background image

GE MEDICAL SYSTEMS

2127661

LOGIQ   400 SERVICE MANUAL

TM

8–57

OPTIONS

REV 5

8–5–8 Operational Check–out

1.

Connect the ECG patient cable with the connector on the ECG panel as shown in ILLUSTRATION 8–62.

CONNECTING ECG PATIENT CABLE

ILLUSTRATION 8–62

2.

Connect the electrodes of ECG patient cable with the ECG simulator.

3.

Plug in the system and turn the system power ON.

Содержание LOGIQ 400

Страница 1: ...2127661 Revision 9 LOGIQ 400 Service Manual Copyright 1995 1996 1997 1998 1999 2000 by General Electric Company ...

Страница 2: ...0 Software version 5 01 release and error correction PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV Title page 9 GE Logo page P9030CB 2 P9030CC 4 P9030CD 0 A 9 i 8 ii to x 9 Chapter 1 1 1 8 1 2 to 1 3 0 1 4 2 1 5 to 1 8 8 1 9 7 1 10 to 20 8 Chapter 2 2 1 5 2 2 to 2 4 0 2 5 5 2 6 to 2 9 0 2 10 9 2 11 to 2 16 0 Chapter 3 3 1 9 3 2 0 3 3 to 3 16 9 Chapter 4 4 1 6 4 2 0 4 3 2 4 4 1 4 5 0 4 6 1 4 7 0 4 8...

Страница 3: ...tice on Service 1 19 1 4 ADDRESS 1 20 CHAPTER 2 INSTALLATION 2 1 PREINSTALLATION 2 3 2 1 1 Introduction 2 3 2 1 2 Power Line Requirements 2 3 2 1 3 Physical Specifications 2 4 2 1 4 Recommended Ultrasound Room Layout 2 5 2 2 INSTALLATION 2 8 2 2 1 Introduction 2 8 2 2 2 Average Installation Time 2 8 2 2 3 Installation Warnings 2 8 2 2 4 Checking the Components 2 8 2 2 5 Unpacking LOGIQ 400 2 9 2 2...

Страница 4: ... Peripherals 3 15 3 6 VIDEO SPECIFICATIONS 3 16 CHAPTER 4 FUNCTIONAL CHECKS 4 1 INTRODUCTION 4 3 4 1 1 Required Equipment 4 3 4 2 FUNCTIONAL CHECK PROCEDURES 4 4 4 2 1 Basic Controls 4 4 4 3 DIAGNOSTICS 4 10 4 3 1 Service Software Menu 4 10 4 3 2 Diagnosis Test Menu 4 11 4 3 3 Utility Menu 4 16 4 4 POWER SUPPLY ADJUSTMENTS 4 37 4 4 1 Power Supply Access 4 38 4 4 2 Power Supply Adjustment Procedure...

Страница 5: ...ter Clamp Set FRU No 156 6 97 6 2 17 Monitor Bottom Assy FRU No 157 6 98 6 2 18 CRT Assy NTSC FRU No 158 CRT Assy PAL FRU No 159 6 100 6 2 19 CRT Cable Set FRU No 160 6 102 6 2 20 CRT Lamp FRU No 161 6 103 6 2 21 Monitor15 Assy FRU No 170 6 104 6 2 22 Monitor15 Cover Set FRU No 171 6 106 6 2 23 Speaker Assy FRU No 178 6 109 6 2 24 Task Lamp FRU No 180 6 110 6 2 25 CRT Cable Set FRU No 182 6 111 6 ...

Страница 6: ... FRU No 459 6 162 6 2 60 Rotary Encoder FRU No 460 6 163 6 2 61 Rear CONN Panel Assy FRU No 501 6 166 6 2 62 CNTIF Assy FRU No 502 6 168 6 2 63 AVIF Assy or BVIF Assy FRU No 503 6 169 6 2 64 Circuit Protector FRU No 504 15A FRU No 505 7 5A 6 170 6 2 65 ECG Board Assy FRU No 506 6 171 6 2 66 ECG Panel Assy FRU No 507 6 172 6 2 67 Nest Fan Assy FRU No 511 6 174 6 2 68 Probe CONN Set Assy FRU No 512 ...

Страница 7: ...752 6 234 6 2 96 TRIAC Unit FRU No 753 6 236 6 2 97 Power Inlet Unit2 FRU No 754 for 115V FRU No 755 for 220V 6 238 6 2 98 LV3 Unit FRU No 770 6 240 6 3 FUSE REPLACEMENT 6 243 6 3 1 Introduction 6 243 6 3 2 Replacement Procedures 6 244 6 4 SYSTEM SOFTWARE INSTALLATION 6 251 6 4 1 Introduction 6 251 6 4 2 Preparing before Installation 6 252 6 4 3 Initializing Hard Disk 6 253 6 4 4 Installing Softwa...

Страница 8: ...d Continuity 7 8 7 3 3 Chassis Leakage Current Test 7 9 7 3 4 Probe Leakage Current Test 7 12 7 3 5 ECG Leakage Current Test 7 16 7 3 6 When There s Too Much Leakage Current 7 20 CHAPTER 8 OPTIONS 8 1 INTRODUCTION 8 5 8 2 VIDEO CASSETTE RECORDER INSTALLATION 8 6 8 2 1 Foreword 8 6 8 2 2 Tools Required 8 6 8 2 3 Time Required 8 6 8 2 4 Parts Required 8 6 8 2 5 Functional Check out 8 7 8 2 6 Install...

Страница 9: ...Renewal Parts 8 41 8 4 B W PRINTER INSTALLATION 8 42 8 4 1 Foreword 8 42 8 4 2 Tools Required 8 42 8 4 3 Time Required 8 42 8 4 4 Parts Required 8 42 8 4 5 Functional Check out 8 42 8 4 6 Setting DIP Switches 8 43 8 4 7 Installing B W Video Printer on Console 8 43 8 4 8 Connecting Cables 8 47 8 4 9 Operational Check out 8 48 8 4 10 Final Procedures 8 51 8 4 11 Renewal Parts 8 51 8 5 ECG OPTION INS...

Страница 10: ...Check out 8 74 8 6 9 Final Procedures 8 75 8 6 10 Renewal Parts 8 77 8 7 FOOT SWITCH INSTALLATION 8 78 8 7 1 Foreword 8 78 8 7 2 Tools Required 8 78 8 7 3 Time Required 8 78 8 7 4 Parts Required 8 78 8 7 5 Functional Check out 8 78 8 7 6 Connecting Foot Switch 8 79 8 7 7 Operational Check out 8 80 8 7 8 Final Procedures 8 80 8 8 SWIVEL LOCK INSTALLATION 8 81 8 8 1 Foreword 8 81 8 8 2 Tools Require...

Страница 11: ...8 9 2 Tools Required 8 85 8 9 3 Time Required 8 85 8 9 4 Parts Required 8 85 8 9 5 Installing Left Side Probe Holder 8 85 8 9 6 Operational Check out 8 86 8 9 7 Final Procedures 8 86 8 10 MTZ PROBE HOLDER INSTALLATION 8 87 8 10 1 Foreword 8 87 8 10 2 Tools Required 8 87 8 10 3 Time Required 8 87 8 10 4 Parts Required 8 87 8 10 5 Installing MTZ Probe Holder 8 88 8 10 6 Final Procedures 8 88 ...

Страница 12: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 TABLE OF CONTENTS x REV 9 This page is left blank intentionally ...

Страница 13: ... 1 1 SERVICE MANUAL CONTENTS 1 3 1 2 SAFETY 1 4 1 2 1 Warnings 1 4 1 2 2 Specifications 1 17 1 3 EMC Electromagnetic Compatibility 1 18 1 3 1 EMC Performance 1 18 1 3 2 Notice upon Installation of Product 1 18 1 3 3 General Notice 1 19 1 3 4 Countermeasures against EMC related Issues 1 19 1 3 5 Notice on Service 1 19 1 4 ADDRESS 1 20 ...

Страница 14: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 INTRODUCTION 1 2 REV 0 This page is left blank intentionally ...

Страница 15: ...st 3 Chapter 3 System Configuration Contains system configuration and specifications 4 Chapter 4 Functional Checks Contains functional checks that must be performed as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Diagrams Contains block diagrams and functional explanations of the LOGIQ 400 electronics 6 Chapter 6 Renewal Parts Contains a complete li...

Страница 16: ...e cleaning disinfection and sterilization refer to Probe section in LOGIQ 400 User Manual and Caution Sheet supplied with each probe NOTICE This medical equipment is approved in terms of the prevention of radio wave interference to be used in hospitals clinics and other institutions which are environmentally qualified The use of this equip ment in an inappropriate environment may cause some electr...

Страница 17: ...YSTEMS 2127661 INTRODUCTION 1 5 REV 8 1 2 1 Warnings continued OUTSIDE MARKINGS OF LOGIQ 400 For Color Monitor Models ILLUSTRATION 1 1 Note For further details regarding the cautions above refer to 2 2 10 MOVING INTO POSITION in Chap ter 2 ...

Страница 18: ...ngs continued OUTSIDE MARKINGS OF LOGIQ 400 For B W Monitor Models with S W V3 40 or earlier ILLUSTRATION 1 2 Note For further details regarding the cautions above refer to 2 2 10 MOVING INTO POSITION in Chap ter 2 Note B W system is applied the color monitor from software version 4 01y ...

Страница 19: ...Warnings continued CAUTION Possible Injury Placing objects on top of the monitor may cause the monitor to tilt with the falling objects resulting in injury to the operator Do not place any objects on the monitor OUTSIDE MARKINGS OF LOGIQ 400 For Color Monitor Models ILLUSTRATION 1 3 ...

Страница 20: ... 1 8 REV 8 1 2 1 Warnings continued OUTSIDE MARKINGS OF LOGIQ 400 For USA ILLUSTRATION 1 4 Note For the symbols shown in the illustration above refer to latter pages in this chapter The CAUTION label for the radio influence is attached on the console from April 1996 ...

Страница 21: ...inued OUTSIDE MARKINGS OF LOGIQ 400 For Europe ILLUSTRATION 1 5 Note For the symbols shown in the illustration above refer to latter pages in this chapter The CAUTION label for the radio influence is attached on the console from April 1996 The GOST label is attached on the console from June 1998 ...

Страница 22: ...Danish Greek and Turkish supplied with European Console EUROPEAN CAUTION LABELS FOR MAIN CAUTION LABEL ILLUSTRATION 1 6 Note The labels shown in ILLUSTRATION 1 6 are supplied with the consoles for Europe They shall be attached on the console over the existing labels as necessary Refer to the installation instructions supplied with the labels ...

Страница 23: ...IQ 400 For Japan ILLUSTRATION 1 7 Note For the symbols shown in the illustration above refer to latter pages in this chapter The CAUTION label for the radio influence is attached on the console from April 1996 The Japanese EMC label is attached on the console wiith the software version 4 10 or later instead of the VCIM label ...

Страница 24: ...ICE MANUAL GE MEDICAL SYSTEMS 2127661 INTRODUCTION 1 12 REV 8 1 2 1 Warnings continued OUTSIDE MARKINGS OF LOGIQ 400 For Korea ILLUSTRATION 1 8 Note ILLUSTRATION 1 8 shows the labels attached on the console for Korea ...

Страница 25: ...h ECG ILLUSTRATION 1 9 Note This label is attached only on the LOGIQ 400 console with the optional ECG unit Labels including English Russian Swedish Danish Greek and Turkish supplied with European Console EUROPEAN LANGUAGE LABEL ON ECG LABEL ILLUSTRATION 1 10 Note The labels shown in ILLUSTRATION 1 10 are supplied with the consoles for Europe They shall be attached on the console over the existing...

Страница 26: ...JURY OR DEATH OR THE POWER SUPPLY TO BE DAMAGED TURN OFF THE POWER AND CHECK THE RESIDUAL VOLTAGE OF CAPACITORS BEFORE ACCESSING THE POWER SUPPLY UNIT CAREFULLY WORK WHILE ACCESSING THE POWER SUPPLY UNIT OUTSIDE MARKINGS OF LOGIQ 400 ON POWER SUPPLY BOX ILLUSTRATION 1 11 Note Same labels are attached on both left and right outside of the power supply unit ...

Страница 27: ... protection suitable for direct cardiac contact Laboratory logo or labels denoting conformance with industry safety standards such as UL or IEC The system is not designed for use with flammable anesthetic gases CAUTION The equilateral triangle is usually used in combina tion with other symbols to advise or warn the user ATTENTION Consult accompa nying documents is intended to alert the user to ref...

Страница 28: ...power on position of the mains power switch ON Indicates the power on posi tion of the power switch CAUTION This Power Switch DOES NOT ISO LATE Mains Supply Protective Earth Indicates the protective earth grounding terminal Equipotentiality Indicates the ter minal to be used for connecting equi potential conductors when intercon necting grounding with other equip ment Non Ionizing Radiation indica...

Страница 29: ...xed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION 2 Type BF EQUIPMENT TYPE B EQUIPMENT with an F TYPE APPLIED PART TYPE B EQUIPMENT EQUIPMENT providing a particular degree of protection against electric shock par ticularly regarding allowable LEAKAGE CURRENT Normal mode Single failure mode Patient leakage c...

Страница 30: ...of the prod uct The product must be installed as stipulated in 1 3 2 Notice upon Installation of Product In case of issues related to EMC please follow procedures stated in 1 3 4 Countermeasures against EMC related Issues 1 3 2 Notice upon Installation of Product 1 Use either power supply cords provided by GEMS or ones designated by GEMS Products equipped with power source plug should be plugged i...

Страница 31: ...n any cover for some reason be sure to shut it before starting resuming operation Operating the system with any cover open may affect EMC performance 1 3 4 Countermeasures against EMC related Issues Generally it is very difficult to grapple with issues related to EMC It may take much time and cost General countermeasures Electromagnetic interference with other equipment 1 Electromagnetic interfere...

Страница 32: ...ervice Engineer as needed GE Medical Systems Ultrasound Business Group 4855 W Electric Ave Milwaukee WI 53219 USA TEL 1 800 437 1171 FAX 1 414 647 4090 CANADA TEL 1 800 668 0732 LATIN SOUTH AMERICA TEL 1 305 735 2304 GE Ultrasound Europe GE Ultraschall Deutschland GmbH Co KG Beethovenstr 239 42655 Solingen GERMANY TEL OLC Europe Toll Free Numbers or English German Hotline 49 212 2802 207 English G...

Страница 33: ...ed Ultrasound Room Layout 2 5 2 2 INSTALLATION 2 8 2 2 1 Introduction 2 8 2 2 2 Average Installation Time 2 8 2 2 3 Installation Warnings 2 8 2 2 4 Checking the Components 2 8 2 2 5 Unpacking LOGIQ 400 2 9 2 2 6 Probe Cable Arm Installation 2 12 2 2 7 MTZ Probe Holder Installation 2 13 2 2 8 Transducer Connection 2 13 2 2 9 Powering Up Procedure 2 14 2 2 10 Moving into Position 2 15 2 2 11 Adjusti...

Страница 34: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 INSTALLATION 2 2 REV 0 This page is left blank intentionally ...

Страница 35: ... before installation We recommend that you use an analyzer Dranetz Model 606 3 or Dranetz Model 626 PARAMETER LIMITS Voltage Range Japan 100 VAC 10 90 110 VAC Europe 220 240 VAC 10 198 264 VAC USA 120 VAC 10 108 132 VAC Power Japan MAX 1000 VA Europe MAX 1000 VA USA MAX 1000VA Line Frequency All applications 50 60Hz 2Hz Power Transients Less than 25 of nominal peak voltage for less than 1 millisec...

Страница 36: ...cal and electronic equipment may produce EMI unintentionally as the result of a malfunction These sources include medical lasers cauterizing guns computers monitors fans gel warmers microwave ovens and cellular phones The presence of broadcast station or van may also cause interference Carefully read the following precautions before installing the unit 1 Connect the power plug for any other equipm...

Страница 37: ...CUIT BREAKER RADIATION SHIELDING NONE REQUIRED for ULTRASOUND ENERGY TEMPERATURE 20 26 DEG C 68 79 DEG F for PATIENT COMFORT HUMIDITY 50 to 70 for PATIENT COMFORT HEAT DISSIPATION 2000 BTU Hr for LOGIQ 400 FLOOR LOADING Approximately 680 800 kg m2 without Accessories FLOOR CONDITION Gradient WITHIN 5 degrees LOGIQ 400 Weight 145 kg 320lbs without Accessories POWER SOURCE 220 240VAC 50Hz SINGLE PHA...

Страница 38: ...rol for overhead lights 4 Film viewer 5 Film and linen storage 6 Medical equipment storage 7 Hospital grade equipment electrical outlet 8 Analog telephone line for connection to InSite 9 Document storage area for operating and service manuals 10 Nearby waiting room dressing room lavatory 11 Trash bin Recommended and Alternate Ultrasound Console Floor Plans ILLUSTRATION 2 1 provides a recommended s...

Страница 39: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 INSTALLATION 2 7 REV 0 2 1 4 Recommended Ultrasound Room Layout Continued RECOMMENDED ULTRASOUND FLOOR PLAN ILLUSTRATION 2 1 ...

Страница 40: ...pproximately 320 lbs 145 kg without options preferably two people should unpack it Two people are also preferable for installing any additional bulky items 2 There are no operator serviceable components To prevent shock do not remove any covers or panels Should problems or malfunctions occur unplug the power cord Only qualified service personnel should carry out servic ing and troubleshooting Note...

Страница 41: ...ove the INNER SLEEVE 7 Remove the PLASTIC BAG 8 Lift the Monitor up by pressing the UP DOWN Release button located on the Monitor Arm 9 Remove the MONITOR SUPPORTER 10 Remove the adhesive tapes attached at the four corners of the TOP COVER 11 Put the TOP COVER on floor and prepare the slope to put the console down as shown in ILLUSTRATION 2 2 12 Unlock the brakes by stepping down on the brake pads...

Страница 42: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 INSTALLATION 2 10 REV 9 2 2 5 Unpacking LOGIQ 400 Continued or VINYL BAG UNPACKING LOGIQ 400 ILLUSTRATION 2 2 ...

Страница 43: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 INSTALLATION 2 11 REV 0 2 2 5 Unpacking LOGIQ 400 Continued LABELS ON PACKAGE ILLUSTRATION 2 3 ...

Страница 44: ...th the LOGIQ 400 console except to Japanese console This provides the proce dures for installing the probe cable arm onto console 1 Set the arm spacer to the axis of the probe cable arm Refer to ILLUSTRATION 2 4 2 Insert the probe cable arm into the hole onto console as shown in ILLUSTRATION 2 4 PROBE CABLE ARM INSTALLATION ILLUSTRATION 2 4 ...

Страница 45: ...TION ILLUSTRATION 2 5 2 2 8 Transducer Connection Connect a transducer to the upper transducer receptacle as follows 1 Ensure that the transducer twist lock lever points towards the 3 o clock position 2 Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface 3 Twist the transducer twist lock lever to the 7 o clock position to lock it in place Twis...

Страница 46: ...N POWER SWITCH ILLUSTRATION 2 6 3 Plug the main power cable to a hospital grade power receptacle with the proper rated voltage checked during preinstallation Note Never use a three to two prong adapter this defeats the safety ground 4 Set the circuit breaker to ON position upper side 5 Turn the main power switch ON LOGIQ 400 will start the power up sequence Note Follow the procedures below when un...

Страница 47: ... two people should move the machine when large humps grooves or grades will be encountered It is better to pull from the rear rather than push from the front of the unit Before moving store all loose parts in the unit Wrap transducers in soft cloth or foam to prevent damage Although LOGIQ 400 is a compact and mobile machine two people should move it over rough surfaces or up and down grades 2 2 11...

Страница 48: ...mer Information the Product Locator Installation Card Refer to ILLUSTRATION 2 7 Mail this Installation Card Product Locator to the address corresponding to your pole Note The Product Locator Installation Card shown in ILLUSTRATION 2 7 may not be same as the pro vided Product Locator card PRODUCT LOCATOR INSTALLATION CARD ILLUSTRATION 2 7 ...

Страница 49: ...TION 3 3 3 2 DIMENSIONS 3 3 3 3 ELECTRICAL SPECIFICATIONS 3 6 3 3 1 Power Supply 3 6 3 3 2 Facility Power Receptacle 3 6 3 4 STORAGE AND OPERATION REQUIREMENTS 3 6 3 5 OPTIONAL PERIPHERALS 3 7 3 5 1 Peripherals Accessories Connector Panel 3 7 3 5 2 List of Optional Peripherals 3 11 3 5 3 Power Consumption of Optional Peripherals 3 15 3 6 VIDEO SPECIFICATIONS 3 16 ...

Страница 50: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL SYSTEM CONFIGURATION 3 2 REV 0 This page is left blank intentionally ...

Страница 51: ...system configuration and specifications 3 2 DIMENSIONS Regarding LOGIQ 400 dimensions Refer to ILLUSTRATION 3 1 and 3 2 for planning the position of your LOGIQ 400 NOTE LENGTH mm inches ABERRATION 10 WEIGHT 145 kg 320 lbs OVERALL DIMENSIONS Color Monitor Models and B W Monitor Models with S W V4 01 or later ILLUSTRATION 3 1 ...

Страница 52: ... ABERRATION 10 WEIGHT 136 kg 300 lbs OVERALL DIMENSIONS B W Monitor Models with S W V3 40 or earlier ILLUSTRATION 3 2 Note The ILLUSTRATION 3 2 shows the dimensions for the B W console with the software version 3 40 or earlier For the dimensions of the B W console with the software version 4 01 or later refer to ILLUSTRATION 3 1 ...

Страница 53: ...STRATION 3 3 shows the dimensions for the LOGIQ 400 console with the software version 5 01y or later Note The tall type console is 75mm taller than the normal height console NOTE LENGTH mm inches ABERRATION 10 WEIGHT 145 kg 320 lbs OVERALL DIMENSIONS 15 INCH Monitor Models with S W V5 01 or later ILLUSTRATION 3 3 ...

Страница 54: ...ded The specific power receptacle used depends on your country s power line standards The receptacle should have International Electrotechnical Commission IEC approval or equivalent 3 4 STORAGE AND OPERATION REQUIREMENTS The LOGIQ 400 is shipped in a single container excluding PROBES Shipping weight is approximately 395 lbs 179kg B W monitor console 375lbs 170kg The size of the container is D1100 ...

Страница 55: ...r The serial ports for controlling recording devices video input output connectors audio input output connectors cam era expose connectors foot switch connector and power outlet connectors for peripherals are located on the rear panel This section indicates the pin assignment for each connector 1 6 in ILLUSTRATION 3 4 at pages 3 7 through 3 8 REAR CONNECTOR PANEL ILLUSTRATION 3 4 Note The B W cons...

Страница 56: ...ection is available on the console with the software version 3 40 or later and the optional DICOM Upgrade Set Note Each outer case ground line of peripheral accessory connectors are protectively grounded Signal ground lines are Not Isolated except the Service port 1 All of signal lines include signal GND of the Service port are isolated Note Output level of RS232C signals High 3V to 15V Low 15V to...

Страница 57: ...L SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL SYSTEM CONFIGURATION 3 9 REV 9 3 5 1 Peripherals Accessories Connector Panel Continued Note Output level of control signals indicated in the above tables are TTL level ...

Страница 58: ...L SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL SYSTEM CONFIGURATION 3 10 REV 9 3 5 1 Peripherals Accessories Connector Panel continued RN 12V Ethernet Note Output level of RS232C signals High 3V to 15V Low 15V to 0V ...

Страница 59: ...AL Color Video Printer SONY UP 1850MD NTSC SONY UP 1850EPM PAL SONY UP 2950MD NTSC SONY UP 2850P PAL SONY UP 2900MD NTSC SONY UP 2800P PAL Video Graphic Printer SONY UP890MD NTSC SONY UP890CE PAL SONY UP890MDG NTSC PAL Note If the customer need to connect the other peripherals with LOGIQ 400 console please contact the address shown in Chapter 1 of this manual Note The Quaf Frm format is available ...

Страница 60: ...recording devices with LOGIQ 400 console TABLE 3 3 LIST OF CONNECTING CABLES NAME PART NO FIGURE NOTE POWER CABLE P9509EE For recording device P9509LH Used only for installing B W Video Printer on front of console RS232C CABLE P9509MN For control signals 2150410 For Insite connection MINI PLUG CABLE P9509BE For control signals used only for B W Video Printer Ethernet Cable 2195662 For DICOM capabi...

