Powering On the System
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
3–22
Circuit breaker
The Circuit Breaker is located on the back of the console, at the
bottom of the system. On supplies main power to all internal
systems. Off removes main power from all internal systems.
The circuit breaker automatically shuts off power to the system
in case of a power overload.
If a power overload occurs:
1.
Turn off all peripheral devices.
2.
Turn off the Main Power Switch to the console.
3.
Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the On position; DO
NOT hold the switch in the On position. If the Circuit Breaker
switch remains On, follow the
Power On procedure previously
described.
NOTE: If the Circuit Breaker switch does not remain in the On
position or trips again:
1.
Disconnect the Power Cable.
2.
Call Service immediately.
DO NOT attempt to use the system.
1
0
.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...