Section 8-8 - PC Block
8-8-1-12
Check the Device Driver Configuration.(BEP2)
After System boot up. confirm the installed device information.
Double click Go Device Manager Icon.
1.) Check The PCI Bridge Between FEC Assy and PC2IP Assy under multi function adapters
2.) Check the Sound Driver is On board Sound (Sound Max).
3.) Check the VIC Assy Driver is GE Medical SuperSonic Diagnostic Device Driver and GEMSK
Multimedia Device
4.) Check The USB Device Driver.
Figure 8-116 Device Manager
Figure 8-117 Multi Function Adapter
Figure 8-118 Sound Driver
Device Manger
VIC assy
On board sound