Страница 61: ...202SF C386 NORYL CONVEX H45202CC No need I739 NORYL LINEAR H45202JG T739 NORYL INTRAOPERATIVE LINEAR H45202TG LA39 NORYL SUPERFICIAL LINEAR H45202LA B510 PU CARDIAC H45202BT PA51 S222 NORYL TRANSCRANIAL SECTOR H45202TC CWD2 NORYL CARDIAC CWD H45202DB No need CWD5 NORYL CARDICAC CWD H45202DE TYPE FAMILY PROBES PART No HEAD SHELL IS SAME AS CBF HEAD SHELL IS SAME AS CAE HEAD SHELL IS SAME AS MTZ HEA...

Страница 62: ...ers The number of upper row indicates the part number of probes for the region other than Japan The number of lower row indi cates the part number of probes for Japan Probes which have only one part number are not available in Japan Note 546L S317 S611 C386 I739 T739 and LA39 probes are available on LOGIQ 400 console from software version 3 00 Note B510 S222 CWD2 and CWD5 probes are available on L...

Страница 63: ...mption of each optional peripheral for LOGIQ 400 TABLE 3 6 POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES DEVICE MODEL POWER CONSUMPTION Video Cassette Recorder SVO 9500MD 64W SVO 9500MDP Color Video Printer UP 1850MD 140W UP 1850EPM UP 2950MD 220W UP 2850P at printing UP 2900MD 180W UP 2800P at printing B W Video Printer UP890MD 110W UP890CE UP890MDG 120W at printing ...

Страница 64: ...07 0 by 157 3 207 0 by 157 3 239 5 by 180 239 5 by 180 Total Horizontal Line Time s 63 56 64 00 31 78 32 00 Horizontal Display s 49 54 48 81 24 77 24 41 Front Porch Width s 2 76 3 09 1 39 1 54 Sync Pulse Width s 4 73 4 68 2 36 2 34 Back Porch Width s 6 53 7 42 3 26 3 71 Total Horizontal Blanking s 14 02 15 19 7 01 7 59 Vertical Blanking Interval H 31 50 38 50 31 50 38 50 Vertical Front Porch Width...

Страница 65: ...TLE PAGE 4 1 INTRODUCTION 4 3 4 1 1 Required Equipment 4 3 4 2 FUNCTIONAL CHECK PROCEDURES 4 4 4 2 1 Basic Controls 4 4 4 3 DIAGNOSTICS 4 10 4 3 1 Service Software Menu 4 10 4 3 2 Diagnosis Test Menu 4 11 4 3 3 Utility Menu 4 16 4 4 POWER SUPPLY ADJUSTMENTS 4 37 4 4 1 Power Supply Access 4 38 4 4 2 Power Supply Adjustment Procedure 4 40 ...

Страница 66: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 2 REV 0 This page is left blank intentionally ...

Страница 67: ...using the built in service software and power supply adjustments 4 1 1 Required Equipment To perform these tests you ll need any of a sector linear or a convex transducer Note The soft menu displayed on the CRT monitor screen disappears automatically by the time set with the CRT Menu Auto Erase Interval of the System Parameter page 1 6 When the Inf is selected the soft menu does not disappear auto...

Страница 68: ...key 7 Press key The B mode sub menu should appear on CRT monitor screen as follows 8 Select Dynamic Range section The Dynamic Range selection is displayed in reverse using or key video 9 Change Dynamic Range up Image grows softer and harder depending on position down using or key 10 Press Top Menu Select key The menus on CRT monitor screen should disappear twice 11 Rotate B M Gain knob Image grows...

Страница 69: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 5 REV 0 4 2 1 Basic Controls Continued B MODE DISPLAY SCREEN ILLUSTRATION 4 1 ...

Страница 70: ...th of B mode image should change wider or narrower 20 Press M D Cursor key The indicator LED located at the left side of the M Cursor key should be ON and the M mode cursor should become active again 21 Press Top Menu Select key The top menu should appear on the CRT monitor screen twice as follows 22 Select M section using or The M selection is displayed in reverse video key 23 Press key The M mod...

Страница 71: ...SPLAY SCREEN ILLUSTRATION 4 2 Note You can select the two types of display format Side by Side with the 2 D display to the left of the M mode display and Top Bottom with the 2 D display on top of the M mode display by using the Preset Menu For the Preset Menu refer to Customizing Your System in the LOGIQ 400 User Man ual ...

Страница 72: ...Freeze key The image should freeze 31 Rotate Cine Scroll knob The Cine gauge should appear under the B mode image 32 Press Freeze key again The image revives acquisition and the Cine gauge should disappear 33 Press CFM key again The B mode screen should appear as shown in ILLUSTRATION 4 1 CFM MODE DISPLAY SCREEN ILLUSTRATION 4 3 Note You can select the four types of CFM window Survey Survey Detail...

Страница 73: ...Rotate Gain knob The Doppler noise should increase decrease 37 Rotate Audio Volume knob The Doppler audio volume should increase decrease 38 Press B key The B mode screen should appear as shown in ILLUSTRATION 4 1 PULSE DOPPLER MODE DISPLAY SCREEN ILLUSTRATION 4 4 Note You can select the two types of display format Side by Side with the 2 D display to the left of the PD mode display and Top Bottom...

Страница 74: ...p Menu using the or key of Sub Menu Select keys on the keyboard The Set Up selec tion is displayed in reverse video as shown in ILLUSTRATION 4 6 3 Press key to select the Set Up Menu The Set Up Menu is displayed on the CRT monitor screen as shown in ILLUSTRATION 4 7 SET UP MENU ILLUSTRATION 4 7 4 Select the Diag section using the or key and press key to use the User Diagnostics function The Diagno...

Страница 75: ... the Diag section using the or key and press key to use the User Diagnostics function The Diagnosis Menu is displayed on the CRT monitor screen as shown in ILLUSTRATION 4 8 DIAGNOSIS TEST MENU ILLUSTRATION 4 8 Note Press and hold the Control key and press R to display the previous menu in the User Diagnostics mode ...

Страница 76: ...y when the Diagnosis Test Menu is displayed The system test will start automatically After checking each test item Passed or Failed is displayed depending on the actual test re sult SYSTEM TEST 1 ILLUSTRATION 4 9 The system beeps three times after checking all the tests If system does not pass the diagnostic test please contact the qualified service engineer in your area Please refer to Chapter 1 ...

Страница 77: ...e pattern permits the serviceman to accurately adjust the contrast of an optional equipment such as a camera to see the details in the lightest and darkest areas of the image Type 2 and then press Enter key when the Diagnosis Menu is displayed The Test Pattern Black White is dis played as shown in ILLUSTRATION 4 10 TEST PATTERN BLACK WHITE ILLUSTRATION 4 10 ...

Страница 78: ...t of an optional equipment such as a color video printer Type 3 and then press Enter key when the Diagnosis Menu is displayed The Test Pattern Color is displayed as shown in ILLUSTRATION 4 11 TEST PATTERN COLOR ILLUSTRATION 4 11 Note In case of B W monitor system this function is invalid The message Not tested is displayed on the CRT monitor screen when the option number 3 Test Pattern Color is se...

Страница 79: ...e vertical linearity and graphics perimeter pattern should be centered on the optional equipment display This ensures all whitescale graphics overlays will be visible Type 4 and then press Enter key when the Diagnosis Menu is displayed The Test Pattern Graphics is displayed as shown in ILLUSTRATION 4 12 TEST PATTERN GRAPHICS ILLUSTRATION 4 12 ...

Страница 80: ...U ILLUSTRATION 4 13 Note Press and hold the Control key and press R to display the previous screen in the User Utility mode Note Item 12 for V4 01 V4 02 or later and item 11 for V3 40 or V3 41 Network Error Log Display is the additional function for the console with the software version 3 40 or later This menu is displayed and available only when the Storage Print DICOM option is enabled Note Item...

Страница 81: ...ate and time previously set is displayed as shown in ILLUSTRATION 4 14 TIME ADJUSTMENT SCREEN ILLUSTRATION 4 14 Procedure 1 Input the time difference using the numeric keys on keyboard and then press Enter The input format for the time is HH MM You can change the current time within 12 hours Note Put a character in front of the time difference to decrease the current time After changing the curren...

Страница 82: ...E DISPLAY ILLUSTRATION 4 15 Procedure for Saving File 1 Type Y and then press Return key 2 Insert an 128MB MO Magneto Optical disk or 230MB disk for the conosle with software version 3 10 or later into the MOD drive of console and press any key Note Use new initialized MO disk It is impossible to save the error log file into the disk which another file is already stored Note The 230MB MO disk is a...

Страница 83: ...or by using the image recall function Note Press and hold the Control key and press W to save the trouble image into the hard disk You can save the maximum 8 images into the hard disk Note To save the trouble image into the hard disk the optional VIPB Assy must be installed on the console instead of VIDO Assy Otherwise the message This function is not available will be displayed on the CRT monitor...

Страница 84: ...uc ceeded or Failed is displayed on the CRT monitor Then the message Press Y to continue Control R to quit is displayed Type Y and press Return to continue to use this function Press and hold the Control key and press R to quit this function Procedure for Loading Trouble Images 1 Type 2 and then press Return key The message Set a media to drive and then press any key is displayed 2 Set an 128MB MO...

Страница 85: ...re for Displaying Trouble Images 1 Type 3 and then press Return key The Trouble Image File Menu will be displayed on the CRT monitor screen Refer to ILLUSTRATION 4 16 2 Input the file number you desire to display on the CRT monitor screen 3 Press Return key Press and hold the Control key and press Q to display the Trouble Image Menu on the CRT monitor screen ...

Страница 86: ...ology Data 5 Application Cardiology Data 6 Application Vascular Data 7 Application Ulorogy Data 8 Application Small Parts Data 9 All System Application Data System data includes System Preset Data File User Define Data File VCR Registration Information Data File VCR Patient Information Data File and Image Archive Search Data File Each application data includes Application Preset Data File Probe Pr...

Страница 87: ...enu into one MO disk Choose the item 09 All System Application Data Procedure for Saving User Data 1 Type 1 and then press the Return key The message Set a media to drive and then press any key is dis played 2 Insert an initialized 128MB MO Magneto Optical disk or 230MB disk for the conosle with software version 3 10 or later into the MOD drive of console and press the Return key The User Data Men...

Страница 88: ... drive and then press any key is dis played 2 Insert an initialized 128MB MO Magneto Optical disk or 230MB disk for the conosle with software version 3 10 or later into the MOD drive of console and press the Return key The User Data Menu is displayed on the CRT monitor screen Refer to ILLUSTRATION 4 18 3 Input the file number you desire to load from the MO disk 4 Press Return key The system starts...

Страница 89: ...ay is displayed as shown in ILLUSTRATION 4 19 USER OPTION DISPLAY ILLUSTRATION 4 19 Note The status of options indicated on the user option display is different depending on the system con figuration Note This function can only display the option names It is not possible for the user to change the option settings Note The following three options are not available on the LOGIQ 400CL systems Item 09...

Страница 90: ...ollowing four 4 options are available only for the console with the software version 3 40 ot later Item 29 3DvieW 1 Item 30 3DvieW 2 Item 31 Storage Print DICOM Item 32 Worklist DICOM Note Thefollowingthree 3 optionsareavailableonlyfortheconsolewiththesoftwareversion4 01otlater Item 33 ACE 2 Item 34 Realtime Doppler Calculation Item 35 MR3 Startup Display Note The ATO option Item 36 is not availab...

Страница 91: ...plays the version and capacity of memory only operator usable memory for each installed board Type 6 and then press Return key when the Utility Menu is displayed The Board Configuration Display is dis played as shown in ILLUSTRATION 4 20 BOARD CONFIGURATION DISPLAY ILLUSTRATION 4 20 Note The board configuration and version of each board indicated on the board configuration display is different dep...

Страница 92: ...ns of all system software installed in the console Type 7 and then press Return key when the Utility Menu is displayed The Software Configuration Display is displayed as shown in ILLUSTRATION 4 21 SOFTWARE CONFIGURATION DISPLAY ILLUSTRATION 4 21 Note The configuration indicated on the software configuration display is different depending on the soft ware version of console ...

Страница 93: ...ILLUSTRATION 4 22 SOFTWARE CONFIGURATION DISPLAY ILLUSTRATION 4 22 Procedure for Changing Each Settings 1 Input the item number you desire to change the settings The choices for the selected item are displayed 2 Type the appropriate number and press Return Note This function is valid only when the USrobotics modem is selected by the option setting function of the service software If the Motolora m...

Страница 94: ...k for the conosle with software version 3 10 or later and press the Return key The system automatically starts initializing the MO disk It takes approximately 15 minutes to complete initializing a disk Note The 230MB MO disk is available on the console with software version 3 10 or later During the system is initializing MO disk Initializing is displayed on the CRT monitor After initializing Suc c...

Страница 95: ... Menu Continued J SYSTEM ID ENTRY DISPLAY This displays the current system ID number and sets the new one Type 10 and then press the Return key when the Utility Menu is displayed The System ID Entry Display menu is displayed as shown in ILLUSTRATION 4 24 SYSTEM ID ENTRY DISPLAY ILLUSTRATION 4 24 ...

Страница 96: ...RT monitor screen as shown in ILLUSTRATION 4 24 2 Type any 5 digit number within the limits of 00000 to 16383 and press the Return key The new system ID number is set Procedure for Displaying Current System ID 1 Type 2 and then press the Return key The current system ID number is displayed on the CRT monitor screen as shown in ILLUSTRATION 4 24 2 Type Y and then press the Return key to continue th...

Страница 97: ...the settings of the Contrast and the Brightness on the CRT Monitor suitably Note This function is available only for the console with the system software version 4 01 or later Type 11 and then press Return key when the Utility Menu is displayed The Monitor Adjustment menu is dis played as shown in ILLUSTRATION 4 25 MONITOR ADJUSTMENT ILLUSTRATION 4 25 ...

Страница 98: ...ter of the Monitor Assy by pushing the top center of the door 3 Move Contrast and Brightness slide potentiometers rightwards to the maximum to increase both settings 4 Decrease brightness by moving Brightness slide potentiometer leftwards by degrees until B in the gray scale bar is not visible 5 Decrease contrast by moving Contrast slide potentiometer leftwards by degrees until A in the gray scale...

Страница 99: ...ystem software version 3 40 or later Type 11 12 on the console with the software V4 01 or later and then press Return key when the Utility Menu is displayed The Network Error Log Display menu is displayed as shown in ILLUSTRATION 4 27 NETWORK ERROR LOG DISPLAY ILLUSTRATION 4 27 Procedure for Saving File 1 Type Y and then press Return key 2 Insert a MO disk into the MOD drive and press any key Duri...

Страница 100: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 36 REV 6 This page is left blank intentionally ...

Страница 101: ...r both types of Power Supply Unit Before beginning the power supply adjustments procedure make sure the power outlet should provide the power line standards Refer to Chapter 2 Installation Note If the adjustment pot is over turned clockwise the power supply output is down to protect the circuits against the over voltage In that case power the LOGIQ 400 OFF and turn the pot counterclock wise all th...

Страница 102: ...S 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 38 REV 6 4 4 1 Power Supply Access 1 Unscrew four screws and remove the Side Cover Left as shown in the ILLUSTRATION 4 28 SIDE COVER LEFT REMOVAL ILLUSTRATION 4 28 ...

Страница 103: ...to the ILLUSTRATION 4 29 PS COVER LEFT REMOVAL ILLUSTRATION 4 29 3 Unscrew two screws and pull the LV Unit out Refer to the ILLUSTRATION 4 30 Note For the LV2 Unit and the LV3 Unit the points for measuring and adjusting are located at the front of the unit It is not necessary to pull the LV unit out of the Power Supply Unit2 and Power Supply Unit3 LV UNIT ILLUSTRATION 4 30 ...

Страница 104: ...Unit 3 respectively TABLE 4 1 POWER SUPPLY MEASUREMENT LOCATION POWER SUPPLY MEASURE AT RETURN AT 5V for Digital 5V 12V 5V for Analog 12V 5V Terminal on Power Supply See ILLUSTRATION 4 31 5V Terminal on Power Supply See ILLUSTRATION 4 31 5V Terminal on Power Supply See ILLUSTRATION 4 31 12V Terminal on Power Supply See ILLUSTRATION 4 31 12V Terminal on Power Supply See ILLUSTRATION 4 31 GND Termin...

Страница 105: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 41 REV 6 4 4 2 Power Supply Adjustment Procedure continued POWER SUPPLIES ON LV UNIT ILLUSTRATION 4 31 ...

Страница 106: ...Supply See ILLUSTRATION 4 32 5V Terminal on Power Supply See ILLUSTRATION 4 32 GND Terminal on Power Supply See ILLUSTRATION 4 32 GND Terminal on Power Supply See ILLUSTRATION 4 32 GND Terminal on Power Supply See ILLUSTRATION 4 32 GND Terminal on Power Supply See ILLUSTRATION 4 32 15V 15V Terminal on Power Supply See ILLUSTRATION 4 32 GND Terminal on Power Supply See ILLUSTRATION 4 32 TABLE 4 4 P...

Страница 107: ...pply Adjustment Procedure continued CAUTION The potentiometers 1 and 2 indicated in ILLUSTRATION 4 32 are used only for the sup plier adjustment Do not touch and rotate these potentiometers POWER SUPPLIES ON LV2 UNIT ILLUSTRATION 4 32 Note Use a non conductive trim pot to adjust the power supply ...

Страница 108: ...upply See ILLUSTRATION 4 33 GND Terminal on Power Supply See ILLUSTRATION 4 33 12V 12V Terminal on Power Supply See ILLUSTRATION 4 33 GND Terminal on Power Supply See ILLUSTRATION 4 33 5V 5V Terminal on Power Supply See ILLUSTRATION 4 33 GND Terminal on Power Supply See ILLUSTRATION 4 33 15V 15V Terminal on Power Supply See ILLUSTRATION 4 33 GND Terminal on Power Supply See ILLUSTRATION 4 33 TABLE...

Страница 109: ...5 REV 6 4 4 2 Power Supply Adjustment Procedure continued POWER SUPPLIES ON LV3 UNIT ILLUSTRATION 4 33 Note Check the 48V output as necessary Refer to ILLUSTRATION 4 33 for the measuring points Torelance 48V 2 4V Note Use a non conductive trim pot to adjust the power supply ...

Страница 110: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL FUNCTIONAL CHECKS 4 46 REV 6 This page is left blank intentionally ...

Страница 111: ... DIAGRAMS TABLE OF CONTENTS SECTION TITLE PAGE 5 1 INTRODUCTION 5 3 5 2 LOGIQ 400 SYSTEM 5 3 5 3 BLOCK DIAGRAM 5 4 5 4 WIRING DIAGRAM 5 8 5 5 POWER SUPPLY BLOCK DIAGRAM 5 12 5 6 POWER SUPPLY2 BLOCK DIAGRAM 5 14 5 7 POWER SUPPLY2 BLOCK DIAGRAM 5 16 5 8 CIRCUIT BOARD DESCRIPTION 5 18 ...

Страница 112: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...

Страница 113: ...console ease of use Digital circuitry enables more reliable and precise operation System Features The key design goals of this system are High Image Quality Increased User Productivity Multiple Clinical Applications Planned Upgradeability High Mobility Types of Applications The system supports many clinical uses Scan and display parameters may be user selected to default to desired values for each...

Страница 114: ...AGRAMS REV 6 5 3 BLOCK DIAGRAM LOGIQ 400 SYSTEM BLOCK DIAGRAM For Color Monitor Model and B W Model with S W V4 01 or Later ILLUSTRATION 5 1 a Note The DOPC Assy is supplied to the B W console with the softtware V4 01 or later with the optional Doppler Upgrade ...

Страница 115: ... continued NOTE The DICOM option is available for the console with software ver sion 3 40 or later NOTE The VIPP is installed on the con sole with software version 5 01 or later LOGIQ 400 SYSTEM BLOCK DIAGRAM For Color Monitor Model and B W Model with S W V4 01 or Later ILLUSTRATION 5 1 b ...

Страница 116: ...V 6 5 3 BLOCK DIAGRAM continued LOGIQ 400 SYSTEM BLOCK DIAGRAM For B W Monitor Model with S W V3 40 or Earlier ILLUSTRATION 5 2 a Note The B W console with the software version 4 01 or later has the same board configurations with the CFM console except to the DOPC Assy ...

Страница 117: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 7 DIAGRAMS REV 6 5 3 BLOCK DIAGRAM continued LOGIQ 400 SYSTEM BLOCK DIAGRAM For B W Monitor Model with S W V3 40 or Earlier ILLUSTRATION 5 2 b ...

Страница 118: ...alled in the console with the new 3 5 inch HDD Quantum Fireball 1280S and Quan tum Fireball EL because the HDD LED Assy is not supplied NOTE VIPP ASSY is installed in the console with software version 5 01y or later LOGIQ 400 SYSTEM WIRING DIAGRAM For Color Monitor Model and B W Model with S W V4 01 or Later ILLUSTRATION 5 3 a ...

Страница 119: ...DICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 9 DIAGRAMS REV 9 5 4 WIRING DIAGRAM continued LOGIQ 400 SYSTEM WIRING DIAGRAM For Color Monitor Model and B W Model with S W V4 01 or Later ILLUSTRATION 5 3 b ...

Страница 120: ...IAGRAM continued NOTE The CB46 and CB45N are not installed in the console with the new 3 5 inch HDD Quantum Fireball 1280S and Quantum Fireball EL because the HDD LED Assy is not supplied LOGIQ 400 SYSTEM WIRING DIAGRAM For B W Monitor Model with S W V3 40 or Earlier ILLUSTRATION 5 4 a ...

Страница 121: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 11 DIAGRAMS REV 6 5 4 WIRING DIAGRAM continued LOGIQ 400 SYSTEM WIRING DIAGRAM For B W Monitor Model with S W V3 40 or Earlier ILLUSTRATION 5 4 b ...

Страница 122: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 12 DIAGRAMS REV 0 5 5 POWER SUPPLY BLOCK DIAGRAM LOGIQ 400 SYSTEM POWER SUPPLY BLOCK DIAGRAM ILLUSTRATION 5 5 a ...

Страница 123: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 13 DIAGRAMS REV 0 5 5 POWER SUPPLY BLOCK DIAGRAM Continued LOGIQ 400 SYSTEM POWER SUPPLY BLOCK DIAGRAM ILLUSTRATION 5 5 b ...

Страница 124: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 14 DIAGRAMS REV 2 5 6 POWER SUPPLY2 BLOCK DIAGRAM LOGIQ 400 SYSTEM POWER SUPPLY2 BLOCK DIAGRAM ILLUSTRATION 5 6 a ...

Страница 125: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 15 DIAGRAMS REV 2 5 6 POWER SUPPLY2 BLOCK DIAGRAM Continued LOGIQ 400 SYSTEM POWER SUPPLY2 BLOCK DIAGRAM ILLUSTRATION 5 6 b ...

Страница 126: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 16 DIAGRAMS REV 6 5 7 POWER SUPPLY3 BLOCK DIAGRAM LOGIQ 400 SYSTEM POWER SUPPLY3 BLOCK DIAGRAM ILLUSTRATION 5 7 a ...

Страница 127: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 5 17 DIAGRAMS REV 6 5 7 POWER SUPPLY3 BLOCK DIAGRAM continued LOGIQ 400 SYSTEM POWER SUPPLY3 BLOCK DIAGRAM ILLUSTRATION 5 7 b ...

Страница 128: ...rocessor and Cine As sembly 1 1 TDCB ASSY Time Line Memory Processor Cine and Digi tal Scan Converter for B W Assembly 1 A10 DSCC ASSY Digital Scan Converter CFM Assembly 1 1 A11 VIPB ASSY Video Output Processor with VTR Playback Memory Assembly 1 VIPPASSY Video Assy with Interlace Progressive Scan Coonverter and VCR Playback 1 VIDM ASSY Video Output Processor for B W Assembly 1 A11 VIDM ASSY Vide...

Страница 129: ...SPR ASSY ASPR ASSY ASPR3 ASSY ASPR3 ASSY 2123603 used foe console with software version 2 11 or former 2123603 2 From console with software version 3 00 Old ASPR Assy 2123603 can be replaced with new ASPR Assy 2123603 2 2192676 From console with software version 4 01 or later 2192676 2 From console with software version 4 31 Old ASPR3 Assy 2192676 can be replaced with new ASPR3 Assy 2192676 2 4 8 ...

