Section 2-3 - Facility Needs
2-3-3
Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
LINEN SUPPLY
EXAMINA
TION
TABLE
FILM VIEWER
COUNTER TOP
FILE
CABINET
SECRETARYS OR
DOCTOR’S DESK
FILM
PROCESSING
ROOM,
OVERHEAD
LIGHTS DIMMER
DEDICATED
POWER
RECEPTACLE
EXTERNAL
PERIPHERALS
SINK
COUNTER
TO
P
FILM VIEWER
FILM SUPPLIES
SINK
LINEN SUPPLY
PROBES/SUPPLIES
24 IN.
(61 CM)
EXAMINARION
TABLE
76 IN.
(193 CM)
DOOR
30 IN.
(76 CM)
An 8 by 10 foot Minimal Floor Plan
DEDICATED POWER
OUTLETS
DEDICATED ANALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
A 14 by 17 foot Recommended Floor Plan
GE CABINET
FOT SOFTWARE
AND MANUALS
Scale : Each square equals one square foot
DOOR
42 IN.
(107 CM)
DEDICATED ALALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
SUCTION LINE
EMERGECY
OXYGEN
24 IN.
(61 CM)
76 IN.
(193 CM)
PATIENT
TOILET
FACILITY
FOOT
SW
FOOT
SW
LOGIQ 5
CONSOLE
CONSOLE
18 IN.
(46 CM)
STOOL
STOOL