Quick Start 11
LOGIQ
t
200PRO Series Quick Start
2234974–100 Rev. 0
B-Mode Controls (cont’d)
B-Mode Optimization
Adjustments for...
Do the Following...
Adjustments for...
Do the Following...
Image too grainy
1.
Increase Dynamic Range.
2.
Increase Frame Average.
3.
Decrease Edge Enhance.
4.
Change Map.
Image too soft
1.
Decrease Dynamic Range.
2.
Increase Edge Enhance.
3.
Decrease Frame Average.
4.
Change Map.
Image too noisy
1.
Decrease B/M GAIN.
2.
Decrease Dynamic Range.
3.
Increase Frame Average.
4.
Increase Edge Enhance.
Improve Uniformity
1.
Increase the number of focal zones.
2.
Decrease Scan Width.
3.
Adjust TGC to compensate for
attenuation.
Cystic Imaging
1.
Decrease the B/M GAIN.
2.
Decrease Dynamic Range.
3.
Use Scan Width to reduce image
width.
4.
Change Combi Focus to increase
number of focal zones.
5.
Optimize focal zone placement.
Technically Difficult
Patients
1.
Select the proper probe.
2.
Maintain a lower Dynamic Range.
3.
Decrease Scan Width for faster
frame rates.