Страница 130: ... ASSY 2123337 3 0 4 5 or 6 8 9 or a TDCB ASSY TDCB ASSY TDCB ASSY TDCB ASSY 2123315 2123315 2 Old TDCB Assy 2123315 can be replaced with new TDCB Assy 2123315 2 2123315 3 Old TDCB Assy 2123315 and 2123315 2 can be replaced with new TDCB Assy 2123315 3 0 4 5 or 6 26 or 28 DSCC ASSY DSCC ASSY DSCC ASSY 2123313 2 2123313 3 Old DSCC Assy 2123313 2 can be replaced with new DSCC Assy 2123313 3 4 8 9 a o...

Страница 131: ... Assy PAL FRU No 151 6 90 6 2 13 Monitor Cover FRU No 152 6 92 6 2 14 Escutcheon Assy FRU No 153 6 93 6 2 15 Fixing Metal Plate FRU No 154 6 96 6 2 16 CRT Filter FRU No 155 CRT Filter Clamp Set FRU No 156 6 97 6 2 17 Monitor Bottom Assy FRU No 157 6 98 6 2 18 CRT Assy NTSC FRU No 158 CRT Assy PAL FRU No 159 6 100 6 2 19 CRT Cable Set FRU No 160 6 102 6 2 20 CRT Lamp FRU No 161 6 103 6 2 21 Monitor...

Страница 132: ...b Set FRU No 404 6 151 6 2 52 Trackball FRU No 405 6 152 6 2 53 Gain Encoder FRU No 410 6 153 6 2 54 Keyboard Panel Assy FRU No 450 6 155 6 2 55 Keyboard Cover Assy FRU No 451 SW PWB FRU No 462 Rubber Key FRU No 452 453 454 and 455 6 156 6 2 56 Keyboard Knob Set FRU No 464 6 159 6 2 57 Trackball FRU No 457 6 160 6 2 58 TGC Assy FRU No 458 6 161 6 2 59 Freeze Key Assy FRU No 459 6 162 6 2 60 Rotary...

Страница 133: ...Nest Mother Assy FRU No 615 6 208 6 2 84 Mother IC FRU No 616 6 210 6 2 85 High Voltage Assy FRU No 701 6 214 6 2 86 HV Fan FRU No 702 6 216 6 2 87 Low Voltage Unit FRU No 703 6 218 6 2 88 Power Control Unit FRU No 704 6 220 6 2 89 SSR Unit FRU No 705 6 222 6 2 90 Power Inlet Unit FRU No 706 for 115V FRU No 707 for 220V 6 224 6 2 91 PW Air Filter FRU No 708 6 226 6 2 92 Filter Cover Set FRU No 709...

Страница 134: ...3 6 4 4 Installing Software 6 254 6 4 5 Final Procedures 6 255 6 5 POWER SUPPLY REPLACEMENT 6 257 6 5 1 Introduction 6 257 6 5 2 Replacement Procedures 6 257 6 6 REPLACING 2 5 INCH HDD ASSY WITH 3 5 INCH HDD ASSY 6 263 6 6 1 Introduction 6 263 6 6 2 Time Required 6 263 6 6 3 Parts Required 6 263 6 6 4 Procedures 6 263 6 7 REPLACING LV2 UNIT WITH LV3 UNIT 6 276 6 7 1 Introduction 6 276 6 7 2 Time R...

Страница 135: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 5 REV 9 This page is left blank intentionally ...

Страница 136: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 6 REV 9 LIST OF ABBREVIATIONS 1 Stock item 2 Non stock item ...

Страница 137: ...V NTSC VER 2 11 OPERATOR CONSOLE ASSY 2127803 1 B W 220V PAL VER 2 11 PROBE CABLE ARM P9514FT 1 1 1 1 1 1 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 1 1 1 SWIVEL LOCK U0144AE 1 1 CRT FILTER P9522WD 1 1 VINYL COVER ASSY 2131324 1 B W VINYL COVER ASSY 2131427 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 1 AQUASON...

Страница 138: ...R ASSY P9522XM 1 1 1 1 1 1 1 1 SWIVEL LOCK U0144AE 1 1 CRT FILTER P9522WD 1 1 VINYL COVER ASSY 2131324 1 B W VINYL COVER ASSY 2131427 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 1 1 1 OPERATION MANUAL 2156967 100 1 1 1 1 English OPERATION MANUAL 2156967 140 ...

Страница 139: ... 1 1 CRT FILTER P9522WD 1 1 VINYL COVER ASSY 2131324 1 B W VINYL COVER ASSY 2131427 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 1 1 1 OPERATION MANUAL 2177612 100 1 1 1 1 English OPERATION MANUAL 2177612 140 1 1 Japanese QUICK GUIDE 2137070 100 1 1 English Q...

Страница 140: ... LOGIQ 400CL Software Version 3 20 PROBE CABLE ARM P9514FT 1 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 VINYL COVER ASSY 2131324 1 128MB MO DISK 2131693 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 OPERATION MANUAL 2202294 100 1 1 English OPERATION MANUAL 2202294 140 1 Japanese SERVICE MANU...

Страница 141: ... 1 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 CRT FILTER P9522WD 1 1 1 VINYL COVER ASSY 2131324 1 1 B W VINYL COVER ASSY 2131427 1 1 V3 40Am PRESET MO 2225754 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 1 1 1 1 1 1 3 PROBE PORT OPTION 2177136 1 1 MV...

Страница 142: ...TATION CALC S W 2165098 1 1 1 1 1 1 1 1 1 ADVANCED VASCULAR 2165099 1 1 1 1 1 1 1 CALCULATION S W START UP MD DISPLAY 2165100 1 1 1 1 1 1 1 1 1 1 1 1 FETAL TREND S W 2136313 1 1 IMAGE ARCHIVE S W 2136312 1 CFM CAPTURE S W 2136314 1 Power Doppler Imaging S W 2136317 1 1 1 Advanced Cardiac Calc S W 2136315 1 PART NAME PART NO QUANTITY DESCRIPTION USA RA1412LB 2212967 2 EUROPE RA1413LE 2212967 3 JAPA...

Страница 143: ...OOT SWITCH P9509NB 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 OPERATION MANUAL 2209831 100 1 1 English OPERATION MANUAL 2209831 140 1 1 Japanese QUICK GUIDE 2211144 140 1 1 Japanese SERVICE MANUAL 2127661 1 1 1 1 COLOR ENHANCEMENT S W 2165097 1 1 ADVANCED VASCULAR S W 2165099 1 1 ...

Страница 144: ... 1 ARM SPACER P9510HK 1 1 1 1 1 1 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 CRT FILTER P9522WD 1 1 VINYL COVER ASSY 2131324 1 1 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 1 1 1 1 1 1 3 PROBE PORT OPTION ...

Страница 145: ... 1 1 1 1 1 ADVANCED VASCULAR 2165099 1 1 1 1 1 1 CALCULATION S W START UP MD DISPLAY 2165100 1 1 1 1 1 1 1 1 1 1 1 1 1 FETAL TREND S W 2136313 1 IMAGE ARCHIVE S W 2136312 1 CFM CAPTURE S W 2136314 1 Power Doppler Imaging S W 2136317 1 1 Real Time Doppler Calc S W 2241560 1 1 1 1 ACE 2 S W 2241561 1 1 1 1 1 1 1 1 1 1 1 1 1 PART NAME PART NO QUANTITY DESCRIPTION USA RA1413LJ 2239787 2 USA RA1414LF 2...

Страница 146: ...9NB 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 OPERATION MANUAL 2230620 100 1 1 English OPERATION MANUAL 2241837 140 1 1 Japanese QUICK GUIDE 2211144 140 1 1 Japanese SERVICE MANUAL 2127661 1 1 1 1 COLOR ENHANCEMENT S W 2165097 1 1 ADVANCED VASCULAR S W 2165099 1 1 1 1 1 BASIC VAS...

Страница 147: ...9510HK 1 1 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 1 1 CRT FILTER P9522WD 1 1 1 1 VINYL COVER ASSY 2131324 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 1 1 3 PROBE PORT OPTION 2177136 1 1 1 OPERATION MANUAL 2250488 100 1 1 1 1 Engl...

Страница 148: ... 1 1 ARM SPACER P9510HK 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 VINYL COVER ASSY 2131324 1 1 128MB MO DISK 2131693 1 1 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 OPERATION MANUAL 2250488 100 1 1 English OPERATION MANUAL 2250488 140 1 1 Japanese QUICK START GUIDE 2237879 ...

Страница 149: ...400CL Software Version 4 32 PROBE CABLE ARM P9514FT 1 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 VINYL COVER ASSY 2131324 1 1 128MB MO DISK 2131693 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 OPERATION MANUAL 2251967 140 1 1 Japanese QUICK GUIDE 2211144 140 1 1 Japanese SERVICE MANUAL 212766...

Страница 150: ... CABLE ARM P9514FT 1 1 1 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 1 3 PROBE PORT OPTION 2177136 1 1 1 1 128MB MO DISK 2131693 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 ADD CINE MEMORY P9523UC 1 1 1 BASIC OP MANUAL 2260260 100 1 1 1 ADVANCED O...

Страница 151: ...C VER 5 01 PROBE CABLE ARM P9514FT 1 1 1 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 1 1 FOOT SWITCH P9509NB 1 1 1 3 PROBE PORT OPTION 2177136 1 1 1 1 128MB MO DISK 2131693 1 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 1 1 1 BASIC OP MANUAL 2260260 100 1 1 1 ADVANCED OP MANUAL 22602...

Страница 152: ...are Version 5 01 PROBE CABLE ARM P9514FT 1 1 1 1 ARM SPACER P9510HK 1 1 1 1 MTZ HOLDER ASSY P9522XM 1 1 1 1 FOOT SWITCH P9509NB 1 1 1 128MB MO DISK 2131693 1 1 1 1 3 5 INCH MO CLEANING KIT U0055ZA 1 1 1 1 3 5 INCH MO HEAD CLEANER 2148392 1 1 1 1 AQUASONIC GEL U0403BD 1 1 1 1 BASIC OP MANUAL 2260260 100 1 1 1 1 ADVANCED OP MANUAL 2260261 100 1 1 1 1 English QUICK START GUIDE 2260274 100 1 1 1 1 Eng...

Страница 153: ... 6 5 REAR RIGHT SIDE ACCESS FRU501 509 See ILLUSTRATION 6 8 KEYBOARD BLOCK FRU400 410 See ILLUSTRATION 6 7 FRONT UPPER LEFT SIDE ACCESS FRU511 580 See ILLUSTRATION 6 9 P C BOARDS FRU601 618 See ILLUSTRATION 6 10 COVERS FRU300 420B See ILLUSTRATION 6 6 POWER SUPPLY FRU701 512 See ILLUSTRATION 6 11 POWER SUPPLY 2 FRU750 562 See ILLUSTRATION 6 12 15 INCH MONITOR FRU170 182 See ILLUSTRATION 6 4 LOGIQ ...

Страница 154: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 24 REV 9 OPERATOR CONSOLE ASSY 1 12 EXPLODED VIEW a COLOR MONITOR ILLUSTRATION 6 2 ...

Страница 155: ...223 1 1 6 2 1 101 MONITOR COVER SET 2128722 1 2 6 2 2 MONITOR COVER SET 2162082 2 1 2 Available for all console 6 2 2 MONITOR COVER SET 2203089 1 2 6 2 2 102 ESCUTCHEON ASSY 2126888 1 2 6 2 2 ESCUTCHEON ASSY 2162083 1 2 6 2 2 ESCUTCHEON ASSY 2200222 1 2 6 2 2 FRU NO FRU Only for console with soft ware V2 11 or former excluding FRU109 110 Only for console with soft ware V2 11 or former Including FR...

Страница 156: ... 2 2 105 MONITOR COVER R ASSY 2168466 1 2 including caution label 6 2 2 106 MONITOR COVER TOP P9523NH 1 2 including caution label 6 2 2 107 ESCUTCHEON FRONT DOOR P9524JJ 1 1 6 2 4 ESCUTCHEON FRONT DOOR 2193942 1 1 6 2 4 108 CRT CAP SET P9523NB 1 2 2 Knobs and Lamp Switch Keytop 6 2 5 109 CRT FILTER P9522WD 1 2 includes FRU110 6 2 6 110 CRT FILTER CLAMP SET P9523NC 1 2 6 2 6 111 ACCESSORY ASSY 2114...

Страница 157: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 27 REV 9 This page is left blank intentionally ...

Страница 158: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 28 REV 9 OPERATOR CONSOLE ASSY 2 12 EXPLODED VIEW b B W MONITOR ILLUSTRATION 6 3 ...

Страница 159: ...nsole 6 2 12 151 MONITOR ASSY PAL 2140516 2 1 1 6 2 12 MONITOR ASSY PAL 2164506 1 1 Available for all B W console 6 2 12 152 MONITOR COVER 2140517 1 2 6 2 13 153 ESCUTCHEON ASSY 2140518 2 1 2 6 2 14 ESCUTCHEON ASSY 2164507 1 2 6 2 14 154 FIXING METAL PLATE 2144045 1 1 6 2 15 155 CRT FILTER P9514BY 1 2 6 2 16 156 CRT FILTER CLAMP SET 2140521 2 1 2 6 2 16 157 MONITOR BOTTOM ASSY 2126884 1 2 6 2 17 F...

Страница 160: ...58 2250267 V4 31 2197064 2198442 2198445 LOGIQ 400CL V3 20 2213911 2213915 LOGIQ 400CL V3 41 2239832 2239836 LOGIQ 400CL V4 02 2250270 2250271 2255918 2255919 LOGIQ 400CL V4 32 MATERIAL LIST 2 12 continued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE B W MONITOR BLOCK CONTINUED 158 CRT ASSY NTSC 2143774 1 2 6 2 18 159 CRT ASSY PAL 2143775 1 2 6 2 18 160 CRT CABLE SET 2135183 1 1 CB81B 6...

Страница 161: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 31 REV 9 This page is left blank intentionally ...

Страница 162: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 32 REV 9 OPERATOR CONSOLE ASSY 3 12 EXPLODED VIEW c 15 INCH MONITOR ILLUSTRATION 6 4 ...

Страница 163: ...Y 2267945 1 2 173 MONITOR COVER L2 ASSY 2250407 1 2 including caution label 6 2 22 174 MONITOR COVER R2 ASSY 2250409 1 2 including caution label 6 2 22 175 TRAY2 ASSY 2250410 1 2 including caution label 6 2 22 178 SPEAKER ASSY 2250413 1 1 6 2 23 179 LAMP SHADE2 2246213 1 2 6 2 24 180 TASK LAMP U0107EH 1 1 6 2 24 181 CRT ASSY 2251431 1 2 including tilt mechanism 6 2 25 182 CRT CABLE SET 2251953 1 1...

Страница 164: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 34 REV 9 OPERATOR CONSOLE ASSY 4 12 EXPLODED VIEW c NECK BLOCK ILLUSTRATION 6 5 ...

Страница 165: ...L V3 41 2239832 2239836 LOGIQ 400CL V4 02 2250270 2250271 2255918 2255919 LOGIQ 400CL V4 32 2265944 2265947 LOGIQ 400CL V5 01 MATERIAL LIST 4 12 SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE NECK BLOCK 201 NECK ASSY 2127542 1 1 for Color monitor 6 2 26 NECK ASSY 2126885 1 1 for B W monitor 6 2 26 NECK ASSY 2265686 1 1 for 15 Inch monitor 6 2 21 203 NECK GRIP P9524HJ 1 2 excluding screw 6...

Страница 166: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 36 REV 9 OPERATOR CONSOLE ASSY 5 12 EXPLODED VIEW d COVERS ILLUSTRATION 6 6 ...

Страница 167: ... 2 6 2 32 SIDE COVER R 2261595 1 2 From V5 System 6 2 32 303 REAR COVER 2115618 1 2 excluding Rear Door Assy 6 2 33 REAR COVER ASSY 2265557 1 2 From V5 System 6 2 33 304 REAR DOOR ASSY 2126886 1 2 6 2 34 305 FRONT BASE COVER P9523PC 1 2 6 2 35 306 FRONT COVER P9524CE 1 2 6 2 36 FRONT COVER ASSY 2211257 1 2 6 2 36 FRONT COVER ASSY 2253599 1 2 From V5 System 6 2 36 307 KEYBOARD BOTTOM COVER P9524CN ...

Страница 168: ...0CL V5 01 MATERIAL LIST 5 12 Continued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE COVERS CONTINUED 315 HANDLE 2115622 1 2 6 2 43 316 AIR FILTER P9524BJ 1 2 6 2 44 AIR FILTER 2259546 1 2 6 2 44 317 CABLE ARM P9514FT 1 2 6 2 45 318 FRONT CASTER ASSY P9523PD 1 2 6 2 46 319 REAR CASTER ASSY P9523PE 1 2 6 2 47 320 CASTER COVER ASSY P9523NE 1 2 covers for 1 caster 6 2 48 318B FRONT CASTER R...

Страница 169: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 39 REV 9 This page is left blank intentionally ...

Страница 170: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 40 REV 9 OPERATOR CONSOLE ASSY 6 12 EXPLODED VIEW e KEYBOARD BLOCK ILLUSTRATION 6 7 ...

Страница 171: ...YBOARD ASSY 2200224 1 1 For LOGIQ 400CL 6 2 50 KEYBOARD2 ASSY 2206006 1 1 6 2 50 402 KEYBOARD COVER 2126999 1 2 6 2 50 403 KEY SHEET 2144759 1 2 With metal plate 6 2 50 KEY SHEET 2200221 1 2 For LOGIQ 400CL 6 2 50 404 KEYBOARD KNOB SET 2126998 1 2 6 Caps and 8 TGC knobs 6 2 51 405 TRACKBALL 2141741 1 1 6 2 52 410 GAIN ENCODER 2164871 1 2 6 2 53 GAIN ENCODER 2216951 1 2 6 2 53 GAIN ENCODER 2216954 ...

Страница 172: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 42 REV 9 OPERATOR CONSOLE ASSY 7 12 EXPLODED VIEW e KEYBOARD BLOCK ILLUSTRATION 6 8 ...

Страница 173: ...922 1 2 6 2 55 KEYBOARD COVER 2 CL ASSY 2268955 1 2 For LOGIQ 400CL 452 RUBBER KEY AN_E 2268923 1 2 For English 6 2 55 RUBBER KEY AN_R 2269083 1 2 For Russian RUBBER KEY AN_G 2269084 1 2 For Greek RUBBER KEY AN_T 2269085 1 2 For Turkish 453 RUBBER KEY TOP MENU 2268926 1 2 6 2 55 454 RUBBER KEY MODE 2268927 1 2 6 2 55 455 RUBBER KEY TRACKBALL 2268928 1 2 6 2 55 457 TRACKBALL ASSY 2235387 1 1 6 2 57...

Страница 174: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 44 REV 9 OPERATOR CONSOLE ASSY 8 12 EXPLODED VIEW f REAR RIGHT SIDE ACCESS ILLUSTRATION 6 9 ...

Страница 175: ...SSY 2138820 1 1 For 120V B W console 6 2 61 REAR CONN PANEL ASSY 2138821 1 1 For 220V B W console REAR CONN PANEL2 ASSY 2115619 2 1 1 For 100V CFM console REAR CONN PANEL2 ASSY 2115620 2 1 1 For 120V CFM console REAR CONN PANEL2 ASSY 2115621 2 1 1 For 220V CFM console 502 CNTIF ASSY 2123327 2 1 2 6 2 62 CNTIF2 ASSY 2212958 1 2 6 2 62 503 AVIF ASSY 2123329 1 2 For CFM console 6 2 63 BVIF ASSY 21236...

Страница 176: ...2265947 LOGIQ 400CL V5 01 MATERIAL LIST 8 12 continued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE 506 ECG BOARD ASSY P9522WZ 1 1 6 2 65 ECG BOARD2 ASSY 2207806 1 1 6 2 65 ECG BOARD2 ASSY 2207806 2 1 1 From V5 System 6 2 65 507 ECG PANEL ASSY P9522XB 1 1 6 2 66 ECG CONN PNL ASSY 2251836 1 1 From V5 System 6 2 66 508 ECG CABLE ASSY P9509KG 1 2 ECG CABLE P9509KC 1 2 System side ECG CABLE...

Страница 177: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 47 REV 9 This page is left blank intentionally ...

Страница 178: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 48 REV 9 OPERATOR CONSOLE ASSY 8 12 EXPLODED VIEW g FRONT UPPER LEFT SIDE ACCESS ILLUSTRATION 6 10 ...

Страница 179: ... 68 PROBE CONN SET ASSY 2265572 1 1 From V5 System 6 2 68 514 CONNECTOR COVER U0477FD 1 2 6 2 69 515 SHIELD FINGER LONG P9524DW 1 1 6 2 70 516 SHIELD FINGER SHORT P9524DX 1 1 6 2 70 517 PROBE CONN 1 ASSY 2231575 1 2 For 3 slots model 6 2 71 518 DCON ASSY 2123321 1 2 For 2 slots model 6 2 71 519 PRAG ASSY 2123323 1 2 6 2 72 520 MODD U0052ZA 1 1 6 2 73 MODD 2187012 1 1 Refer to Note below 6 2 73 MOD...

Страница 180: ...85221 4 1 1 6 2 75 3 5 INCH V3 30By INIT ED HDD KIT 2208953 4 1 1 6 2 75 3 5 INCH V3 40Dy INIT ED HDD KIT 2219189 6 1 1 6 2 75 3 5 INCH V3 41Dy INIT ED HDD KIT 2219190 6 1 1 6 2 75 3 5 INCH FRU NO FRU 3 5 inch HDD which V R 2 11y software is installed 3 5 inch HDD which V R 2 50y software is installed 3 5 inch HDD which V R 3 00 Cy software is installed 3 5 inch HDD which V R 3 10Cy software is in...

Страница 181: ...ntinued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE FRONT UPPER LEFT BLOCK CONTINUED 523 V4 01Cy INIT ED HDD KIT 2239423 4 1 1 6 2 75 3 5 INCH V4 02Cy INIT ED HDD KIT 2239426 4 1 1 6 2 75 3 5 INCH V4 31By INIT ED HDD KIT 2248939 2 1 1 6 2 75 3 5 INCH V4 32By INIT ED HDD KIT 2248942 2 1 1 6 2 75 3 5 INCH V5 01y INT ED HDD KIT 2264062 1 1 6 2 75 FRU NO FRU 3 5 inch HDD which V R 4 02Cy s...

Страница 182: ...POWER SWITCH ASSY P9522WL 1 1 6 2 77 POWER SWITCH ASSY 2251835 1 1 From V5 System 6 2 77 528 FLAT CABLE P9522ME 1 1 CB21 529 FLAT CABLE 2126890 1 1 CB10A 530 FLAT CABLE 2126891 1 1 CB23A 531 FLAT CABLE 2126892 1 1 CB24A 532 FLAT CABLE 2126893 1 1 CB29A 533 FLAT CABLE 2126894 1 1 CB30A 534 CB6M CABLE 2122996 1 2 MODD MOTHER 535 FLAT CABLE 2145889 1 1 CB24N for 3 5 inch HDD 536 HDD HOLDER2 ASSY 2145...

Страница 183: ... SERVICE MANUAL RENEWAL PARTS 6 53 REV 9 OPERATOR CONSOLE ASSY 9 12 609 VIDO2 Assy 610 VIPB Assy VIPB2 Assy VIPP Assy 611 VIDM Assy 603 ASPR Assy ASPR3 Assy 605 DOPC Assy DOPC2 Assy DOPC3 Assy EXPLODED VIEW h P C BOARDS ILLUSTRATION 6 11 ...

Страница 184: ...EFERENCE P C BOARDS 601 TRDR ASSY 2123305 1 1 6 2 78 TRDR ASSY 2123305 2 1 1 6 2 78 602 HBFR ASSY 2123307 1 1 6 2 78 603 ASPR ASSY 2123603 1 1 6 2 78 ASPR ASSY 2123603 2 1 1 6 2 78 ASPR3 ASSY 2192676 1 1 6 2 78 ASPR3 ASSY 2192676 2 1 1 6 2 78 FRU NO FRU This is one of 2 HBFR as semblies installed in a con sole Fromsoftwareversion3 00 OldTRDR 2123305 canbe replacedwithnewTRDR 2123305 2 Fromsoftware...

Страница 185: ...8 605 DOPC ASSY 2123311 2 1 1 only for color monitor model 6 2 78 DOPC ASSY 2123311 3 1 1 6 2 78 DOPC2 ASSY 2192564 1 1 6 2 78 DOPC3 ASSY 2248949 1 1 6 2 78 606 TLMC ASSY 2158390 2 1 1 6 2 78 607 TDCB ASSY 2123315 3 1 1 6 2 78 FRU NO FRU From Software version 4 01 for MD and 4 02 for CL Fromsoftwareversion3 00 OldDOPC 2123311 2 can bereplacedwithnewDOPC 2123311 3 For color monitor model excluding ...

