includes the battery s current capacity Power Assistant Battery Status Icon AC unplugged Battery 90 100 AC unplugged Battery 50 89 AC unplugged Battery 20 49 AC unplugged Battery 0 19 AC plugged charging Battery 90 100 AC plugged charging Battery 50 89 AC plugged charging Battery 20 49 AC Plugged charging...

Страница 5: ...ansfer patient data Using AC power is strongly recommended in these instances The system can unexpectedly shut down if the battery power is too low resulting in patient data loss Note the message in the status bar when Scan on Battery is activated System is using battery power Make sure the battery has enough capacity to prevent potential data loss Switch to Power Saving Mode The system switches t...

Страница 6: ... AC power when the fuel gage icon is yellow or red Battery Charging Whether the system is on or off the battery will charge as long as the system is connected to AC power via the power cord and the breaker is on A fan will run while the battery is charging The fan will stop when the battery is done charging While the battery is charging the front LED will blink When the battery is fully charged th...

Страница 7: ...outlet and turn on the circuit breaker 9 Wait until the battery is fully charged It takes about 2 5 hours to fully charge the battery Refreshing the Battery helps maintain the battery life and accuracy of the fuel gauge and helps to avoid unexpected shutdowns Recommended every 6 months Battery Deterioration When the system detects battery deterioration the following message appears If this Message...