GE M
EDICAL
S
YSTEMS
D
IRECTION
2317229, R
EVISION
3
LOGIQ™ 180 B
ASIC
S
ERVICE
M
ANUAL
3 - 26
Section 3-6 - System Configuration
3-6-6
Video Specification
Video specifications may be needed to be able to connect laser cameras to the LOGIQ™ 180 .
General
SL
Parameter
NTSC
PAL
1
Total Number of Horizontal Lines
525[Frame]
625[Frame]
2
Vertical Field Frequency
60[Hz]
50[Hz]
3
Horizontal Scanning Frequency
15.734[kHz]
15.625[KHz]
4
Vertical Scanning Method
Interlaced
Interlaced
5
Sync Input
Composite
Composite
6
Pixel Clock
12.272715[MHz]
12.25[MHz]
7
Display width and Height [pixel * line]
608 X 464
608 X 472
Horizontal Timing & Video Amplitude
SL
Parameter
NTSC
PAL
1
Total H-Line Time
63.556[micro sec.]
64.00[micro sec.]
2
H-Sync Pulse Width
4.725[micro sec.]
4.68[micro sec.]
3
Back Porch
4.725[micro sec.]
4.68[micro sec.]
4
Total Active H-line time
52.64[micro sec.]
52.07[micro sec.]
5
Front Porch
1.46[micro sec.]
1.49[micro sec.]
6
Video Amplitude (back porch to peak)
0.7Vpp
0.7Vpp
7
Sync Amplitude (back porch to peak)
0.3Vpp
0.3Vpp
Vertical Timing
SL
Parameter
NTSC
PAL
1
Total V-Line Time
262.5[H]
312.5[H]
2
V-Sync Pulse Width
1st Field 3[lines]
1st Field 2.5[lines]
2nd Field 3[lines]
2nd Field 2.5[lines]
3
V-front porch
1st Field 7.5[lines]
1st Field 6.25[lines]
2nd Field 7.5[lines]
2nd Field 6.25[lines]
4
V-Blanking
1st Filed 30.5[lines]
1st Filed 32[lines]
2nd Field 30.5[lines]
2nd Field 32[lines]