Страница 186: ...32 3 1 1 For MVME147 010 6 2 78 610 VIPB ASSY P9525VA 1 1 6 2 78 VIPB ASSY 2149141 1 1 For MVME147 010 6 2 78 VIPB2 ASSY 2212616 1 1 6 2 78 VIPB2 ASSY 2212616 2 1 1 6 2 78 VIPP ASSY 2245652 1 1 6 2 78 FRU NO FRU For color monitor model without VTR playback DSCCASSYisusedonlyfor color monitor model OldDSCC 2123313 2 can bereplacedwithnewDSCC 2123313 3 For color monitor model with VTR playback avail...

Страница 187: ...ontinued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE 611 VIDM ASSY 2123337 1 1 For MVME147 SRF 6 2 78 VIDM ASSY 2123337 2 1 1 For MVME147 010 6 2 78 VIDM ASSY 2123337 3 1 1 6 2 78 612 MPU ASSY 2123339 4 1 1 6 2 77 MPU A ASSY 2195746 3 1 1 6 2 77 FRU NO FRU For MVME147 010 Old VIDM 2123337 2 can be replaced with new VIDM 2123337 3 MASC Assy ADPT Assy OBC Assy Old MPU Assy 2123339 212333...

Страница 188: ...5627 2 1 1 6 2 83 616 MOTHER IC P9522PK 1 1 6 2 84 FRU NO FRU Optional expansion Cine memory From software V3 10 console This Assy has backward compatibility with old Assy 2115627 Note There are two types of OBC One Board Computer ASSY included in MPU ASSY MVME147SRF and MVME147 010 Those two assemblies have functional compatibility However the type of time keeper RAM mounted on each board is diff...

Страница 189: ...5 LOGIQ 400CL V3 20 2213911 2213915 LOGIQ 400CL V3 41 2239832 2239836 LOGIQ 400CL V4 02 2250270 2250271 2255918 2255919 LOGIQ 400CL V4 32 2265944 2265947 LOGIQ 400CL V5 01 MATERIAL LIST 9 12 continued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE 617 STCG ASSY 2123325 1 1 STCG ASSY 2123325 2 1 1 618 TDCWD ASSY 2158393 1 1 FRU NO FRU installed on PRAG Assy supplied with optional Stereable...

Страница 190: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 60 REV 9 OPERATOR CONSOLE ASSY 10 12 EXPLODED VIEW i POWER SUPPLY UNIT ILLUSTRATION 6 12 ...

Страница 191: ...GIQ 400CL V4 32 2265944 2265947 LOGIQ 400CL V5 01 MATERIAL LIST 10 12 SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE POWER SUPPLY BLOCK 701 HIGH VOLTAGE ASSY 2126983 2 1 1 6 2 85 702 HV FAN 2126990 1 1 6 2 86 703 LOW VOLTAGE UNIT 2126989 1 1 6 2 87 704 POWER CONTROL UNIT 2126984 3 1 1 6 2 88 705 SSR UNIT 2126985 1 1 6 2 89 706 POWER INLET UNIT 115 100V 2126987 1 1 6 2 90 707 POWER INLET U...

Страница 192: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 62 REV 9 OPERATOR CONSOLE ASSY 12 12 EXPLODED VIEW j POWER SUPPLY UNIT2 ILLUSTRATION 6 13 ...

Страница 193: ...FOR REFERENCE POWER SUPPLY BLOCK 2 750 HV UNIT 2144488 2 1 1 6 2 93 751 COOLING FAN 2158762 1 1 6 2 94 752 LV2 UNIT 2144482 6 1 1 6 2 95 753 TRIAC UNIT 2144487 1 1 6 2 96 754 POWER INLET UNIT2 115 100V 2144485 1 1 6 2 97 755 POWER INLET UNIT2 220V 2144484 1 1 6 2 97 756 POWER TRANSFORMER 2 2144483 1 2 757 115V POWER SUPPLY UNIT2 2144481 1 2 6 5 758 220V POWER SUPPLY UNIT2 2144480 1 2 6 5 761 PW AI...

Страница 194: ...REW SET P9523NW 1 1 901 PROBE ADAPTER TYPE PA51 2136306 1 1 For Japan PROBE ADAPTER TYPE PA51 2136307 1 1 For USA Europe Asia 902 PROBE ADAPTER TYPE 5S P9602AN 1 1 For Japan PROBE ADAPTER TYPE 5S P9602AT 1 1 For USA Europe Asia 904 V2 11 SOFTWARE MO DISK SET 2136320 4 1 2 V2 11 y 913 V2 50 SOFTWARE MO DISK SET 2171909 2 1 2 V2 50 y 905 V3 00 SOFTWARE MO DISK SET 2164543 4 1 2 V3 00 Cy 906 V3 10 SO...

Страница 195: ...AL LIST 12 12 continued SECTION PART NAME PART NO QTY DESCRIPTION FOR REFERENCE OTHERS Continued 915 V4 31 SYSTEM MO DISK 2248937 2 1 2 V4 31B y 916 V4 32 SYSTEM MO DISK 2248940 2 1 2 917 V5 01 SYSTEM MO DISK 2264060 1 2 V5 01y 910 E721 PROBE HANDLE SET 2132282 1 1 911 PROBE CONNECTOR 2133634 1 1 KNOB SET 950 CBF PROBE P9603AD 1 1 For USA Europe Asia CBF PROBE P9603AA 1 1 For Japan 951 CAE PROBE P...

Страница 196: ...ia LH PROBE P9601AC 1 1 For Japan 955 W PROBE P9600BH 1 1 For USA Europe Asia W PROBE P9600BD 1 1 For Japan 956 C364 PROBE P9607AB 1 1 For USA Europe Asia C364 PROBE P9607AA 1 1 For Japan 957 C551 PROBE P9607AD 1 1 For USA Europe Asia C551 PROBE P9607AC 1 1 For Japan 958 E721 PROBE P9607AF 1 1 For USA Europe Asia E721 PROBE P9607AE 1 1 For Japan 959 739L PROBE 2107460 2 1 1 For USA Europe Asia 739...

Страница 197: ...Europe Asia C386 PROBE 2154186 1 1 For Japan 977 I739 PROBE 2147189 2 1 1 For USA Europe Asia I739 PROBE 2147189 1 1 For Japan 978 T739 PROBE 2147188 2 1 1 For USA Europe Asia T739 PROBE 2147188 1 1 For Japan 980 LA39 PROBE 2155078 2 1 1 For USA Europe Asia LA39 PROBE 2155078 1 1 For Japan 981 B510 PROBE 2123593 1 1 For USA Europe Asia B510 PROBE 2133115 1 1 For Japan 982 S222 PROBE 2159263 1 1 Fo...

Страница 198: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 68 REV 9 This page is left blank intentionally ...

Страница 199: ...S EXISTS AT SEVERAL POINTS INSIDE BECOME THOROUGHLY FA MILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS TO AVOID ACCI DENTAL CONTACT CAUTION Do not wear the ESD wrist strap when you remove a part of power supply unit Turn OFF power and unplug the power cord before removing a part of power supply unit However be sure to turn off power and wear the strap before you remove a circuit boards ...

Страница 200: ...ocated beneath the horizontal part of the Neck Assy 4 Remove the cover located at the bottom of the Neck Assy 5 Unscrew two screws 5 and 6 and remove the plate fixing the cables 6 Disconnect the four connectors 7 10 7 Remove the screw cap located at position 11 8 Loosen one screw 11 and pull the Monitor Assy upwards Note When replacing the monitor with the new Monitor Assy 2162081 be sure to attac...

Страница 201: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 71 REV 9 6 2 1 Monitor Assy FRU No 100 continued MONITOR ASSY DISASSEMBLY ILLUSTRATION 6 14 ...

Страница 202: ...8 11 10 Disconnect one connector of task lamp switch 11 Remove the Escutcheon Assy FRU 102 Note Monitor Cover Set FRU 101 includes the Escutcheon Assy FRU 102 Escutcheon Latch FRU 103 Monitor Cover L Assy FRU 104 Monitor Cover R Assy FRU 105 Monitor Cover Top Assy FRU 106 and the Escutcheon Front Door FRU 107 Note When replacing the Escutcheon with the new Monitor Cover Set 2162082 or replacing th...

Страница 203: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 73 REV 9 6 2 2 Monitor Cover Set FRU No 101 continued MONITOR COVER SET DISASSEMBLY ILLUSTRATION 6 15 ...

Страница 204: ...tcheon Assy be sure to attach the proper label on the Escutcheon as follows a Choose the proper label LOGIQ 400 or LOGIQ 400 MD indicating the same model name as the label on the removed Escutcheon b Remove the lower pasteboard from the proper label Refer to ILLUSTRATION 6 16 REMOVING LOWER PASTEBOARD ILLUSTRATION 6 16 c Set and attach the label on the Escutcheon Adjust the position of the label a...

Страница 205: ...E MANUAL 2127661 RENEWAL PARTS 6 75 REV 9 6 2 2 Monitor Cover Set FRU No 101 continued d Press down the label all around e Remove the upper pasteboard from the label refer to ILLUSTRATION 6 18 REMOVING UPPER PASTEBOARD ILLUSTRATION 6 18 ...

Страница 206: ...me Required 10 Minute Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 19 1 Turn OFF the system 2 Remove the Escutcheon Assy FRU 102 Refer to 6 2 2 on page 6 72 3 Remove the Escutcheon Latch FRU 103 from the Escutcheon Assy FRU 102 as shown in ILLUSTRATION 6 19 ESCUTCHEON LATCH REMOVAL ILLUSTRATION 6 19 ...

Страница 207: ...efer to ILLUSTRATION 6 20 1 Turn OFF the system 2 Pull out the Escutcheon Front Door FRU 107 from the Escutcheon Assy FRU 102 as shown in ILLUSTRATION 6 20 ESCUTCHEON FRONT DOOR REMOVAL ILLUSTRATION 6 20 Note The color of the Escutcheon for the LOGIQ 400CL is different from the other color monitor assy Use the proper FRU when replacing the monitor cover set for the LOGIQ 400CL ...

Страница 208: ...ute Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 21 1 Turn OFF the system 2 Remove the Escutcheon Front Door FRU 107 Refer to 6 2 4 on page 6 77 3 Pull out two CRT Caps from the front side of the Monitor Assy CRT CAP SET REMOVAL ILLUSTRATION 6 21 Note The CRT Cap Set FRU108 includes the key top of lamp switch ...

Страница 209: ...ed 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 22 1 Turn OFF the system 2 Slide the left side filter clamp leftwards and pull it out 3 Slide the right side filter clamp rightwards and pull it out 4 Remove the CRT Filter CRT FILTER REMOVAL ILLUSTRATION 6 22 Note The CRT Filter FRU109 includes the CRT Filter Clamps FRU110 ...

Страница 210: ...Time Required 0 5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 23 1 Turn OFF the system 2 Remove the Monitor Cover Set FRU 101 Refer to 6 2 2 on page 6 72 3 Unscrew five screws 1 5 4 Disconnect two connectors 6 and 7 5 Remove the Accessory Assy ACCESSORY ASSY DISASSEMBLY ILLUSTRATION 6 23 ...

Страница 211: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 81 REV 9 This page is left blank intentionally ...

Страница 212: ...over L Assy FRU 104 and Monitor Cover R Assy FRU 105 Refer to 6 2 2 on page 6 72 3 Cut off the tie wrap fixing the CRT Cable Set FRU 116 to the speaker 4 Unscrew two screws 1 and 2 5 Loosen two screws 3 and 4 6 Disconnect one connector 5 from the back side of monitor 7 Remove the speaker at the left 8 Unscrew two screws 6 and 7 9 Loosen two screws 8 and 9 10 Disconnect one connector 10 from the ba...

Страница 213: ...MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 83 REV 9 6 2 8 Speaker Assy FRU No 112 continued SPEAKER ASSY DISASSEMBLY ILLUSTRATION 6 24 Note Left and right speakers are interchangeable ...

Страница 214: ...quired Screwdriver Procedure Refer to ILLUSTRATION 6 25 1 Turn OFF the system 2 Remove the Monitor Cover L Assy FRU 104 and Monitor Cover R Assy FRU 105 Refer to 6 2 2 on page 6 72 3 Turn the CRT lamp counterclockwise and remove it 4 Unscrew two screws 1 and 2 5 Disconnect one connector 3 6 Remove the CRT Lamp Assy CRT LAMP ASSY DISASSEMBLY ILLUSTRATION 6 25 ...

Страница 215: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 85 REV 9 This page is left blank intentionally ...

Страница 216: ...S SPRING LOADED RELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVEMENT KEEP YOUR HEAD AND BODY AWAY FROM ABOVE NECK ASSY Refer to ILLUSTRATION 6 26 on page 6 87 1 Turn OFF the system 2 Remove the Monitor Assy FRU 100 Refer to 6 2 1 on page 6 70 3 Remove the Monitor Cover Set 1 FRU 101 Refer to 6 2 2 on page 6 72 4 Remove the Speaker Assy 2 FRU 112 Refer to 6 2 8 on page 6 82...

Страница 217: ... MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 87 REV 9 6 2 10 CRT Assy FRU No 115 continued CRT ASSY DISASSEMBLY ILLUSTRATION 6 26 Note The CRT Assy does not include the tilt mechanism ...

Страница 218: ...ol Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 27 1 Turn OFF the system 2 Remove the Monitor Assy FRU 100 Refer to 6 2 1 on page 6 70 3 Remove the Monitor Cover Set FRU 101 Refer to 6 2 2 on page 6 72 4 Disconnect three connectors 1 3 from the back side of the CRT Assy CRT ASSY DISASSEMBLY ILLUSTRATION 6 27 ...

Страница 219: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 89 REV 9 This page is left blank intentionally ...

Страница 220: ...page 6 91 1 Turn OFF the system 2 Unscrew four hexagonal screws 1 4 located beneath the horizontal part of the Neck Assy 3 Remove the cover located at the bottom of the Neck Assy 4 Unscrew two screws 5 and 6 and remove the plate fixing the cables 5 Disconnect the three connectors 7 9 6 Remove two screw caps located at position 10 and 11 7 Loosen two screws 10 and 11 and pull the Monitor Assy upwar...

Страница 221: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 91 REV 9 6 2 12 Monitor Assy NTSC FRU No 150 Monitor Assy PAL FRU No 151 continued MONITOR ASSY NTSC PAL DISASSEMBLY ILLUSTRATION 6 28 ...

Страница 222: ... Cover FRU No 152 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 29 1 Turn OFF the system 2 Remove the Monitor Bottom Assy Refer to 6 2 17 on page 6 98 3 Unscrew four screws 1 4 4 Remove the Monitor Cover FRU 152 MONITOR COVER DISASSEMBLY ILLUSTRATION 6 29 ...

Страница 223: ...stem 2 Remove the Monitor Cover FRU 152 Refer to 6 2 13 on page 6 92 3 Unscrew four screws 1 4 4 Remove the Escutcheon Assy FRU 153 Note When replacing the Escutcheon with the new Escutcheon Assy 2164507 Be sure to attach the proper label included with the Escutcheon Assy on to the Escutcheon To attach the new label refer to the procedures described in the following pages ESCUTCHEON ASSY REMOVAL I...

Страница 224: ...proper label on to the Escutcheon as follows a Choose the proper label LOGIQ 400 or LOGIQ 400 MD indicating the same model name as the label on the removed Escutcheon b Remove the lower pasteboard from the proper label Refer to ILLUSTRATION 6 31 REMOVING LOWER PASTEBOARD ILLUSTRATION 6 31 c Set and attach the label on the Escutcheon Adjust the position of the label attached as shown in ILLUSTRATIO...

Страница 225: ...CE MANUAL 2127661 RENEWAL PARTS 6 95 REV 9 6 2 14 Escutcheon Assy FRU No 153 continued d Press down the label all around e Remove the upper pasteboard from the label refer to ILLUSTRATION 6 33 REMOVING UPPER PASTEBOARD ILLUSTRATION 6 33 ...

Страница 226: ...U No 154 Time Required 5 Minute Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 34 1 Turn OFF the system 2 Remove the Monitor Bottom Assy FRU 157 Refer to 6 2 17 on page 6 98 3 Pull out the Fixing Metal Plate FRU 154 from the Escutcheon Assy FRU 153 FIXING METAL PLATE REMOVAL ILLUSTRATION 6 34 ...

Страница 227: ...TRATION 6 35 1 Turn OFF the system 2 Slide the filter clamp at left side leftwards and pull it out 3 Slide the filter clamp at right side rightwards and pull it out 4 Remove the CRT Filter FRU 155 CRT FILTER CLAMP AND CRT FILTER REMOVAL ILLUSTRATION 6 35 Note When reinstalling the CRT filter you will hear a click sound when the filter clamp is secured by the lock mechanism This indicates it is pro...

Страница 228: ...ttom Assy FRU No 157 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 36 1 Turn OFF the system 2 Unscrew one screw 1 3 Disconnect two connectors 2 and 3 4 Remove the Monitor Bottom Assy FRU 157 5 Remove the Stud MONITOR BOTTOM ASSY DISASSEMBLY ILLUSTRATION 6 36 ...

Страница 229: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 99 REV 9 This page is left blank intentionally ...

Страница 230: ...TALLED WILL CAUSE SUDDEN UPWARD MOVEMENT KEEP YOUR HEAD AND BODY AWAY FROM ABOVE NECK ASSY Refer to ILLUSTRATION 6 37 on page 6 101 1 Turn OFF the system 2 Remove the Monitor Assy NTSC PAL FRU 150 151 Refer to 6 2 12 on page 6 90 3 Remove the Monitor Bottom Assy FRU 157 Refer to 6 2 17 on page 6 98 4 Remove the Monitor Cover FRU 152 Refer to 6 2 13 on page 6 92 5 Remove the Escutcheon Assy FRU 153...

Страница 231: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 101 REV 9 6 2 18 CRT Assy NTSC FRU No 158 CRT Assy PAL FRU No 159 continued CRT ASSY NTSC PAL DISASSEMBLY ILLUSTRATION 6 37 ...

Страница 232: ...quired Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 38 1 Turn OFF the system 2 Remove the Monitor Assy NTSC PAL FRU 150 151 Refer to 6 2 12 on page 6 90 3 Remove the Monitor Bottom Assy FRU 157 Refer to 6 2 17 on page 6 98 4 Disconnect one connector 1 and remove the CRT Cable Set CRT CABLE SET DISASSEMBLY ILLUSTRATION 6 38 ...

Страница 233: ...uired Screwdriver Procedure Refer to ILLUSTRATION 6 39 1 Turn OFF the system 2 Remove the Monitor Bottom Assy FRU 157 Refer to 6 2 17 on page 6 98 3 Unscrew two screws 1 and 2 to remove the bracket of Monitor Bottom Assy as shown in ILLUSTRATION 6 39 4 Turn the CRT Lamp counterclockwise and remove it CRT LAMP DISASSEMBLY ILLUSTRATION 6 39 ...

Страница 234: ...IDEO MONITOR SUPPORT ARM Refer to ILLUSTRATION 6 40 on page 6 105 1 Turn OFF the system 2 Remove the CRT Filter Refer to ILLUSTRATION 6 35 on page 6 97 3 Unscrew four hexagonal screws 1 4 located beneath the horizontal part of the Neck Assy 4 Remove the cover located at the bottom of the Neck Assy 5 Unscrew two screws 5 and 6 and remove the plate fixing the cables 6 Disconnect the four connectors ...

Страница 235: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 105 REV 9 6 2 21 Monitor15 Assy FRU No 170 continued MONITOR15 BLOCK ASSY DISASSEMBLY ILLUSTRATION 6 40 ...

Страница 236: ... Remove the LAMP SHADE2 FRU179 4 Unscrew six screws 3 8 5 Remove the Monitor Cover L2 Assy FRU 173 and R2 Assy FRU 174 6 Remove the Tray2 Assy FRU 175 7 Loosen four screws 9 12 8 Remove the Escutcheon2 Assy FRU 172 9 Disconnect one connector 13 10 Disconnect one connector 14 from the back side of the Escutcheon2 Assy Note Monitor15 Cover Set includes the Monitor Cover L2 Assy FRU 173 Monitor Cover...

Страница 237: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 107 REV 9 6 2 22 Monitor15 Cover Set FRU No 171 continued MONITOR COVER SET DISASSEMBLY ILLUSTRATION 6 41 ...

Страница 238: ...tach the proper label on to the Escutcheon as follows a Choose the proper label LOGIQ 400 PRO indicating the same model name as the label on the removed Escutcheon b Remove the lower pasteboard from the proper label Refer to ILLUSTRATION 6 42 REMOVING LOWER PASTEBOARD ILLUSTRATION 6 42 c Set and attach the label on the Escutcheon Adjust the position of the label attached as shown in ILLUSTRATION 6...

Страница 239: ...fer to ILLUSTRATION 6 44 1 Turn OFF the system 2 Remove the Monitor Cover L2 Assy FRU 173 and Monitor Cover R2 Assy FRU 174 Refer to 6 2 22 on page 6 106 4 Unscrew eight screws 1 8 5 Disconnect one connector for each speaker from the back side of the monitor 6 Remove the Speakers SPEAKER ASSY DISASSEMBLY ILLUSTRATION 6 44 Note Left and right speakers are interchangeable ...

Страница 240: ...10 REV 9 6 2 24 TASK Lamp FRU No 180 Time Required 0 5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 45 1 Turn OFF the system 2 Unscrew two screws 1 and 2 3 Remove the Lamp Shade2 4 Remove the TASK LAMP FRU 180 TASK LAMP DISASSEMBLY ILLUSTRATION 6 45 ...

Страница 241: ...equired Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 46 1 Turn OFF the system 2 Remove the Monitor15 Assy FRU 170 Refer to 6 2 21 on page 6 104 3 Remove the Monitor15 Cover Set FRU 171 Refer to 6 2 22 on page 6 106 4 Disconnect three connectors 1 3 from the back side of the CRT Assy CRT ASSY DISASSEMBLY ILLUSTRATION 6 46 ...

Страница 242: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 112 REV 9 This page is left blank intentionally ...

Страница 243: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 113 REV 9 This page is left blank intentionally ...

Страница 244: ...t position 3 Remove the Monitor Assy FRU No 100 Monitor Assy NTSC FRU No 150 or Monitor Assy PAL FRU No 151 Refer to 6 2 1 on page 6 70 or 6 2 12 on page 6 90 4 Remove the Neck Space Plate 5 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 6 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 Note The Linear Slide Cover is installed from the production unit of LOGIQ ...

Страница 245: ... for B W Monitor continued 11 Disconnect one connector 1 12 Disconnect two connectors 2 and 3 from the P25A and P26A connectors on Mother Board 13 Disconnect one flat cable 4 14 Cut off the tie wraps to bind cables 15 Remove thirteen hexagonal bolts 5 17 16 Remove the Neck Assy from console NECK ASSY DISASSEMBLY ILLUSTRATION 6 47 ...

Страница 246: ... MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 116 REV 9 6 2 26 Neck Assy FRU No 201 for Color Monitor No 202 for B W Monitor continued LINEAR SLIDE COVER DISASSEMBLY ILLUSTRATION 6 48 ...

Страница 247: ... PARTS 6 117 REV 9 6 2 27 Neck Grip FRU No 203 Time Required 5 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 49 1 Turn OFF the system 2 Loosen two hexagonal screws 1 and 2 3 Remove the Neck Grip NECK GRIP DISASSEMBLY ILLUSTRATION 6 49 ...

Страница 248: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 118 REV 9 This page is left blank intentionally ...

Страница 249: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 119 REV 9 This page is left blank intentionally ...

Страница 250: ...Refer to ILLUSTRATION 6 50 on page 6 121 1 Turn OFF the system 2 Remove the Monitor Assy FRU No 100 Monitor Assy NTSC FRU No 150 or Monitor Assy PAL FRU No 151 Refer to 6 2 1 on page 6 70 or 6 2 12 on page 6 90 3 Remove the Neck Space Plate 2 Refer to 6 2 29 on page 6 122 The Neck Space Plate 2 is used only for the B W monitor console In case of color monitor console skip this step 4 Remove the Ro...

Страница 251: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 121 REV 9 6 2 28 Neck Space Plate FRU No 204 continued NECK SPACE PLATE DISASSEMBLY ILLUSTRATION 6 50 ...

