MA-025
–
DGA 900 Operator Guide
–
Rev 2.0
12-Apr-19
Page 81 of 92
First isolate the product through the external circuit breaker or external switch
and apply LOTO before opening the door.
Remove the eight M5 aerotight nuts that secure the Louvre catchment tray to the
top rear of the front door as shown in Figure C-7, and then lift the tray away from the
neoprene 3 mm gasket as shown in Figure C-8.
Remove the used white filter from the tray (it’s disposable and can be pulled out) as
shown in Figure C-9. Insert a new Louvre air filter as shown in Figure C-10 into the
tray ensuring that the filter evenly fills the entire space.
Refit the tray to the gasket on the front door and secure in place using the eight nuts
previously removed as shown in Figure C-7.
Figure C-7: Louvre catchment tray
–
attached
Figure C-8: Louvre gasket
–
tray removed
Figure C-9: Louvre catchment tray
–
detached &
filter removed
Figure C-10: Louvre air filter
C.3
Oil Filter Cleaning
The oil is filtered to prevent particles from entering the product or being returned to the
transformer. Therefore, this filter may need occasional cleaning.
Note:
A non-critical error in the product data file, such as error code 18 or 20 can be
suggestive of such action.
Figure C-11 shows the oil connections on the base of the Analysis module.