DGA 900 Operator Guide
–
Rev 2.0
12-Apr-19
Page 20 of 92
Figure 6-5: Dashboard
The Last DGA Measurement panel shows the measurement date and time stamp, gas
levels (with measurements in PPM) displayed numerically and represented graphically on a
bar chart marked with the configured Caution (H for High) and Alarm (HH for High High)
levels. Blue signifies a reading within the normal limits, yellow a caution
—
the reading has
exceeded the High (H) level, and red an alarm
–
the reading has exceeded the High-High
level. Rest the pointer on any gas to obtain the full name in a tooltip. After the next
measurement completes, the details automatically update.
The System Info gives the current date, time, product state, scheduler mode (Normal,
Alarm, Caution) and next scheduled measurement. Indicators show the status of the alarm,
caution and service LEDs as well as the six relays. There is also a free text ID field to assign
a meaningful name to the product. The default is the product
’s
serial number.
Note: Only
Operator or Administrator changes will be saved.
The product
’s
serial and firmware version
numbers are also listed.