Страница 252: ...crewdriver Hexagonal Wrench Procedure WARNING PERSONAL INJURY HAZARD THE NECK ASSY IS SPRING LOADED RELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVEMENT KEEP YOUR HEAD AND BODY AWAY FROM ABOVE NECK ASSY Refer to ILLUSTRATION 6 51 on page 6 123 1 Turn OFF the system 2 Remove the Monitor Assy NTSC FRU No 150 or Monitor Assy PAL FRU No 151 Refer to 6 2 12 on page 6 90 3 Remov...

Страница 253: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 123 REV 9 6 2 29 Neck Space Plate 2 FRU No 205 continued NECK SPACE PLATE 2 DISASSEMBLY ILLUSTRATION 6 51 ...

Страница 254: ...exagonal Wrench Procedure WARNING PERSONAL INJURY HAZARD THE NECK ASSY IS SPRING LOADED RELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVEMENT KEEP YOUR HEAD AND BODY AWAY FROM ABOVE NECK ASSY Refer to ILLUSTRATION 6 52 on page 6 125 1 Turn OFF the system 2 Remove the Monitor Assy NTSC FRU No 150 or Monitor Assy PAL FRU No 151 Refer to 6 2 12 on page 6 90 3 Unscrew one screw...

Страница 255: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 125 REV 9 6 2 30 Rotation Spacer FRU No 206 continued ROTATION SPACER DISASSEMBLY ILLUSTRATION 6 52 ...

Страница 256: ...GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL 2127661 RENEWAL PARTS 6 126 REV 9 This page is left blank intentionally ...

Страница 257: ...31 Side Cover Left FRU No 301 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 53 1 Turn OFF the system 2 Remove two screw caps located at 3 and 4 and unscrew four screws 1 4 3 Remove the Side Cover Left SIDE COVER LEFT DISASSEMBLY ILLUSTRATION 6 53 ...

Страница 258: ...Screwdriver Procedure Refer to ILLUSTRATION 6 54 1 Turn OFF the system 2 Remove the Front Base Cover FRU No 305 Refer to 6 2 35 on page 6 131 3 Remove the two screw caps located at 3 and 4 and unscrew five screws 1 5 5 4 Remove the Rear Door Assy Refer to 6 2 34 on page 6 130 5 Remove the Side Cover Right SIDE COVER RIGHT DISASSEMBLY ILLUSTRATION 6 54 ...

Страница 259: ... FRU No 303 Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 55 1 Turn OFF the system 2 Pull the Air Filter out 3 Remove the Rear Door Assy Refer to 6 2 34 on page 6 130 4 Unscrew eight screws 1 8 5 Remove the Rear Cover REAR COVER DISASSEMBLY ILLUSTRATION 6 55 ...

Страница 260: ...EV 9 6 2 34 Rear Door Assy FRU No 304 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 56 1 Turn OFF the system 2 Turn the knob on the Rear Door Assy counterclockwise 3 Remove the Rear Door Assy REAR DOOR ASSY DISASSEMBLY ILLUSTRATION 6 56 ...

Страница 261: ... 9 6 2 35 Front Base Cover FRU No 305 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 57 1 Turn OFF the system 2 Pull the Air Filter out 3 Unscrew three screws 1 3 4 Remove the Front Base Cover FRONT BASE COVER DISASSEMBLY ILLUSTRATION 6 57 ...

Страница 262: ...t Cover FRU No 306 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 58 1 Turn OFF the system 2 Remove the Front Base Cover FRU No 305 Refer to 6 2 35 on page 6 131 3 Unscrew four screws 1 4 4 Remove the Front Cover FRONT COVER DISASSEMBLY ILLUSTRATION 6 58 ...

Страница 263: ...ssary to attach the Blind Label on the Front Cover to cover the openings for the HDD LED in case of the console with the new type of 3 5 inch HDD Quantum FireBall TM 1280S Because the new type of 3 5 inch HDD does not support the HDD LED function to indicate the access status of HDD To attach the Blind Label to the Front Cover refer to ILLUSTRATION 6 59 ATTACHING BLIND LABEL TO FRONT COVER ILLUSTR...

Страница 264: ... Procedure Refer to ILLUSTRATION 6 60 1 Turn OFF the system 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Front Cover FRU 306 Refer to 6 2 36 on page 6 132 4 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 5 Unscrew six screws 1 6 6 Remove the Keyboard Bottom Cover KEYBOARD BOTTOM COVER DISASSEMBLY ILLUSTRATION 6 60 ...

Страница 265: ...eet 3 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 4 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 5 Remove the Rear Cover Assy FRU 303 Refer to 6 2 33 on page 6 129 6 Pull the Cable Arm Assy out Refer to 6 2 45 on page 6 142 7 Unscrew two screws 1 and 2 from the upper side of top cover and five screws 3 7 from the lower side of top cover 8 Remove the Top Co...

Страница 266: ...dure Refer to ILLUSTRATION 6 62 1 Turn OFF the system 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Unscrew three hexagonal bolts from each Caster Bumper 1 3 4 Remove the Front Bumpers FRONT BUMPER SET DISASSEMBLY ILLUSTRATION 6 62 Note Reuse the hexagonal bolts removed to install the new front bumper s Note Tall version console is used the BUMPER LONG SET ASSY ...

Страница 267: ...ocedure Refer to ILLUSTRATION 6 63 1 Turn OFF the system 2 Remove the Keyboard Panel Assy FRU 400 Refer to 6 2 49 on page 6 147 3 Unscrew inside three hexagonal bolts 1 3 4 Unscrew outside five screws 4 8 5 Remove the Corner Guard upwards CORNER GUARD DISASSEMBLY ILLUSTRATION 6 63 Note Reuse the hexagonal bolts and screws removed to install the new corner guard ...

Страница 268: ...8 REV 9 6 2 41 Probe Holder FRU No 312 Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 64 1 Turn OFF the system 2 Pull the latch located at bottom of the probe holder 3 Pull the Probe Holder out PROBE HOLDER REMOVAL ILLUSTRATION 6 64 ...

Страница 269: ...older FRU No 313 Gel Holder Bottom FRU No 314 Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 65 1 Turn OFF the system 2 Pull the Gel Holder out 3 Remove the Gel Holder Bottom from the Gel Holder GEL HOLDER AND GEL HOLDER BOTTOM REMOVAL ILLUSTRATION 6 65 ...

Страница 270: ...Handle FRU No 315 Time Required 15 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 66 1 Turn OFF the system 2 Remove the Rear Cover FRU No 303 Refer to 6 2 33 3 Unscrew four hexagonal bolts 1 4 4 Remove the Handle HANDLE DISASSEMBLY ILLUSTRATION 6 66 ...

Страница 271: ... 6 141 REV 9 6 2 44 Air Filter FRU No 316 Time Required 5 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 67 1 Turn OFF the system 2 Pull out the two Air Filters 1 and 2 located at front and rear of console AIR FILTERS REMOVAL ILLUSTRATION 6 67 ...

Страница 272: ...S 2127661 RENEWAL PARTS 6 142 REV 9 6 2 45 Cable Arm Assy FRU No 317 Time Required 1 Minute Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 68 1 Turn OFF the system 2 Pull the Cable Arm Assy out CABLE ARM REMOVAL ILLUSTRATION 6 68 ...

Страница 273: ...t Caster Assy FRU No 318 Time Required 15 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 69 1 Turn OFF the system 2 Put the system on a box 3 Unscrew four hexagonal bolts 1 4 4 Remove the Front Caster Assy FRONT CASTER ASSY DISASSEMBLY ILLUSTRATION 6 69 ...

Страница 274: ... 47 Rear Caster Assy FRU No 319 Time Required 15 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 70 1 Turn OFF the system 2 Put the system on a box 3 Unscrew four screws 1 4 4 Remove the Rear Caster Assy REAR CASTER DISASSEMBLY ILLUSTRATION 6 70 ...

Страница 275: ...er Cover Assy FRU No 320 Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 71 1 Turn OFF the system 2 Remove four labels two for Logo and two for plain 3 Unscrew five screws 1 5 4 Remove the Caster Cover Assy CASTER COVER ASSY DISASSEMBLY ILLUSTRATION 6 71 ...

Страница 276: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 146 REV 9 This page is left blank intentionally ...

Страница 277: ...tem 2 Unscrew three screws 1 3 of the Keyboard Panel Assy 3 Lift the Keyboard Assy up 4 Disconnect the connector 4 5 Remove the Keyboard Panel Assy FRU400 KEYBOARD PANEL ASSY KEYBOARD BOTTOM COVER KEYBOARD PANEL ASSY DISASSEMBLY ILLUSTRATION 6 72 Note The color of the Key Sheet for the LOGIQ 400CL is different from the other color monitor assy Use the proper FRU when replacing the Keyboard Panel A...

Страница 278: ...nd eight TGC knobs Refer to ILLUSTRATION 6 79 on page 6 159 3 Remove the Keyboard Panel Assy Refer to 6 2 49 on page 6 147 4 Unscrew nine screws 1 9 to remove two metal plates 5 Unscrew ten screws 10 19 6 Remove the Keyboard Cover FRU 402 7 Unscrew twelve screws 20 31 8 Remove the Key Sheet FRU 403 out of the Keyboard Assy or the Keyboard Panel2 Assy FRU 401 Note The color of the Key Sheet for the...

Страница 279: ...ANUAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 149 REV 9 6 2 50 Keyboard Assy FRU No 401 Keyboard Cover FRU No 402 Key Sheet FRU No 403 continued KEYBOARD ASSY KEYBOARD COVER AND KEY SHEET DISASSEMBLY ILLUSTRATION 6 73 ...

Страница 280: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 150 REV 9 This page is left blank intentionally ...

Страница 281: ...nob Set FRU No 404 Time Required 20 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 74 1 Turn OFF the system 2 Loosen two hexagonal bolts for each encoder knob 3 Remove six Encoder Knobs 1 6 4 Pull the eight TGC Pot Knobs 7 out KEYBOARD KNOBS DISASSEMBLY ILLUSTRATION 6 74 ...

Страница 282: ...ired 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 75 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to 6 2 49 on page 6 147 3 Unscrew four screws 1 4 4 Disconnect one connector 5 5 Unscrew four screws 6 9 6 Remove The Holding Metal Plate TRACKBALL DISASSEMBLY ILLUSTRATION 6 75 ...

Страница 283: ...6 76 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to 6 2 49 on page 6 147 3 Unscrew four screws 1 4 4 Disconnect the gain encoder connector 5 GAIN ENCODER DISASSEMBLY ILLUSTRATION 6 76 Note The appearance of new Gain Encoder 2206005 is different from the old one 2164871 Note The new Gain Encoder 2206005 can be used only for the console with the software version 3 30 or later ...

Страница 284: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 154 REV 9 This page is left blank intentionally ...

Страница 285: ... 77 1 Turn OFF the system 2 Unscrew three screws 1 3 of the Keyboard Panel Assy 3 Lift the Keyboard Assy up 4 Disconnect the connector 4 5 Remove the Keyboard Panel Assy KEYBOARD PANEL ASSY DISASSEMBLY ILLUSTRATION 6 77 Note The color of the Keyboard Panel for the LOGIQ 400CL is different from the other color monitor assy Use the proper FRU when replacing the Keyboard Panel Assy for the LOGIQ 400C...

Страница 286: ...efer to ILLUSTRATION 6 79 on page 6 159 3 Remove the Keyboard Panel Assy Refer to ILLUSTRATION 6 77 on page 6 155 4 Unscrew twenty seven screws 1 27 4 Remove nine conectors 28 36 6 Remove the SW PWD Assy FRU 462 7 Remove the Rubber Key Assy FRU 452 453 454 and 455 Note The color of the Key Sheet for the LOGIQ 400CL is different from the other color monitor assy Use the proper FRU when replacing th...

Страница 287: ...UAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 157 REV 9 6 2 55 Keyboard Cover Assy FRU No 451 SW PWB FRU No 462 Rubber Key FRU No 452 453 454 and 455 continued KEYBOARD ASSY SW PWB RUBBER KEY DISASSEMBLY ILLUSTRATION 6 78 ...

Страница 288: ...LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS 2127661 RENEWAL PARTS 6 158 REV 9 This page is left blank intentionally ...

Страница 289: ...nob Set FRU No 464 Time Required 20 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 79 1 Turn OFF the system 2 Loosen two hexagonal bolts for each encoder knob 3 Remove six Encoder Knobs 1 6 4 Pull the eight TGC Pot Knobs 7 out KEYBOARD KNOBS DISASSEMBLY ILLUSTRATION 6 79 ...

Страница 290: ... 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 80 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to ILLUSTRATION 6 77 on page 6 155 3 Unscrew four screws 1 3 4 Disconnect one connector 4 5 Unscrew four screws 5 8 6 Remove The Holding Metal Plate TRACKBALL DISASSEMBLY ILLUSTRATION 6 80 ...

Страница 291: ... Assy FRU No 458 Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 81 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to ILLUSTRATION 6 77 on page 6 155 3 Unscrew four screws 1 4 4 Disconnect one connector 5 TGC DISASSEMBLY ILLUSTRATION 6 81 ...

Страница 292: ...me Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 82 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to ILLUSTRATION 6 77 on page 6 155 3 Unscrew four screws 1 3 4 Disconnect one connector 4 5 Remove The Holding Metal Plate FREEZE KEY ASSY DISASSEMBLY ILLUSTRATION 6 82 ...

Страница 293: ...ol Required Screwdriver Procedure Refer to ILLUSTRATION 6 83 Note The same four rotary encoders are used in the keyboard Panel Assy 1 Turn OFF the system 2 Remove the Keyboard Panel Assy Refer to ILLUSTRATION 6 77 on page 6 155 3 Unscrew four screws 1 2 4 Disconnect the gain encoder connector 3 ROTARY ENCODER DISASSEMBLY ILLUSTRATION 6 83 ...

Страница 294: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 164 REV 9 This page is left blank intentionally ...

Страница 295: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 165 REV 9 This page is left blank intentionally ...

Страница 296: ...ve the Rear Cover Assy FRU 303 Refer to 6 2 33 on page 6 129 Note The Linear Slide Cover is installed from the production unit of LOGIQ 400 console with software Ver sion 5 01y or later Refer to ILLUSTRATION 6 48 on page 6 116 Remove the Linear Slide Cover Note Unscrew the two screws 1 and 2 to remove the Rear CONN Spacer Refer to ILLUSTRATION 6 84 The Rear CONN Spacer is installed from the produc...

Страница 297: ...ble 5 out of the CNTIF Assy or CNTIF2 Assy Note The CNTIF2 Assy includes the Ethernet port for the DICOM option 7 Disconnect one flat cable 6 out of the AVIF Assy or BVIF Assy 8 Disconnect two connectors 7 and 8 One of them is connected with the Power Supply Unit and another is connected with the B W Printer Panel 9 Remove the Rear CONN Panel Assy REAR CONN PANEL ASSY DISASSEMBLY ILLUSTRATION 6 85...

Страница 298: ...exagonal box wrench Procedure Refer to ILLUSTRATION 6 86 1 Turn OFF the system 2 Remove the Rear CONN Panel Assy FRU 501 Refer to 6 2 61 on page 6 166 3 Unscrew six hexagonal studs 1 6 using box wrench 4 Remove the CNTIF Assy or the CNTIF2 Assy CNTIF 2 ASSY DISASSEMBLY ILLUSTRATION 6 86 Note The CNTIF2 Assy includes the Eternet port for the optional DICOM network connection ...

Страница 299: ...ired Screwdriver Procedure Refer to ILLUSTRATION 6 87 1 Turn OFF the system 2 Remove the Rear CONN Panel Assy FRU 501 Refer to 6 2 61 on page 6 166 3 Disconnect three connectors 1 3 Note In case of the BVIF Assy disconnect two connectors 1 and 3 4 Unscrew eleven screws 4 14 5 Remove the AVIF Assy or BVIF Assy AVIF BVIF ASSY DISASSEMBLY ILLUSTRATION 6 87 ...

Страница 300: ...quired 1 Hour Tool Required Screwdriver Soldering Iron Procedure Refer to ILLUSTRATION 6 88 1 Turn OFF the system 2 Remove the Rear CONN Panel Assy FRU 501 Refer to 6 2 61 on page 6 166 3 Remove the soldered cord 1 and 2 4 Remove the Circuit Protector toward the direction indicated by the arrow CIRCUIT PROTECTOR DISASSEMBLY ILLUSTRATION 6 88 ...

Страница 301: ...n OFF the system 2 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 3 Remove the ECG Panel Assy FRU 511 Refer to 6 2 66 on page 6 172 4 Disconnect the connector 1 5 Unscrew two screws 2 and 3 6 Remove the ECG Board Assy ECG BOARD ASSY DISASSEMBLY ILLUSTRATION 6 89 Note New type of ECG Board Assy 2207806 has the reserved connector P69 This connector is not used for the LOGIQ 400 co...

Страница 302: ...TE The new type of ECG Panel Assy is installed from production unit of LOGIQ 400 console with soft ware Version 5 01 y or later Refer to page 6 173 1 Turn OFF the system 2 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 3 Disconnect three connectors 1 3 4 Unscrew two screws 4 and 5 5 Loosen two screws 6 and 7 6 Remove the ECG Panel Assy ECG PANEL ASSY DISASSEMBLY a ILLUSTRATION 6...

Страница 303: ... 6 2 66 ECG Panel Assy FRU No 507 continued NOTE To remove the new type of ECG Panel Assy unscrew two screws 1 and 2 Loosen two screws 3 and 4 Remove the new type of ECG Panel Assy after disconnect three connectors 5 7 1 2 3 4 5 6 7 ECG PANEL ASSY DISASSEMBLY b ILLUSTRATION 6 91 ...

Страница 304: ...tem 2 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 3 Disconnect two connectors 1 and 2 4 Unscrew two screws 3 and 4 5 Pull out the Nest Fan Assy leftwards NOTE The FRAME BASE is installed from the production unit of the LOGIQ 400 console with software ver sion 5 01 y or later Refer to ILLUSTRATION 6 92 Unscrew six screws 1 6 to remove the FRAME BASE NEST FAN ASSY DISASSEMBLY a ...

Страница 305: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 175 REV 9 6 2 67 Nest Fan Assy FRU No 511 continued NEST FAN ASSY DISASSEMBLY ILLUSTRATION 6 93 ...

Страница 306: ...U 306 Refer to 6 2 36 on page 6 132 4 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 5 Remove the Keyboard Bottom Cover FRU 307 Refer to 6 2 37 on page 6 134 6 Remove the HDD LED Assy FRU 526 Refer to 6 2 76 on page 6 196 7 Remove the Power Switch Assy FRU 527 Refer to 6 2 77 on page 6 197 8 Disconnect one connector 1 9 Disconnect four connectors 2 5 10 Loosen two screws 6 and 7...

Страница 307: ... SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 177 REV 9 6 2 68 Probe CONN Set Assy FRU No 512 for 3 slots model FRU No 513 for 2 slots model continued PROBE CONN SET ASSY DISASSEMBLY ILLUSTRATION 6 94 ...

Страница 308: ...PARTS 6 178 REV 9 6 2 69 Connector Cover FRU No 514 Time Required 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 95 1 Turn OFF the system 2 Unscrew four screws 1 4 3 Remove the Connector Cover CONNECTOR COVER DISASSEMBLY ILLUSTRATION 6 95 ...

Страница 309: ...e Required 45 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 96 1 Turn OFF the system 2 Remove the Front Cover FRU 306 Refer to 6 2 36 on page 6 132 3 Unscrew eight screws 1 8 4 Remove the connector frame 5 Remove the Shield Finger Long FRU 515 or Shield Finger Short FRU 516 SHIELD FINGER LONG SHORT DISASSEMBLY ILLUSTRATION 6 96 ...

Страница 310: ...cedure Refer to ILLUSTRATION 6 97 on page 6 181 1 Turn OFF the system 2 Remove the Probe CONN Set Assy FRU 512 513 Refer to 6 2 68 on page 6 176 3 Remove the PRAG Assy FRU 519 Refer to 6 2 72 on page 6 182 4 Unscrew eight screws 1 8 5 Remove the connector frame 6 Unscrew twelve screws 9 20 7 Unscrew three screws 21 23 8 Unscrew two studs 24 and 25 9 Remove the Probe CONN 1 Assy or DCON Assy ...

Страница 311: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 181 REV 9 6 2 71 Probe CONN 1 Assy FRU No 517 or DCON Assy FRU No 518 continued PROBE CONN 1 DCON ASSY DISASSEMBLY ILLUSTRATION 6 97 ...

Страница 312: ...Required Screwdriver Procedure Refer to ILLUSTRATION 6 98 1 Turn OFF the system 2 Remove the Probe CONN Set Assy FRU 512 513 Refer to 6 2 68 on page 6 176 3 Unscrew seven screws 1 7 4 Unscrew three screws 8 10 to remove the metal bracket 5 Disconnect two connectors 11 and 12 6 Remove the PRAG Assy PRAG ASSY DISASSEMBLY ILLUSTRATION 6 98 ...

Страница 313: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 183 REV 9 This page is left blank intentionally ...

Страница 314: ...Cable 4 Unscrew two screws 1 and 2 5 Remove the MODD Fan Assy 6 Remove the HDDB Assy FRU 525 Refer to 6 2 75 on page 6 195 7 Unscrew four screws 3 6 8 Remove the upper metal plate 9 Disconnect the SCSI Cable and MODD Power Cable 10 Remove the MODD Holder Assy 11 Unscrew four screws 7 10 12 Remove the MODD Note There are two types of MODD for the spare parts The folding method of CB24A or CB24N cab...

Страница 315: ...SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 185 REV 9 6 2 73 MODD FRU No 520 MODD Fan FRU No 521 MODD Holder Assy FRU No 522 continued MODD MODD FAN ASSY AND MODD HOLDER ASSY REMOVAL ILLUSTRATION 6 99 ...

Страница 316: ...install a new MODD onto the system ILLUSTRATION 6 100 shows the settings of jumper switches when the 3 5 inch HDD is installed on the console Jumper Switches for another type of MODD 2187012 MODD U0052ZA SETTINGS OF SWITCHES ON MODD ILLUSTRATION 6 100 Note If the 2 5 inch HDD is installed on the console set the jumper switches as shown in ILLUSTRATION 6 101 MODD U0052ZA Jumper Switches for another...

Страница 317: ...to the console with the 3 5 inch HDD make sure that the settings of the DIP switches and the jumper switches are as shown in ILLUSTRATION 6 102 MODD 2187012 2 SETTINGS OF SWITCHES ON MODD ILLUSTRATION 6 102 Note When installing the MODD 2187012 2 to the console with the 2 5 inch HDD set the jumper switches as shown in ILLUSTRATION 6 103 MODD 2187012 2 SETTINGS OF SWITCHES ON MODD ILLUSTRATION 6 10...

Страница 318: ...n eject the disk following the procedure below CAUTION Use the emergency ejection procedure only in emergency situations If this procedure is used while a MO disk is inserted data in the disk can be destroyed 1 Turn the system power OFF 2 Unfold and straighten one end of a large wire paper clip 3 Insert the straightened end of the clip into the emergency eject hole and press firmly This will eject...

Страница 319: ...ith new one the following works have to be done a Set up the system preset parameters without regard to current settings For set up procedures refer to LOGIQ 400 Proprietary Manual 2127662 a 1 Set the Country Zone properly It takes approximately 40 minutes to complete this step a 2 Set other three parameters Video Signal Power Frequency and Modem Kind a 3 Reset the system using System Reset functi...

Страница 320: ...played on the CRT monitor screen frequently because the system is initialized as a CFM console The required changes are 1 The following items in the Option Set Menu shall be set to the Not Installed 02 PW Doppler 03 Color Flow Mapping 04 B Color 05 Dop CFM for Intravaginal Probe 06 Segmented Doppler 07 Real Time Auto Doppler Trace 17 Heading VCR Playback 2 The following parameters in the System Pa...

Страница 321: ...5 3 5 Inch HDD Disassembly Refer to ILLUSTRATION 6 106 1 Make sure the customer data is saved on MO disk Refer to 4 3 3 D USER DATA BACK UP on page 4 22 2 Turn OFF the system 3 Remove the Keyboard Panel Assy FRU 400 Refer to 6 2 49 on page 6 147 4 Unscrew two screws 1 and 2 5 Pull out the HDD Holder2 Assy 6 Disconnect three connectors CB24N CB5N and CB45N Refer to ILLUSTRATION 6 106 7 Unscrew four...

Страница 322: ...nutes to complete this step a 2 Set other three parameters Video Signal Power Frequency and Modem Kind a 3 Reset the system using System Reset function of the Service Menu Refer to LOGIQ 400 Proprietary Manual 2127662 b Install the customer data into new HDD using the back up disk s after resetting the system Refer to 4 3 3 D USER DATA BACK UP on page 4 22 c Set the User Option properly To set the...

Страница 323: ...operly and the terminator module IC is NOT mounted on RP1 as shown in ILLUSTRATION 6 107 If the terminator module IC is installed on the 3 5 inch HDD re move it out of the IC socket Fujitsu M1603SA BOTTOM VIEW OF 3 5 INCH HDD ILLUSTRATION 6 107 Note The setting of jumper switches for the new 3 5 inch HDD Assy Quantum Fireball 1280S is differ ent from the old one Refer to ILLUSTRATION 6 108 to chec...

Страница 324: ... 523 HDD Holder Assy FRU No 524 continued 3 5 Inch HDD Disassembly continued Note When using the type Quantum Fireball EL of HDD Assy Refer to ILLUSTRATION 6 109 to check and set the jumper switches on the HDD Assy HDD Assy Quantum Fireball EL SETTING OF JUMPER SWITCHES OF HDD ASSY ILLUSTRATION 6 109 ...

Страница 325: ...he console with the 3 5 inch HDD Assy Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 110 1 Turn OFF the system 2 Remove the Keyboard Panel Assy FRU 400 Refer to 6 2 49 on page 6 147 3 Disconnect three connectors 1 3 4 Unscrew four screws 4 7 5 Remove the HDDB Assy HDDB ASSY DISASSEMBLY ILLUSTRATION 6 110 ...

Страница 326: ...m FireBall 1280S and Quantum FireBall EL Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 111 1 Turn OFF the system 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Front Cover FRU 306 Refer to 6 2 36 on page 6 132 4 Unscrew two screws 1 and 2 5 Disconnect one connector 3 6 Remove the HDD LED Assy HDD LED ASSY DISASSEMBLY ILLUSTRA...

Страница 327: ...quired Screwdriver Procedure Refer to ILLUSTRATION 6 112 1 Turn OFF the system 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Front Cover FRU 306 Refer to 6 2 36 on page 6 132 4 Unscrew two screws 1 and 2 5 Disconnect one connector 3 6 Remove the Power Switch Assy POWER SWITCH ASSY DISASSEMBLY ILLUSTRATION 6 112 ...

Страница 328: ...er to 6 2 31 on page 6 127 3 Unscrew the five Nest Cover screws 4 Remove the Nest Cover 5 Disconnect two connectors if removing the TRDR Assy Otherwise skip this step 6 Disconnect one connector if removing one of two HBFR Assemblies Otherwise skip this step 7 Move the upper and lower stopper in the direction indicated by the arrow 1 8 Pull out the board do not bend it 2 Note After replacing a CPU ...

Страница 329: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 199 REV 9 6 2 78 P C Board s FRU No 601 through 612 continued P C BOARD S REMOVAL ILLUSTRATION 6 113 ...

Страница 330: ...e Nest Cover 5 Remove the MPU Assy FRU 612 Refer to 6 2 78 on page 6 198 6 Remove the Time Keeper RAM mounted on the CPU Assy Note After replacing a time keeper RAM with new one it is necessary to install the system ID number into the new time keeper RAM and set up two items among four system parameters Video Signal and Power Frequency For system ID installation procedures refer to Section 4 3 3 J...

Страница 331: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 201 REV 9 6 2 79 Time Keeper RAM FRU No 613 continued TIME KEEPER RAM REMOVAL ILLUSTRATION 6 114 ...

Страница 332: ... 3 Unscrew the four Nest Cover screws 4 Remove the Nest Cover 5 Remove the MPU Assy FRU 612 Refer to 6 2 78 on page 6 198 6 Remove the Time Keeper Battery Note After replacing a time keeper battery with new one it is necessary to install the system ID number into the time keeper RAM and to set up the factory preset data The LOGIQ 400 can be installed the system ID number into new time keeper RAM u...

Страница 333: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 203 REV 9 6 2 80 Time Keeper Battery FRU No 613B continued TIME KEEPER BATTERY REMOVAL ILLUSTRATION 6 115 ...

Страница 334: ...to ILLUSTRATION 6 116 on page 6 205 1 Turn OFF the system 2 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 3 Unscrew the four Nest Cover screws 4 Remove the Nest Cover 5 Remove the CPU A Assy FRU 610 Refer to 6 2 78 on page 6 198 6 Remove the Time Keeper RAM on the SBC A Assy MVME167 002B Note After replacing a time keeper RAM with new one it is necessary to install the system ID...

Страница 335: ...SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 205 REV 9 6 2 81 Time Keeper RAM for MVME167 002B FRU No 613C continued MPU A ASSY MVME167 002B TIME KEEPER RAM for MVME167 002B REMOVAL ILLUSTRATION 6 116 ...

Страница 336: ... dismounting the piggyback boards Surface mount IC boards are very susceptible to damage from flex torque Refer to ILLUSTRATION 6 117 on page 6 207 1 Turn OFF the system 2 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 3 Unscrew five Nest Cover screws 4 Remove the Nest Cover 5 Remove the TLMC Assy FRU 606 Refer to 6 2 78 on page 6 198 6 Unscrew one screw 1 7 Remove the OMEM Assy ...

Страница 337: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 207 REV 9 6 2 82 OMEM Assy FRU No 614 continued OMEM ASSY DISASSEMBLY ILLUSTRATION 6 117 ...

Страница 338: ... Remove the Front Cover FRU 306 Refer to 6 2 36 on page 6 132 7 Remove the Top Cover FRU 308 Refer to 6 2 38 on page 6 135 8 Unscrew five Nest Cover screws 9 Remove the Nest Cover 10 Remove all the P C Boards Refer to 6 2 78 on page 6 198 11 Remove the Rear CONN Panel Assy FRU 501 Refer to 6 2 61 on page 6 166 12 Remove the HDD LED Assy FRU 526 Refer to 6 2 76 on page 6 196 13 Remove the Power Swi...

Страница 339: ...V 9 6 2 83 Nest Mother Assy FRU No 615 continued 21 Unscrew four hexagonal bolts 1 3 on the left side of the console 22 Unscrew three hexagonal bolts 4 6 on the right side of the console 23 Remove the Nest Mother Assy FRU No 615 NEST MOTHER ASSY DISASSEMBLY ILLUSTRATION 6 118 ...

Страница 340: ... Wrench Procedure Refer to ILLUSTRATION 6 119 on page 6 211 1 Turn OFF the system 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 4 Remove the Rear Cover FRU 303 Refer to 6 2 33 on page 6 129 5 Remove the Neck Assy FRU 201 Refer to 6 2 26 on page 6 114 6 Remove the Mother IC ...

Страница 341: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 211 REV 9 6 2 84 Mother IC FRU No 616 continued MOTHER IC REMOVAL ILLUSTRATION 6 119 ...

Страница 342: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 212 REV 9 This page is left blank intentionally ...

Страница 343: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 213 REV 9 This page is left blank intentionally ...

Страница 344: ...e 6 215 1 Turn OFF the system and unplug the main power cord 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 4 Unscrew four screws 1 4 5 Remove the PS Cover Right 6 Unscrew two screws 5 and 6 7 Disconnect five connectors 7 11 8 Pull the High Voltage Assy out 9 Disconnect the HV Fan Cable 12 10 Remove the High V...

Страница 345: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 215 REV 9 6 2 85 High Voltage Assy FRU No 701 continued HIGH VOLTAGE ASSY DISASSEMBLY ILLUSTRATION 6 120 ...

Страница 346: ... 702 Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 121 1 Turn OFF the system and unplug the main power cord 2 Remove the High Voltage Assy FRU 701 Refer to 6 2 85 on page 6 214 3 Unscrew four screws 1 4 4 Remove the HV Fan HV FAN DISASSEMBLY ILLUSTRATION 6 121 ...

Страница 347: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 217 REV 9 This page is left blank intentionally ...

Страница 348: ...wer cord 2 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 3 Unscrew four screws 1 4 4 Remove the Left PS Cover 5 Unscrew two screws 5 and 6 6 Pull the Low Voltage Unit outwards 7 Unscrew fifteen screws 7 21 to remove wires 8 Pull the Low Voltage Unit out 9 Remove the Plastic Cover as shown in ILLUSTRATION 6 122 on page 6 219 10 Disconnect one connector 22 11 Unscrew two screws 23...

Страница 349: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 219 REV 9 6 2 87 Low Voltage Unit FRU No 703 continued LOW VOLTAGE UNIT DISASSEMBLY ILLUSTRATION 6 122 ...

Страница 350: ...driver Procedure Refer to ILLUSTRATION 6 123 on page 6 221 1 Turn OFF the system and unplug the main power cord 2 Remove the Side Cover Left FRU 301 Refer to 6 2 31 on page 6 127 3 Remove the Left PS Cover Refer to 6 2 87 on page 6 218 4 Unscrew two screws 1 and 2 5 Disconnect eight connectors on the Power Control Unit 3 and 10 6 Remove the Power Control Unit ...

Страница 351: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 221 REV 9 6 2 88 Power Control Unit FRU No 704 continued POWER CONTROL UNIT DISASSEMBLY ILLUSTRATION 6 123 ...

Страница 352: ...Procedure Refer to ILLUSTRATION 6 124 on page 6 223 1 Turn OFF the system and unplug the main power cord 2 Remove the Power Control Unit FRU 704 Refer to 6 2 88 on page 6 220 3 Unscrew two screws 1 and 2 4 Pull the SSR Unit outwards 5 Unscrew eight screws 3 10 to remove eight wires 6 Pull four crimp contacts 11 14 out of the SSR Unit 7 Remove the SSR Unit ...

Страница 353: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 223 REV 9 6 2 89 SSR Unit FRU No 705 continued SSR UNIT DISASSEMBLY ILLUSTRATION 6 124 ...

Страница 354: ...tes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 125 on page 6 225 1 Turn OFF the system and unplug the main power cord 2 Remove the Rear Cover FRU 303 Refer to 6 2 33 on page 6 129 3 Unscrew six screws 1 6 4 Pull the Power Inlet Unit out 5 Disconnect one connector 7 6 Unscrew two screws 8 and 9 to remove two FG wires 7 Remove the Power Inlet Unit ...

Страница 355: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 225 REV 9 6 2 90 Power Inlet Unit FRU No 706 for 115V FRU No 707 for 220V continued POWER INLET UNIT DISASSEMBLY ILLUSTRATION 6 125 ...

Страница 356: ...6 REV 9 6 2 91 PW Air Filter FRU No 708 Time Required 5 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 126 1 Turn OFF the system and unplug the main power cord 2 Remove the Retainer 3 Remove the PW Air Filter PW AIR FILTER REMOVAL ILLUSTRATION 6 126 ...

Страница 357: ...quired 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 127 1 Turn OFF the system and unplug the main power cord 2 Remove the PW Air Filter 3 Unscrew four screws 1 4 4 Remove the Guard Note The Filter Cover Set includes Retainer PW Air Filter and Guard FILTER COVER SET DISASSEMBLY ILLUSTRATION 6 127 ...

Страница 358: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 228 REV 9 This page is left blank intentionally ...

Страница 359: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 229 REV 9 This page is left blank intentionally ...

Страница 360: ...m and unplug the main power cord 2 Remove the Front Base Cover FRU 305 Refer to 6 2 35 on page 6 131 3 Remove the Side Cover Right FRU 302 Refer to 6 2 32 on page 6 128 4 Unscrew four screws 1 4 5 Remove the RIGHT PS COVER 6 Unscrew three screws 5 7 7 Disconnect one flat cable 8 from the PCONT ASSY on the HV Unit 8 Pull the HV Unit outwards 9 Disconnect eight 8 connectors 9 16 out of the HV Unit 1...

Страница 361: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 231 REV 9 6 2 93 HV Unit FRU No 750 continued HV UNIT DISASSEMBLY ILLUSTRATION 6 128 ...

Страница 362: ...ure Refer to ILLUSTRATION 6 129 1 Turn OFF the system and unplug the main power cord 2 Remove the Side Cover Left Refer to 6 2 31 on page 6 127 3 Remove the Left PS Cover Refer to 6 2 95 on page 6 234 4 Unscrew four screws 1 4 5 Pull the Cooling fan outwards 6 Disconnect the cooling fan connector 5 7 Remove the Cooling Fan COOLING FAN DISASSEMBLY ILLUSTRATION 6 129 ...

Страница 363: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 233 REV 9 This page is left blank intentionally ...

Страница 364: ...e Side Cover left Refer to 6 2 31 on page 6 127 3 Unscrew four screws 1 4 4 Remove the Left PS Cover 5 Unscrew two screws 5 and 6 6 Unscrew fourteen 14 screws 7 20 to remove wires 7 Disconnect five connectors 21 25 8 Pull the LV2 Unit outwards 9 Disconnect one connector 26 from the rear of LV2 Unit 10 Unscrew three screws 27 29 11 Remove the LV2 Unit Note When replacing the LV2 Unit re use the met...

Страница 365: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 235 REV 9 6 2 95 LV2 Unit FRU No 752 continued LV2 UNIT REMOVAL ILLUSTRATION 6 131 ...

Страница 366: ...n power cord 2 Remove the Side Cover left Refer to 6 2 31 on page 6 127 3 Remove the Left PS Cover Refer to 6 2 95 on page 6 234 4 Unscrew two screws 1 and 2 5 Disconnect eleven 11 connectors 3 13 from the component side of the TRIAC Unit 6 Pull the TRIAC Unit outwards 7 Disconnect five connectors 14 18 from the NF2 located on the back side of the TRIAC Unit 8 Disconnect two connectors 19 and 20 f...

Страница 367: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 237 REV 9 6 2 96 TRIAC Unit FRU No 753 continued TRIAC UNIT DISASSEMBLY ILLUSTRATION 6 132 ...

Страница 368: ...tes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 133 on page 6 239 1 Turn OFF the system and unplug the main power cord 2 Remove the Rear Cover FRU 303 Refer to 6 2 33 on page 6 129 3 Unscrew six screws 1 6 4 Pull the Power Inlet Unit out 5 Disconnect one connector 7 6 Unscrew two screws 8 and 9 to remove two FG wires 7 Remove the Power Inlet Unit2 ...

Страница 369: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 239 REV 9 6 2 97 Power Inlet Unit2 FRU No 754 for 115V FRU No 755 for 220V continued POWER INLET UNIT2 DISASSEMBLY ILLUSTRATION 6 133 ...

Страница 370: ...ft PS Cover 5 Unscrew two screws 5 and 6 6 Disconnect one connector 7 7 Unscrew fourteen screws 8 21 8 Pull the LV3 Unit Outwards 9 Disconnect one connector 22 from the rear of LV3 Unit 10 Unscrew three screws 23 25 11 Remove the LV3 Unit Note When installing the LV3 Unit it is necessary to remove the HV Unit temporarily Refer to section 6 7 REPLACING LV2 UNIT WITH LV3 UNIT on page 6 276 Note When...

Страница 371: ...2127661 LOGIQ 400 SERVICE MANUAL GE MEDICAL SYSTEMS RENEWAL PARTS 6 241 REV 9 6 2 98 LV3 Unit FRU No 770 continued LV3 UNIT DISASSEMBLY ILLUSTRATION 6 135 ...

Страница 372: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 242 REV 9 This page is left blank intentionally ...

Страница 373: ...Y F501 3 2A 250V Refer to 6 3 2 A CNTIF ASSY F1 1 0A 250V Refer to 6 3 2 B Power Supply Unit F1 5A 300V Refer to 6 3 2 C New power supply unit2 2144481 for 115V and 2144480 for 220V has three replaceable fuses The table below states the items of those fuses TABLE 6 2 REPLACEABLE FUSES FOR POWER SUPPLY UNIT2 LOCATION Ref No RATING REMARKS TRIAC ASSY F301 20A 32V Refer to 6 3 2 D TRIAC ASSY F302 25A...

Страница 374: ... RENEWAL PARTS 6 244 REV 9 6 3 2 Replacement Procedures Parts Required For Power Supply Unit FRU PART NAME QTY PART REMARKS 801 Fuse Set 1 2135619 For Power Supply Unit2 FRU PART NAME QTY PART REMARKS 801B Fuse Set 2 1 2144490 Tools Required Screwdriver ...

Страница 375: ...1 on page 6 127 3 Unscrew five screws and remove the Nest Cover Refer to 6 2 78 on page 6 198 4 Pull the TRDR Assy out of slot 1 Refer to 6 2 78 on page 6 198 5 Replace the defective fuse with new one Make sure that only fuse with the required rated current and of same type is used for replacement For the removal of the F1 fuse refer to ILLUSTRATION 6 136 6 After replacing Fuse re install the TRDR...

Страница 376: ...ve the Rear Cover Refer to 6 2 33 on page 6 129 3 Remove the Rear CONN Panel Assy Refer to 6 2 61 on page 6 166 Note It is not necessary to disconnect cables in order to replace the fuse 4 Replace the defective fuse with new one Make sure that only fuse with the required rated current and of same type are used for replacement 5 After replacing Fuse re install the Rear CONN Panel Assy and Rear Cove...

Страница 377: ...on page 6 127 3 Remove the Left PS Cover Refer to 6 2 87 on page 6 218 4 Remove the Power Control Unit Refer to 6 2 88 on page 6 220 5 Turn the cap 90 degrees counterclockwise and remove the defective fuse as shown in ILLUSTRATION 6 138 6 Install the new fuse Make sure that only fuse with the required rated current and of same type is used for replacement 7 After replacing Fuse re install the Powe...

Страница 378: ... 6 2 87 on page 6 218 4 Remove the defective fuse s located on the TRIAC Unit To remove the F301 or F302 i Pull the defective fuse from the fuse socket on the TRIAC Unit Refer to ILLUSTRATION 6 139 on page 6 249 To remove the F303 ii Turn the cap 90 degrees counterclockwise and pull the cap with the F303 fuse out Refer to ILLUSTRATION 6 139 on page 6 249 5 Install the new fuse Make sure that only ...

Страница 379: ...MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 249 REV 9 6 3 2 Replacement Procedures continued D Fuses on TRIAC Unit of PS Unit2 continued FUSES ON TRIAC UNIT REMOVAL ILLUSTRATION 6 139 ...

Страница 380: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 250 REV 9 This page is left blank intentionally ...

Страница 381: ...LOGIQ 400 Service Software Refer to LOGIQ 400 Proprietary Manual 2127662 Note If you upgrade the system version it is not necessary to backup the user data before installing new software Because the user data remains unchanged after upgrading the system If you install the system software a damaged hard disk it is impossible to backup the user data before installation Therefore it is recommended to...

Страница 382: ...efer to 6 2 74 on page 6 189 for the replacement procedure of hard disk 4 Connect the RS232C cable with the serial port of the Laptop PC Refer to ILLUSTRATION 6 140 5 Connect the connector of another side with the Service port located at the rear side of the LOGIQ 400 as shown in ILLUSTRATION 6 140 Note You may have to disconnect the insite connection CHARACTER TERMINAL CONNECTION ILLUSTRATION 6 1...

Страница 383: ...it Parity None Flow Control None Connector COM1 6 4 3 Initializing Hard Disk CAUTION Do not initialize the hard disk when upgrading the system software version up Initialize the hard disk before installing the system software only when the hard disk has been replaced with new one 1 Turn the main power switch of the LOGIQ 400 system ON The system does not perform the normal power up sequence and th...

Страница 384: ...ompletes the formatting of hard disk The message Format Completed is displayed on the screen of Laptop PC 3 Type the following command and press the return key The hard disk become valid by entering this command VTSH mount_0 0 6 4 4 Installing Software 1 Insert the appropriate system software MO disk into the MO disk drive of LOGIQ 400 2 Type the following command and press the return key The syst...

Страница 385: ...untry Zone properly It takes approximately 40 minutes to complete this step 3 2 Set other three parameters Video Signal Power Frequency and Modem Kind 3 3 Reset the system using System Reset function of the Service Menu Refer to LOGIQ 400 Proprietary Manual 2127662 4 Perform functional checks to verify that the system is functioning properly Refer to Chapter 4 FUNCTIONAL CHECKS in this manual 5 Lo...

Страница 386: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 256 REV 9 This page is left blank intentionally ...

Страница 387: ... through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Possible operational damage Failure to strictly follow ESD Electrostatic Discharge precau tions during this upgrade may cause constant or intermittent operational abnormalities Strictly follow all precautions Check each step as it is completed Parts Required To replace the 115V Powe...

Страница 388: ...urn the system power OFF Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Remove the Side Cover Left Refer to 6 2 31 on page 6 127 3 Remove the Side Cover Right Refer to 6 2 32 on page 6 128 4 Remove the Rear Cover Refer to 6 2 33 on page 6 129 5 Unscrew two screws 1 and 2 to remove the Connector Guard Metal Refer to ILLUSTRATION 6 142 CONNECTOR GUARD METAL DISASSE...

Страница 389: ...WAL PARTS 6 259 REV 9 6 5 2 Replacement Procedures continued 6 Disconnect eight connectors 1 8 from the PS Unit All connectors are located above the Power Inlet unit at rear of the PS Unit Refer to ILLUSTRATION 6 143 CONNECTORS ON PS UNIT ILLUSTRATION 6 143 ...

Страница 390: ...ontinued 7 Support the left side of PS Unit by one person 8 Unscrew three Hexagonal bolts 1 3 at the left side Refer to ILLUSTRATION 6 144 9 Support the right side of PS Unit by one person 10 Unscrew three Hexagonal bolts 4 6 at the right side Refer to ILLUSTRATION 6 144 PS UNIT DISASSEMBLY ILLUSTRATION 6 144 ...

Страница 391: ...al Procedures 1 Plug in the system set the circuit breaker to the ON position and turn the main power switch ON 2 Perform all power supply checks Refer to 4 4 POWER SUPPLY ADJUSTMENT in Chapter 4 3 Perform the leakage current test Refer to 7 3 ELECTRICAL SAFETY TESTS in Chapter 7 4 Re install the covers removed 5 Perform functional checks to verify that the system is functioning properly Refer to ...

Страница 392: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 262 REV 9 This page is left blank intentionally ...

Страница 393: ...HDD to HDD LED Ass through CB46 Tie Wrap 2 This part is required only for LOGIQ 500 Note If the console includes the Quantum FireBall 1280S 3 5 inch HDD Assy it is necessary to use the CB45N Cable Assy for the HDD LED Assy Because the HDD LED Assy is not available on the con sole with this type of 3 5 inch HDD Assy 6 6 4 Procedures A Removing 2 5 inch HDD Assy HDDB Assy and Cables 1 Turn the main ...

Страница 394: ...Q 400 SERVICE MANUAL RENEWAL PARTS 6 264 REV 9 6 6 4 Procedures continued A Removing 2 5 inch HDD Assy HDDB Assy and Cables continued 8 Unscrew four screws and remove the MODD Cover Metal REMOVING MODD COVER METAL ILLUSTRATION 6 145 ...

Страница 395: ...DDB Assy and Cables continued 9 Disconnect the CB24M cable Refer to ILLUSTRATION 6 146 DISCONNECTING CB24 CABLE ILLUSTRATION 6 146 10 Disconnect three cables CB24M CB5 and CB46 from the HDDB Assy Refer to ILLUSTRATION 6 147 REMOVING HDDB ASSY ILLUSTRATION 6 147 11 Unscrew four screws and remove the HDDB Assy Refer to ILLUSTRATION 6 147 ...

Страница 396: ...tinued 12 Unscrew two screws and remove the 2 5 inch HDD Assy with HDD Holder Assy Refer to ILLUSTRATION 6 148 REMOVING 2 5 INCH HDD ASSY ILLUSTRATION 6 148 13 Disconnect the CB24M cable from the Mother Assy and Rear1 Assy and the CB5 cable from the Mother Assy Refer to ILLUSTRATION 6 149 REMOVING CABLES ILLUSTRATION 6 149 14 Remove the CB24M and CB5 cables ...

Страница 397: ...bles 15 Connect the CB24N Cable with the P24A connector on the Mother Assy Refer to ILLUSTRATION 6 150 16 Connect the CB5N Cable with the P5A connector on the Mother Assy Refer to ILLUSTRATION 6 150 REMOVING CABLES ILLUSTRATION 6 150 17 Pass another side of cables through the opening of Frame forward to the MODD and HDD ...

Страница 398: ...ke sure that the Terminator Module IC is not mounted on the 3 5 inch HDD Assy Refer to ILLUSTRATION 6 151 Fujitsu M1603SA REMOVING CABLES ILLUSTRATION 6 151 Note The setting of jumper switches for the new 3 5 inch HDD Assy Quantum Fireball 1280S is differ ent from the old one Refer to ILLUSTRATION 6 152 to check and set the jumper switches properly Quantum Fireball 1280S BOTTOM VIEW OF 3 5 INCH HD...

Страница 399: ... Installing 3 5 Inch HDD Assy Note The setting of jumper switches for the new 3 5 inch HDD Assy Quantum Fireball EL is different from the other one Refer to ILLUSTRATION 6 152 to check and set the jumper switches properly HDD Assy Quantum Fireball EL SETTING OF JUMPER SWITCHES OF HDD ASSY ILLUSTRATION 6 153 ...

Страница 400: ...ecessaey to install those cables to the LOGIQ 400CL console 22 Set the 3 5 inch HDD Assy on the HDD Holder2 Assy Refer to ILLUSTRATION 6 154 23 Screw four screws supplied with the HDD Change Kit to install the 3 5 inch HDD Assy to the HDD Holder2 Assy Refer to ILLUSTRATION 6 154 INSTALLING 3 5 INCH HDD ASSY ON HDD HOLDER2 ASSY ILLUSTRATION 6 154 Note The HDD LED Assy is not available on the consol...

Страница 401: ...ch HDD Assy continued 24 Screw two screws supplied with the HDD Change Kit to install the 3 5 inch HDD Assy with HDD Holder2 Assy on the console Refer to ILLUSTRATION 6 155 25 Connect the CB24N and CB5N cables with the 3 5 inch HDD Assy as shown in ILLUSTRATION 6 155 INSTALLING 3 5 INCH HDD ASSY ON CONSOLE ILLUSTRATION 6 155 ...

Страница 402: ...Note Before connecting the CB24N cable fold the CB24N cable suitably Refer to ILLUSTRATION 6 156 Note The folding method of CB24N cable is different depending on the type of MODD U0052ZA or 2187012 Before connecting the CB24N cable fold the CB24N cable suitably Refer to ILLUSTRATION 6 156 INSTALLING 3 5 INCH HDD ASSY ON CONSOLE ILLUSTRATION 6 156 ...

Страница 403: ... REV 9 6 6 4 Procedures continued C Installing 3 5 Inch HDD Assy continued 26 Change the settings of jumper switches on the MODD Assy as shown in ILLUSTRATION 6 157 Jumper Switches for another type of MODD 2187012 MODD U0052ZA MODD JUMPER SWITCH SETTINGS ILLUSTRATION 6 157 ...

Страница 404: ...TTINGS OF SWITCHES ON MODD ILLUSTRATION 6 158 27 Connect the CB24N with the MODD Assy Refer to ILLUSTRATION 6 146 on page 6 265 28 Screw four screws and install the MODD Cover Metal Refer to ILLUSTRATION 6 145 on page 6 264 29 Re install the Side Cover Right Refer to 6 2 32 on page 6 128 30 Re install the Front Base Cover Refer to 6 2 35 on page 6 131 31 Re install the Keyboard Panel Assy Refer to...

Страница 405: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 275 REV 9 This page is left blank intentionally ...

Страница 406: ...Removing LV2 Unit 1 Turn the LOGIQ 400 system power OFF 2 Turn the circuit breaker located at the rear side of the Power Supply to the OFF position lower side 3 Unplug the unit Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 4 Remove two screw caps located at the place of the 3 and 4 screws 5 Unscrew four screws 1 4 Refer to ILLUSTRATION 6 159 on page 6 277 6 Remove...

Страница 407: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 277 REV 9 6 7 4 Procedures continued A Removing LV2 Unit continued REMOVING COVERS ILLUSTRATION 6 159 ...

Страница 408: ...A Removing LV2 Unit continued 11 Unscrew four screws 1 4 Refer to ILLUSTRATION 6 160 12 Remove the Left PS Cover REMOVING LEFT PS COVER ILLUSTRATION 6 160 13 Unscrew two screws 1 and 2 Refer to ILLUSTRATION 6 161 14 Pull the LV2 Unit outwards Refer to ILLUSTRATION 6 161 REMOVING LV2 UNIT ILLUSTRATION 6 161 ...

Страница 409: ...9 6 7 4 Procedures continued A Removing LV2 Unit continued 15 Unscrew fourteen 14 screws 1 14 to remove the wires Refer to ILLUSTRATION 6 162 16 Disconnect five connectors 15 19 Refer to ILLUSTRATION 6 162 17 Pull the LV2 Unit Outwards FRONT VIEW OF LV2 UNIT ILLUSTRATION 6 162 ...

Страница 410: ...onnect one connector 1 from the rear of LV2 Unit Refer to ILLUSTRATION 6 163 19 Unscrew three screws 2 4 to remove the three wires Refer to ILLUSTRATION 6 163 20 Remove the LV2 Unit REAR VIEW OF LV2 UNIT ILLUSTRATION 6 163 21 Disconnect the CN31 cable from the TRIAC Unit Refer to ILLUSTRATION 6 164 DISCONNECTING CN31 CABLE ILLUSTRATION 6 164 ...

Страница 411: ... LV3 Unit 1 Screw two screws 1 and 2 to fix the two wires LINE L and NEUTRAL N Refer to ILLUSTRATION 6 165 REAR VIEW OF LV3 UNIT ILLUSTRATION 6 165 2 Insert the LV3 Unit as shown in ILLUSTRATION 6 166 3 Screw two screws 1 and 2 to install the LV3 Unit Refer to ILLUSTRATION 6 166 INSTALLING LV3 UNIT ILLUSTRATION 6 166 ...

Страница 412: ...7 5 Remove the Right PS Cover Refer to ILLUSTRATION 6 167 6 Unscrew three 3 screws 5 7 Refer to ILLUSTRATION 6 167 7 Disconnect one connector 8 Refer to ILLUSTRATION 6 167 8 Pull the HV Unit out RIGHT SIDE VIEW OF POWER SUPPLY ILLUSTRATION 6 167 Note It is not necessary to remove the HV Unit completely Pull the HV Unit out to get the openings for accessing the rear of the LV3 Unit by a hand ...

Страница 413: ...ne connector 1 as shown in ILLUSTRATION 6 168 REAR VIEW OF LV3 UNIT b ILLUSTRATION 6 168 10 Screw three 3 screws 5 7 to install the HV Unit Refer to ILLUSTRATION 6 167 on page 6 282 11 Connect one connector 8 Refer to ILLUSTRATION 6 167 on page 6 282 12 Screw four 4 screws 1 4 to install the Right PS Cover Refer to ILLUSTRATION 6 167 on page 6 282 ...

Страница 414: ...SERVICE MANUAL RENEWAL PARTS 6 284 REV 9 6 7 4 Procedures continued B Installing LV3 Unit continued 13 Screw fourteen 14 screws 1 14 to fix the wires to the appropriate terminals Refer to ILLUSTRATION 6 169 FIXING WIRES ILLUSTRATION 6 169 ...

Страница 415: ...LV3 Unit continued 14 Connect the CN31 cable for LV3 Unit supplied with the LV3 Unit to the TRIAC Unit Refer to ILLUSTRATION 6 170 15 Connect another side of CN31 cable for LV3 Unit to the connector located at the front of the LV3 Unit as shown in ILLUSTRATION 6 170 CONNECTING CN31 CABLE FOR LV3 UNIT ILLUSTRATION 6 170 ...

Страница 416: ...USTMENTS in chapter 4 For the tolerances of the power supplies refer to TABLE 4 3 For the measurement locations refer to TABLE 4 4 and ILLUSTRATION 4 32 on page 4 43 5 Adjust the power supplies to the appropriate voltage as necessary if the they are out of tolerance 6 Perform functional checks to verify that the system is functioning properly Refer to Chapter 4 FUNCTIONAL CHECKS in the LOGIQ 400 S...

Страница 417: ...er of the Power Supply Assy over the old label located at the front side of the Power Supply Unit as shown in ILLUSTRATION 6 172 ATTACHING PN LABEL ILLUSTRATION 6 172 Note Two types of Indication labels are supplied with the LV3 Unit One indicates the part number of the 220V Power Supply Unit and another indicates the part number of the 100 120V Power Supply Unit Be sure to choose and attach the a...

Страница 418: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL RENEWAL PARTS 6 288 REV 9 This page is left blank intentionally ...

Страница 419: ... PERIODIC MAINTENANCE PROCEDURE 7 3 7 2 1 Visual Inspection 7 3 7 2 2 Cleaning 7 4 7 2 3 Measurement 7 6 7 2 4 User Data Backup 7 6 7 2 5 Note 7 6 7 3 ELECTRICAL SAFETY TEST 7 7 7 3 1 Outlet Test Wiring Arrangement 7 7 7 3 2 Ground Continuity 7 8 7 3 3 Chassis Leakage Current Test 7 9 7 3 4 Probe Leakage Current Test 7 12 7 3 5 ECG Leakage Current Test 7 16 7 3 6 When There s Too Much Leakage Curr...

Страница 420: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL 7 2 PERIODIC MAINTENANCE REV 0 This page is left blank intentionally ...

Страница 421: ...tenance is based on an estimated equipment usage of 40 hours per week 7 2 PERIODIC MAINTENANCE PROCEDURE 7 2 1 Visual Inspection a Check cables for cracks nicks and abrasions of their insulation b Check that cables sheath are securely clamped in cable glands connector clamps where applicable c Check that cables are correctly routed and secured d Check that all screws are in place and not missing e...

Страница 422: ...ter Set the CRT filter on the monitor insert the CRT filter clamps 1 and slide them inwards securely 2 INSTALLING CRT FILTER AND CLAMPS ILLUSTRATION 7 1 Note Make sure that the filter is securely fixed by the filter clamps These are important to prevent a hazard from the filter falling A click sound will be heard when the filter clamp is secured by the lock mecha nism Note The CRT Filter Clamps an...

Страница 423: ...ger Guards by using a vacuum cleaner and bristled brush c Clean two Air Filters and one PW Air Filter by using a bristled brush If these filters are damaged replace them For the positions of three Air Filters refer to ILLUSTRATION 7 2 AIR FILTERS REMOVAL ILLUSTRATION 7 2 d Clean out the interior of the Console by using a vacuum cleaner ...

Страница 424: ...r to section 4 4 Power Supply Adjustment in Chapter 4 Functional Checks for adjustment proce dure 7 2 4 User Data Backup a Keep user preset data into a MO Magneto Optical disk using the User Data Backup of LOGIQ 400 Utility Menu Refer to 4 3 3 Utility Menu in Chapter 4 FUNCTIONAL CHECKS Note Use new initialized MO disk It is impossible to save any user data files into the disk which another file i...

Страница 425: ...e For more details contact your regional service support center Recommended Tool PART NUMBER NAME DESCRIPTION 46 285652G1 Electric Safety Analyzer DALE 600 For 120V Units 46 328406G2 Electric Safety Analyzer DALE 600E For 220V Units 2113015 Leakage Current Ultrasound Kit For 120V and 220V Units 7 3 1 Outlet Test Wiring Arrangement Test all outlets in the area for proper wiring arrangement by plugg...

Страница 426: ...y Caster Wheel Support GROUND CONTINUITY TEST ILLUSTRATION 7 4 Meter Procedure Follow these steps to test the ground wire resistance 1 Turn the LOGIQ 400 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet Refer to ILLUSTRATION 7 6 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of ...

Страница 427: ...ls must be tested Generic Procedure The test verifies the isolation of the power line from the chassis The testing meter is connected from accessible metal parts of the case to ground Measurements should be made with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current POWER OUTLET H BLACK POLARITY REVERSING SWITCH MOMENTARY SWITCH N WHITE G GR...

Страница 428: ...e meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ 400 unit Refer to ILLUSTRATION 7 6 4 Set the tester s FUNCTION switch to CHASSIS position 5 Follow the test conditions described in the following table for every test point on the unit TABLE 7 1 CHASSIS LEAKAGE CURRENT TEST CONDITIONS TEST CONDITION 1 Mounting screw for transducer receptacle ...

Страница 429: ...0 5mA 0 5mA CAUTION Electric Shock Hazard When the meter s ground switch is OPEN don t touch the unit CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the PO LARITY switch and or the NEUTRAL switch Otherwise the unit may be damaged TABLE 7 3 DATA SHEET FOR CHASSIS SOUR...

Страница 430: ...open and closed with power line polarity normal and reversed and with the unit Off and On For each combination the probe must be active to find the worst case condition POWER OUTLET H BLACK POLARITY REVERSING SWITCH MOMENTARY SWITCH N WHITE G GREEN LEAKAGE TEST METER CONSOLE PROBE SETUP FOR PROBE LEAKAGE CURRENT ILLUSTRATION 7 7 Note Each probe will have some amount of leakage current dependent on...

Страница 431: ...m salt per 1 liter of water If saline is not available a mixture of 1 quart water with one or more grams of table salt mixed thoroughly will substitute CAUTION To avoid probe damage and possible electric shock do not immerse probes into any liquid beyond the level indicated in the probe users manual Do not touch the probe conductive liq uid or any part of the unit under test while the ISO TEST swi...

Страница 432: ...t 1 Turn the LOGIQ 400 unit OFF 2 Plug the unit into the test meter and the meter into the tested AC wall outlet 3 Plug the external probe into the meter s EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Apply a liberal amount of gel to the transducer s lens 7 Wrap the transducer in aluminum foil making sure...

Страница 433: ...USA 0 05mA 0 05mA 0 05mA 0 05mA OTHERS 0 1mA 0 5mA 0 5mA 0 5mA 1 Type BF Applied Part Leakage Current Limits Non conductive floating Surface and Cavity Transducers 2 Type CF Applied Part Leakage Current Limits Surgical Transducer and ECG Connections CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OF...

Страница 434: ...rols such as the lead switch should be operated to find the worst case condition H Black N White G Green Internal circuitry Insulating surface Polarity reversing switch use both positions Appliance power switch use both off and on positions Patient connected leads The connection is at service entrance or on the supply side of a separately derived system Building ground Grounding contact switch use...

Страница 435: ...or anyone while the ISO TEST switch is depressed When the meter s ground switch is OPEN don t touch the unit Follow these steps to test the ECG module for leakage current 1 Turn the LOGIQ 400 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet Refer to ILLUSTRATION 7 10 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis c...

Страница 436: ...eet Refer to TABLE 7 8 8 Set the meter s FUNCTION switch to LEAD TO LEAD position to measure the lead to lead leakage current 9 Select and test each ECG lead positions except ALL of the LEAD selector testing each to the power condition combinations shown in TABLE 7 6 10 Record the lead to lead leakage current measured on the data sheet Refer to TABLE 7 8 11 Set the meter s FUNCTION switch to LEAD ...

Страница 437: ... sure to turn the unit power OFF before switching the polarity using the POLARITY switch Oth erwise the unit may be damaged TABLE 7 8 DATA SHEET FOR ECG LEAKAGE CURRENT ON NORM CLOSED ON REVERSE CLOSED ON NORM OPEN ON REVERSE OPEN OFF NORM CLOSED OFF REVERSE CLOSED OFF NORM OPEN OFF REVERSE OPEN ECG Tester Tester Tester Lead Selector Power Polarity GROUND Switch Switch RL RA LA LL C Patient Lead I...

Страница 438: ... indicate high leakage currents this should be suspected as a possible cause and the outlet wiring should be visually inspected Probe Fails Test the probe in another transducer interface connector If excessive leakage current is slot dependent inspect the Transducer Interface board for bent pins poor connections and ground continuity If the problem still has stayed with the probe replace the probe...

Страница 439: ...tor 8 13 8 2 8 Connecting Cables 8 19 8 2 9 Setting DIP Switches 8 20 8 2 10 Operational Check out 8 20 8 2 11 Final Procedures 8 23 8 2 12 Renewal Parts 8 24 8 3 COLOR VIDEO PRINTER INSTALLATION 8 25 8 3 1 Foreword 8 25 8 3 2 Tools Required 8 25 8 3 3 Time Required 8 25 8 3 4 Parts Required 8 25 8 3 5 Functional Check out 8 26 8 3 6 Installing Color Video Printer on Console 8 26 8 3 7 Connecting ...

Страница 440: ...1 8 5 ECG OPTION INSTALLATION 8 52 8 5 1 Foreword 8 52 8 5 2 Tools Required 8 52 8 5 3 Time Required 8 52 8 5 4 Parts Required 8 52 8 5 5 Functional Check out 8 52 8 5 6 Installing ECG Board Assy 8 53 8 5 7 Installing ECG Panel Assy 8 55 8 5 8 Operational Check out 8 57 8 5 9 Attaching Caution Label 8 66 8 5 10 Final Procedures 8 67 8 5 11 Renewal Parts 8 69 8 6 EXPANDED CINE MEMORY INSTALLATION 8...

Страница 441: ...quired 8 81 8 8 3 Time Required 8 81 8 8 4 Parts Required 8 81 8 8 5 Functional Check out 8 81 8 8 6 Installing Swivel Lock 8 82 8 8 7 Operational Check out 8 84 8 8 8 Final Procedures 8 84 8 9 LEFT SIDE PROBE HOLDER INSTALLATION 8 85 8 9 1 Foreword 8 85 8 9 2 Tools Required 8 85 8 9 3 Time Required 8 85 8 9 4 Parts Required 8 85 8 9 5 Installing Left Side Probe Holder 8 85 8 9 6 Operational Check...

Страница 442: ...GE MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL TM 8 4 OPTIONS REV 0 This page is left blank intentionally ...

Страница 443: ... renewal parts for optional devices The installation procedures for the following optional devices are included in this chapter Video Cassette Recorder Sony SVO 9500 Color Video Printer Sony UP 1850 and UP 2950 Series Printers B W Video Printer Sony UP 890 ECG Expanded Cine Memory Foot Switch Swivel Lock Left Side Probe Holder MTZ Probe Holder ...

Страница 444: ... of monitor Regarding installation procedure for each way refer to 8 2 6 Installing VCR on Console or 8 2 7 Installing VCR on Color Monitor Procedures respectively 8 2 2 Tools Required Standard field Service Tool kit 8 2 3 Time Required Approx 1 hour for installing the VCR on the console Approx 1 5 hour for installing the VCR on the color monitor 8 2 4 Parts Required a When installing the VCR on t...

Страница 445: ... in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Equipment damage possibility If the video cassette recorder is installed on the monitor it is necessary to lock the monitor s tilt and up down mechanism to prevent the equipment dam age Call cus...

Страница 446: ...ower OFF and unplug the unit Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Remove the Top Cover Sheet out of the system Refer to ILLUSTRATION 8 1 3 Raise the monitor up by pushing down the Up Down Release Button located on the Monitor Arm Refer to ILLUS TRATION 8 1 CONSOLE ILLUSTRATION 8 1 ...

Страница 447: ... TM 8 9 OPTIONS REV 0 8 2 6 Installing VCR on Console Continued 4 Put the OPTION TRAY 1 on the console and screw four screws supplied with VTR ASSY 2127668 to fix the OPTION TRAY 1 as shown in ILLUSTRATION 8 2 OPTION TRAY 1 INSTALLATION ILLUSTRATION 8 2 ...

Страница 448: ...IQ 400 SERVICE MANUAL TM 8 10 OPTIONS REV 0 8 2 6 Installing VCR on Console Continued 5 Unscrew four screws fixing the rubber foots of the Video Cassette Recorder as shown in ILLUSTRATION 8 3 VIDEO CASSETTE RECORDER ILLUSTRATION 8 3 ...

Страница 449: ...MANUAL TM 8 11 OPTIONS REV 0 8 2 6 Installing VCR on Console Continued 6 Screw four screws supplied with VTR ASSY 2127668 to install the OPTION TRAY 2 to the Video Cassette Re corder as shown in ILLUSTRATION 8 4 TRAY INSTALLATION ILLUSTRATION 8 4 ...

Страница 450: ...TM 8 12 OPTIONS REV 0 8 2 6 Installing VCR on Console Continued 7 Place the OPTION TRAY 2 with the Video Cassette Recorder on the OPTION TRAY 1 and screw four screws supplied with VTR ASSY 2127668 as shown in ILLUSTRATION 8 5 FIXING TRAY ILLUSTRATION 8 5 ...

Страница 451: ...Q 400 SERVICE MANUAL TM 8 13 OPTIONS REV 0 8 2 7 Installing VCR on Color Monitor 1 Turn the system power OFF and unplug the unit 2 Remove four seals blinding the screw hole from the top of Color Monitor SEALS REMOVAL ILLUSTRATION 8 6 ...

Страница 452: ...Remove the left and right speakers Refer to the 6 2 8 in Chapter 6 Renewal Parts 5 Install the Plates to the Monitor at the both left and right side as shown in ILLUSTRATION 8 7 Use four screws supplied with the VTR ON MONITOR ASSY P9522YE to fix the Plates 6 Attach the Stopper Block on the Monitor Arm so as to cover the Up down Release Button Refer to ILLUSTRA TION 8 7 PLATES AND STOPPER BLOCK IN...

Страница 453: ...ling VCR on Color Monitor Continued 7 Install the speakers removed 8 Install the Monitor Cover Left and Right removed 9 Screw four screws supplied with VTR ON MONITOR ASSY P9522YE to install the VTR BASE onto the console as shown in ILLUSTRATION 8 8 VTR BASE INSTALLATION ILLUSTRATION 8 8 ...

Страница 454: ...VICE MANUAL TM 8 16 OPTIONS REV 5 8 2 7 Installing VCR on Color Monitor Continued 10 Pass the VTR BELT supplied with the VTR on Monitor Assy P9522YE through the two slits of the VTR BASE as shown in ILLUSTRATION 8 9 VTR BELT ILLUSTRATION 8 9 ...

Страница 455: ...SERVICE MANUAL TM 8 17 OPTIONS REV 0 8 2 7 Installing VCR on Color Monitor Continued 11 Put the Video Cassette Recorder onto the VTR BASE and fix the VCR with the VTR BELT Refer to ILLUSTRA TION 8 10 OPTION TRAY FIXATION ILLUSTRATION 8 10 ...

Страница 456: ...E MANUAL TM 8 18 OPTIONS REV 5 8 2 7 Installing VCR on Color Monitor Continued 12 Attach the Caution Label supplied with the VTR on Monitor Assy P9522YE onto the Video Cassette Recorder Refer to ILLUSTRATION 8 11 CAUTION LABEL ILLUSTRATION 8 11 ...

Страница 457: ...lied power by other outlet the system may be damaged by electrical noise transferred with the video signal or may not pass leakage current test 1 Connect the power cable RS232C cable and the two S Video cables between the Video Cassette Recorder and the console as shown in ILLUSTRATION 8 12 CONNECTING CABLES ILLUSTRATION 8 12 Note The arrangement of I O ports on the Rear Panel of B W monitor syste...

Страница 458: ...e control with the LOGIQ 400CL console 8 2 9 Setting DIP Switches 1 Set the DIP switches located on the rear panel of the video cassette recorder as shown in ILLUSTRATION 8 13 Note The LOGIQ 400CL does not provide the VCR remote control function Set the DIP switched as shown in ILLUSTRATION 8 13 DIP SWITCH SETTINGS ILLUSTRATION 8 13 8 2 10 Operational Check out 1 Plug in the system and power the s...

Страница 459: ...ntrol with the following condi tions I Use S Video cable for connecting between LOGIQ 400 and VCR II Connect RS232C cable with port A located at rear panel of LOGIQ 400 a Press the Top Menu Select key located on the keyboard The sub menu selected previously is displayed on the CRT monitor screen SOFTWARE MENU CONTROL KEYS ILLUSTRATION 8 15 b Press the Top Menu Select key again The top menu is disp...

Страница 460: ...LLUSTRATION 8 17 SET UP MENU ILLUSTRATION 8 17 e Select the System Paramtr section using the or key of Sub Menu Select keys The System Paramtr selection is displayed in reverse video as shown in ILLUSTRATION 8 17 f Press key to select the System Parameter Set Up Menu The system parameter set up menu is dis played on the CRT monitor screen as shown in ILLUSTRATION 8 18 SYSTEM PARAMETER SET UP MENU ...

Страница 461: ... parameter and press Set key to select it Refer to ILLUSTRATION 8 19 SYSTEM PARAMETER SET UP MENU 5 6 ILLUSTRATION 8 19 j Move the cursor to SAVE position and press Set button to save the selected parameters k Press the Top Menu Select key twice to exit the System Parameter Set Up menu 6 Make sure that the Video Cassette Recorder is properly controlled by the console For the VCR operation refer to...

Страница 462: ...N 1 OPTION TRAY 1 2127115 1 2 OPTION TRAY 2 2127116 1 3 VTR BASE P9524JY 1 4 VTR BELT P9524JT 1 5 STOPPER BLOCK P9524HH 1 6 PLATE P9524JN 2 7 CAUTION LABEL P9524PU 1 8 POWER CABLE ASSY P9509EE 1 9 RS232C CABLE ASSY P9509MN 1 used when installing VCR on console used when installing VCR on color monitor RENEWAL PARTS ILLUSTRATION 8 20 ...

Страница 463: ...rsion of the console is V3 30B or former V3 20B or former for LOGIQ 400CL In this case we have to use those printers with connecting a mini plug cable to enable the LOGIQ 400 console to control only the print function Other remote functions are not available in the case of the connection with the mini plug cable 8 3 2 Tools Required Standard field Service Tool kit 8 3 3 Time Required Approx 1 hour...

Страница 464: ...atic Discharge precau tions during this upgrade may cause constant or intermittent operational abnormalities Strictly follow all precautions Check each step as it is completed Perform functional checks to verify that the system is functioning properly Refer to Chapter 4 Functional Checks 8 3 6 Installing Color Video Printer on Console 1 Turn the system power OFF and unplug the unit Follow the OSHA...

Страница 465: ... TM 8 27 OPTIONS REV 0 8 3 6 Installing Color Video Printer on Console Continued 4 Put the OPTION TRAY 1 on the console and screw four screws supplied with CP FRONT ASSY 2127666 as shown in ILLUSTRATION 8 22 OPTION TRAY 1 INSTALLATION ILLUSTRATION 8 22 ...

Страница 466: ...LOGIQ 400 SERVICE MANUAL TM 8 28 OPTIONS REV 5 8 3 6 Installing Color Video Printer on Console Continued 5 Set the TRAY on the OPTION TRAY 1 and screw four screws as shown in ILLUSTRATION 8 23 TRAY INSTALLATION ILLUSTRATION 8 23 ...

Страница 467: ...Q 400 SERVICE MANUAL TM 8 29 OPTIONS REV 0 8 3 6 Installing Color Video Printer on Console Continued 6 Put the OPTION TRAY 3 on the TRAY and screw four screws as shown in ILLUSTRATION 8 24 OPTION TRAY 3 INSTALLATION ILLUSTRATION 8 24 ...

Страница 468: ...wo slits of the OPTION TRAY 3 as shown in ILLUSTRATION 8 25 VTR BELT ILLUSTRATION 8 25 8 Put the Color Video Printer onto the OPTION TRAY 3 and fix the Color Video Printer with the VTR BELT as shown in ILLUSTRATION 8 26 PATHWAY OF VTR BELT FOR UP 1850 SERIES PRINTERS PATHWAY OF VTR BELT FOR UP 2950 SERIES PRINTERS COLOR VIDEO PRINTER INSTALLATION ILLUSTRATION 8 26 ...

Страница 469: ...rding devices from the LOGIQ 400 If a peripheral device is supplied power by other outlet the system may be damaged by electrical noise transferred with the video signal or may not pass leakage current test 1 Connect the power cable RS232C cable and the BNC video cables between the Color Video Printer and console as shown in ILLUSTRATION 8 27 CONNECTING CABLES ILLUSTRATION 8 27 ...

Страница 470: ...e software version 3 30A or later and for the CL console with the software version 3 20A or later For the console with the other software version we have to use those printers with connecting a mini plug cable to enable the LOGIQ 400 console to control only the print function Other remote func tions are not available under this condition For the connection of the mini plug cable refer to ILLUSTRAT...

Страница 471: ...rinter Follow the procedures in Section 8 3 9 DIP SWITCH SETTINGS ILLUSTRATION 8 29 8 3 9 Setting Parameters of UP 2950 Series Printers This section provides the information for the required preset parameters of the UP 2950 series printers with the LOGIQ 400 console 1 Plug in the system and power the system ON 2 Power the Color Video Printer ON 3 Open the Front Door of the UP 2950 MD or UP 2850P p...

Страница 472: ...STRATION 8 32 LAYOUT SETUP c LAY p i o SETUP MENU OF PRINTER b ILLUSTRATION 8 32 6 Press the DOWN arrow key twice to display the MULTI PIX menu Refer to ILLUSTRATION 8 33 MULTI PIX 4 SETUP MENU OF PRINTER c ILLUSTRATION 8 33 7 Make sure that the 4 is selected for the MULTI PIX 4 is displayed on the LCD when the 4 is se lected Refer to ILLUSTRATION 8 33 Otherwise select the 4 using the LEFT arrow o...

Страница 473: ...se select the RGB using the LEFT arrow or RIGHT arrow key 14 Press the UP arrow key once to display the INPUT SETUP Top menu 15 Select the PRINTER setup menu using the LEFT arrow or RIGHT arrow key c l PRN i o is displayed on the LCD when the PRINTER setup menu is selected 16 Press the DOWN arrow key four times to display the SYSTEM SETUP menu Refer to ILLUSTRATION 8 36 SYSTEM SETUP PRESS SETUP ME...

Страница 474: ...LUSTRATION 8 39 Otherwise set the IMMED CAP to ON using the LEFT arrow or RIGHT arrow key 25 Press the MENU key to exit the setup menu 8 3 10 Operational Check out 1 Plug in the system and power the system ON 2 Power the Color Video Printer ON 3 Press New Patient key when the PATIENT ENTRY MENU is displayed on the CRT monitor screen of LOGIQ 400 4 Set the system so that the Color Video Printer can...

Страница 475: ...ed for the Color Printer Type parameter on the SYSTEM PARAMETER SETUP menu a Press the Top Menu Select key located on the keyboard The sub menu selected previously is displayed on the CRT monitor screen SOFTWARE MENU CONTROL KEYS ILLUSTRATION 8 40 b Press the Top Menu Select key again The top menu is displayed on CRT monitor screen as shown in ILLUSTRATION 8 41 TOP MENU ILLUSTRATION 8 41 c Select ...

Страница 476: ...LLUSTRATION 8 42 SET UP MENU ILLUSTRATION 8 42 e Select the System Paramtr section using the or key of Sub Menu Select keys The System Paramtr selection is displayed in reverse video as shown in ILLUSTRATION 8 42 f Press key to select the System Parameter Set Up Menu The system parameter set up menu is dis played on the CRT monitor screen as shown in ILLUSTRATION 8 43 SYSTEM PARAMETER SET UP MENU ...

Страница 477: ...D printer h Select the Black and White Printer from among Record1 B W parameters Refer to ILLUSTRATION 8 44 i Select the Color Printer from among Record1 Color parameters Refer to ILLUSTRATION 8 44 Note Use Ellipse keys to select the Record1 and Record2 parameters Other option is displayed by press ing the Ellipse keys when the current selection of Record1parameter or Record2 parameter is dis play...

Страница 478: ...d camera from among the Record1 Color parameter and the Col Pola from among the Plug 2 parameter n Move the cursor to SAVE position and press Set button to save the selected parameters o Press the Top Menu Select key twice to exit the System Parameter Set Up menu 5 Make sure that the Color Video Printer is properly controlled by the console Note If the UP 2950 series printer is connected with the ...

Страница 479: ... 41 OPTIONS REV 5 8 3 12 Renewal Parts ITEM PART NAME PART NO QTY DESCRIPTION 1 OPTION TRAY 1 2127115 1 2 OPTION TRAY 3 P9524ND 1 3 TRAY 2127118 1 4 VTR BELT P9524JT 1 5 POWER CABLE ASSY P9509EE 1 6 RS232C CABLE ASSY P9509MN 1 RENEWAL PARTS ILLUSTRATION 8 45 ...

Страница 480: ...9LH 4 Mini Plug Cable 1 P9509BC 8 4 5 Functional Check out READ and UNDERSTAND these instructions thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Possible operational damage ...

Страница 481: ...TRATION 8 46 Refer to the Sony B W Video Printer operator manual DIP SWITCH SETTINGS ILLUSTRATION 8 46 8 4 7 Installing B W Video Printer on Console 1 Turn the system power OFF and unplug the unit Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Unscrew four screws fixing the rubber foots of the B W Video Printer as shown in ILLUSTRATION 8 47 B W VIDEO PRINTER ILLU...

Страница 482: ... TM 8 44 OPTIONS REV 5 8 4 7 Installing B W Video Printer on Console Continued 3 Screw four screws supplied with the B W Printer Assy to install the Printer PL to the B W Video Printer Refer to ILLUSTRATION 8 48 PRINTER PL INSTALLATION ILLUSTRATION 8 48 ...

Страница 483: ...nstalling B W Video Printer on Console Continued 4 Connect three cables the power cable BNC BNC cable and the mini plug cable between the front CONN panel of the console and the rear panel of the B W Video Printer Refer to 8 4 8 Connecting Cables CONNECTING CABLES ILLUSTRATION 8 49 ...

Страница 484: ...ONS REV 5 8 4 7 Installing B W Video Printer on Console Continued 5 Insert the B W Video Printer with the Printer PL into the Printer Tray as shown in ILLUSTRATION 8 50 6 Screw two screws as shown in ILLUSTRATION 8 50 B W VIDEO PRINTER INSTALLATION ILLUSTRATION 8 50 ...

Страница 485: ...ding devices from the LOGIQ 400 If a peripheral device is supplied power by other outlet the system may be damaged by electrical noise transferred with the video signal or may not pass leakage current test 1 Connect the power cable mini plug cable and the BNC BNC cable between the B W Video Printer and the con sole as shown in ILLUSTRATION 8 51 CABLE CONNECTION ILLUSTRATION 8 51 ...

Страница 486: ...panel of LOGIQ 400 Refer to ILLUSTRATION 8 51 II Set that pressing Record1 key causes the B W video printer to print image including only B W data Image including color data is printed by color video printer III Select B W printer for printing report page IV Select inverse print for printing report page a Press the Top Menu Select key located on the keyboard The sub menu selected previously is dis...

Страница 487: ...LUSTRATION 8 54 SET UP MENU ILLUSTRATION 8 54 e Select the System Paramtr section using the or key of Sub Menu Select keys The System Paramtr selection is displayed in reverse video as shown in ILLUSTRATION 8 54 f Press key to select the System Parameter Set Up Menu The system parameter set up menu is dis played on the CRT monitor screen as shown in ILLUSTRATION 8 55 SYSTEM PARAMETER SET UP MENU 1...

Страница 488: ...ect the Record1 and Record2 parameters Other option is displayed by press ing the Ellipse keys when the current selection of Record1parameter or Record2 parameter is dis played in reverse video j Move the cursor to B W Prnt for the Report parameter and press Set to select it Refer to ILLUSTRATION 8 56 k Move the cursor to B W Prnt for the Plug1 BW parameter and press Set to select it Refer to ILLU...

Страница 489: ... Final Procedures 1 Properly dispose of excess material 2 This completes the installation of the B W Video Printer 8 4 11 Renewal Parts ITEM PART NAME PART NO QTY DESCRIPTION 1 PRINTER PL P9524BZ 1 2 POWER CABLE P9509LH 1 3 MINI PLUG CABLE P9509BC 1 RENEWAL PARTS ILLUSTRATION 8 57 ...

Страница 490: ...atient side 1 P9509KD 2 ECG Cable Assy 1 P9509KH For USA Europe and others ECG Cable system side 1 P9509KE ECG Cable patient side 1 P9509KF 3 PCG sensor 1 P9509KN 8 5 5 Functional Check out READ and UNDERSTAND these instructions thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the p...

Страница 491: ...the unit Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Remove the Side Cover Right Refer to 6 2 27 in Chapter 6 Renewal Parts 3 Unscrew four screws 1 4 and remove the CONN panel 2 Note Keep removed screws because they are required to mount the ECG panel assy onto console CONN PANEL 2 REMOVAL ILLUSTRATION 8 58 ...

Страница 492: ...talling ECG Board Assy Continued 4 Install the ECG Board Assy by screwing two screws 2 and 3 as shown in ILLUSTRATION 8 59 5 Connect the cable 1 CB22 from the Mother Board with the connector on the ECG board assy Refer to ILLUSTRATION 8 59 ECG BOARD ASSY INSTALLATION ILLUSTRATION 8 59 ...

Страница 493: ...l assy with the ECG board assy Refer to ILLUSTRATION 8 60 CONNECTING CABLES ILLUSTRATION 8 60 Note The additional 4 pin connector P69 is supplied with the ECG Board2 Assy for the reserved function Be sure to plug the CB68 cable for the AUX to the P68 connector on the ECG Board2 Assy Otherwise the Auxiliary function shall not be functioning properly ...

Страница 494: ...ERVICE MANUAL TM 8 56 OPTIONS REV 5 8 5 7 Installing ECG Panel Assy Continued 2 Screw four screws 1 4 to install the ECG panel assy as shown in ILLUSTRATION 8 61 INSTALLING ECG PANEL ASSY ILLUSTRATION 8 61 3 Re install the Side Cover Right ...

Страница 495: ...erational Check out 1 Connect the ECG patient cable with the connector on the ECG panel as shown in ILLUSTRATION 8 62 CONNECTING ECG PATIENT CABLE ILLUSTRATION 8 62 2 Connect the electrodes of ECG patient cable with the ECG simulator 3 Plug in the system and turn the system power ON ...

Страница 496: ... SOFTWARE MENU CONTROL KEYS ILLUSTRATION 8 63 6 Press the Top Menu Select key again The top menu is displayed on CRT monitor screen as shown in ILLUSTRATION 8 64 TOP MENU ILLUSTRATION 8 64 7 Select the ECG Menu using the or key of Sub Menu Select keys The ECG selection is displayed in reverse video as shown in ILLUSTRATION 8 64 8 Press key to select the ECG Menu The ECG Menu is displayed on the CR...

Страница 497: ...lect the ECG Wave mode using the or key of Sub Menu Select keys The ECG Wave selection is displayed in reverse video as shown in ILLUSTRATION 8 66 ECG MENU b ILLUSTRATION 8 66 10 Press key to select the ECG Wave 11 Check that the ECG waveform is displayed as shown in ILLUSTRATION 8 67 ECG WAVEFORM a ILLUSTRATION 8 67 ...

Страница 498: ... Sub Menu Select keys 15 Select the ECG Gain section using the or key of Sub Menu Select keys The ECG Gain section is displayed in reverse video 16 Check that the size of the ECG waveform changes by means of pressing the or key of Sub Menu Select keys 17 Check that the Heart Rate is displayed in XXXBPM at the lower right corner of the CRT monitor screen 18 Press the and keys of Sub Menu Select key...

Страница 499: ...en is highlighted and set the delay time 1 S1 to 100 msec The delay time 1 S1 and 2 S2 are displayed at the upper left position on the CRT monitor screen Refer to ILLUSTRATION 8 70 23 Select the Sync Selectn section using the or key of Sub Menu Select keys The Sync Selectn section is displayed in reverse video 24 Press key to set the delay time 2 S2 25 Rotate the trackball and set the delay time 2...

Страница 500: ... Wave section using the or key of Sub Menu Select keys 28 Press key to exit the ECG mode The ECG waveform will disappear 29 Connect the PCG sensor with the ECG panel Refer to ILLUSTRATION 8 71 Note If your customer does not need the PCG sensor or you have not any PCG sensors skip to step 42 CONNECTING PCG SENSOR ILLUSTRATION 8 71 ...

Страница 501: ...Sub Menu Select keys The PCG Wave section is displayed in reverse video as shown in ILLUSTRATION 8 72 ECG MENU e ILLUSTRATION 8 72 31 Press key to select the PCG Wave 32 The PCG waveform is displayed on the CRT monitor screen as shown in ILLUSTRATION 8 73 PCG WAVEFORM ILLUSTRATION 8 73 33 Check that the PCG waveform is displayed by means of patting the PCG sensor ...

Страница 502: ... 38 Check that the size of the PCG waveform changes by means of pressing the or key of Sub Menu Select keys 39 Press the and keys of Sub Menu Select keys simultaneously to display the first menu The ECG first menu is displayed on the CRT monitor screen 40 Select the PCG Wave section using the or key of Sub Menu Select keys 41 Press key to exit the PCG mode The PCG waveform will disappear 42 Connec...

Страница 503: ...out Continued 46 Make sure that the AUX waveform is displayed as shown in ILLUSTRATION 8 75 AUX WAVEFORM ILLUSTRATION 8 75 47 Press the and keys of Sub Menu Select keys simultaneously to display the third menu The ECG third menu is displayed as shown in ILLUSTRATION 8 76 ECG MENU g ILLUSTRATION 8 76 ...

Страница 504: ... section is dis played in reverse video 51 Check that the size of the AUX waveform changes by means of pressing the or key of Sub Menu Select keys 52 Press the and keys of Sub Menu Select keys simultaneously to display the first menu The ECG first menu is displayed on the CRT monitor screen 53 Select the AUX Wave section using the or key of Sub Menu Select keys 54 Press key to exit the AUX mode Th...

Страница 505: ... TM 8 67 OPTIONS REV 5 8 5 10 Final Procedures 1 Remove the Side Cover Left Refer to 6 2 26 in Chapter 6 Renewal Parts 2 Attach the Option Label onto the Nest Cover Refer to ILLUSTRATION 8 78 NEST COVER ILLUSTRATION 8 78 3 Re install the Side Cover Left ...

Страница 506: ...tor Card Refer to ILLUSTRATION 8 79 Mail the installation card Product Locator to the address corresponding to your region PRODUCT LOCATOR CARD ILLUSTRATION 8 79 Note The Product Locator Card indicated in ILLUSTRATION 8 79 may be different from the actually pro vided one 5 Properly dispose of excess material 6 This completes the installation of the LOGIQ 400 ECG Option ...

Страница 507: ...SY 2207806 1 2 ECG PANEL ASSY P9522XB 1 3 ECG CABLE System side P9509KC 1 For Japan ECG CABLE System side P9509KE 1 For USA Europe and others 4 ECG CABLE Patient side P9509KD 1 For Japan ECG CABLE Patient side P9509KF 1 For USA Europe and others 5 PCG SENSOR P9509KN 1 Old ECG Board Assy P9522WZ can be replaced with new ECG Board Assy 2207806 RENEWAL PARTS ILLUSTRATION 8 80 ...

Страница 508: ...2 Option Label 1 8 6 5 Functional Check out READ and UNDERSTAND these instructions thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Possible operational damage Failure to stri...

Страница 509: ...w the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Remove the Side Cover Left Refer to 6 2 26 in Chapter 6 Renewal Parts 3 Unscrew five screws and remove the Nest Cover as shown in ILLUSTRATION 8 81 4 Turn the upper and lower ejectors on slot 9 in the direction indicated by the arrow 1 Refer to ILLUSTRATION 8 81 5 Pull the TLMC Assy straightforwards in the direction indic...

Страница 510: ...SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL TM 8 72 OPTIONS REV 5 8 6 7 Installing OMEM Assy 1 Screw one screw 1 to assemble the stud onto the TLMC Assy as shown in ILLUSTRATION 8 82 ASSEMBLING STUD ILLUSTRATION 8 82 ...

Страница 511: ...ct the two connectors and screw one screw 1 to assemble the OMEM Assy onto the TLMC Assy as shown in ILLUSTRATION 8 83 ASSEMBLING OMEM ASSY ILLUSTRATION 8 83 3 Re install the TLMC Assy into the slot 9 of the Nest and turn the upper and lower ejectors to fix the board firmly 4 Screw five screws to install the Nest Cover ...

Страница 512: ...rea 3 Do the scanning for 30 seconds and press the Freeze button 4 Turn the Cine Scroll potentiometer clockwise 5 Check that the cine gauge appears at the bottom of the B mode image as shown in ILLUSTRATION 8 84 CINE MODE SCREEN ILLUSTRATION 8 84 6 Turn the Cine Scroll potentiometer counterclockwise to move the cursor indicating the current frame to the right most position of the cine gauge 7 Chec...

Страница 513: ...GIQ 400 SERVICE MANUAL TM 8 75 OPTIONS REV 5 8 6 9 Final Procedures 1 Turn the system power OFF 2 Attach the Option Label onto the Nest Cover OPTION LABEL ILLUSTRATION 8 85 3 Install the Side Cover Left Refer to 6 2 26 in Chapter 6 ...

Страница 514: ... ILLUSTRATION 8 86 Mail this installation Card Product Locator to the address corresponding to your pole PRODUCT LOCATOR INSTALLATION CARD ILLUSTRATION 8 86 Note The Product Locator Card indicated in ILLUSTRATION 8 86 may be different from the actually pro vided one 5 Properly dispose of excess material 6 This completes the installation of the Expanded Cine Memory Option for LOGIQ 400 ...

Страница 515: ...E MEDICAL SYSTEMS 2127661 LOGIQ 400 SERVICE MANUAL TM 8 77 OPTIONS REV 5 8 6 10 Renewal Parts ITEM PART NAME PART NO QTY DESCRIPTION 1 OMEM ASSY P9525WU 1 2 STUD U0053AG 1 RENEWAL PARTS ILLUSTRATION 8 87 ...

Страница 516: ...s thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Possible operational damage Failure to strictly follow ESD Electrostatic Discharge precau tions during this upgrade may caus...

Страница 517: ...nnecting Foot Switch 1 Turn the system power OFF Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Open the Rear Door 3 Connect the Foot Switch with the connector on the rear panel as shown in ILLUSTRATION 8 88 FOOT SWITCH CONNECTION ILLUSTRATION 8 88 ...

Страница 518: ...s the image displayed on the CRT monitor screen to freeze 3 Check that depressing the right switch of the Foot Switch causes the recording device assigned to Record 1 by system parameter set up to record the currently displayed image FOOT SWITCH ILLUSTRATION 8 89 8 7 8 Final Procedures 1 Properly dispose of excess material 2 This completes the installation of the LOGIQ 400 Foot Switch Option ...

Страница 519: ...s thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer CAUTION Possible operational damage Failure to strictly follow ESD Electrostatic Discharge precau tions during this upgrade may caus...

Страница 520: ...stalling Swivel Lock 1 Turn the system power OFF Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Unscrew inside two hexagonal bolts fixing the left side front caster to the frame as shown in ILLUSTRATION 8 90 HEXAGONAL BOLTS REMOVAL ILLUSTRATION 8 90 ...

Страница 521: ...27661 LOGIQ 400 SERVICE MANUAL TM 8 83 OPTIONS REV 5 8 8 6 Installing Swivel Lock Continued 3 Reinstall two hexagonal bolts to install the swivel lock as shown in ILLUSTRATION 8 91 SWIVEL LOCK INSTALLATION ILLUSTRATION 8 91 ...

Страница 522: ...lock lever on the front caster faces just forwards the swivel mecha nism is locked 3 Try to roll the LOGIQ 400 console and check that the left side front caster does not rotate 4 Push the back side of swivel lock down by foot 5 Try to roll the LOGIQ 400 console again and check that the swivel lock mechanism is released 8 8 8 Final Procedures 1 Properly dispose of excess material 2 This completes t...

Страница 523: ...ield Service Tool kit 8 9 3 Time Required Approx 30 Minutes 8 9 4 Parts Required ITEM PART NAME QTY PART REMARKS 1 Left Side Probe Holder 1 P9522XD 8 9 5 Installing Left Side Probe Holder 1 Turn the system power OFF 2 Unscrew three screws to remove the keyboard panel assy Refer to sub section 6 2 49 in Chapter 6 RENEWAL PARTS 3 Unscrew two screws and remove the box holder cover as shown in ILLUSTR...

Страница 524: ...TION 8 94 5 Re assemble the keyboard panel assy 8 9 6 Operational Check out 1 Turn the system power ON 2 Perform functional checks to verify that the keyboard panel assy is assembled correctly Refer to Chapter 4 FUNCTIONAL CHECKS 3 Check that the left side probe holder is installed firmly 8 9 7 Final Procedures 1 Properly dispose of excess material 2 This completes the installation of the optional...

Страница 525: ...DER INSTALLATION 8 10 1 Foreword This section describes the installation of the MTZ Probe Holder option for LOGIQ 400 8 10 2 Tools Required Standard field Service Tool kit 8 10 3 Time Required Approx 10 Minutes 8 10 4 Parts Required ITEM PART NAME QTY PART REMARKS 1 MTZ Probe Holder 1 P9522XM ...

Страница 526: ...ith a console except for the Japanese system 1 Install the MTZ probe holder as shown in ILLUSTRATION 8 95 MTZ PROBE HOLDER INSTALLATION ILLUSTRATION 8 95 2 Check that the MTZ probe holder is installed firmly 8 10 6 Final Procedures 1 Properly dispose of excess material 2 This completes the installation of the optional MTZ Probe Holder for LOGIQ 400 ...

Отзывы: