background image

16

Scanning Tips

• Minimum bed height should be 30” to achieve proper compression. The compression range indicator 

arrows on the articulating arm notify the user when the arm is at its lowest point and compression 
cannot be applied. 

• Always check to make sure a membrane has been replaced on the scanner assembly.
• Select proper depth to avoid tissue exclusion. Depth can be evaluated by visualizing the chest wall at 

the time of initial transducer placement. If the chest wall is not visualized a deeper depth should be 
selected.

• Keep the nipple inside the active scan area whenever possible.
• The active scan area should have as much skin contact as possible. 
• Breast tissue should be kept level without tilting or mounding.
• Avoid wedging the breast tissue, which causes the ultrasound beam to penetrate the tissue at an  

angle which may result in shadowing, artifact and false positives. 

• Begin by establishing optimal manual contact and compression of the breast, then activate compression 

assist while gradually releasing manual compression.

• Compression level 2 is recommended for most breast tissue. Once the scan has begun, If the transducer 

slides or lifts, abort the scan and decrease to a lower level of compression. 

• Manual compression is recommended for women who are not able to tolerate any levels of compression, 

such as women with fibrocystic breast tissue, islands of dense breast tissue or women with moderate 
to large size cysts/masses.

• If the breasts are large and tissue extends more than 15 cm from the nipple, it will be necessary to  

obtain views that do not include the nipple in order to ensure coverage of all the tissue. Additional 
views include: SUP, INF and UOQ or a second LAT.

• When the breasts are large, the operator obtains a view that includes as much tissue as possible with 

the nipple visible and then obtains a second volume of the same view farther in the direction that the 
breast tissue extends away from the nipple.

• The transducer should not slide, roll or lift during scanning.
• Remind the patient they can breathe normally but should refrain from talking or moving during the 

scan to avoid motion artifacts.

For a complete summary of Invenia ABUS Scan Station operation and positioning information, please 
refer to the Invenia ABUS Scan Station Basic User Manual (DOC. No. 4700-0014-00).

© 2014 General Electric Company – All rights reserved.

General Electric Company reserves the right to make changes in specifications and  
features shown herein, or discontinue the product described at any time without notice  
or obligation. Contact your GE Representative for the most current information.

GE, GE monogram, Invenia, and Reverse Curve are trademarks of General Electric  
Company.

GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC, a General Electric 
company, doing business as GE Healthcare.

GE Healthcare
447 Indio Way
Sunnyvale, CA 94085-4203
U.S.A.
www.gehealthcare.com

March 2014

4700-0017-00 Rev 01 

ULT-0566-02.14-EN-US

Europe 
GE Healthcare
 
Beethovenstr. 239 
D - 42655 Solingen 
T 49 212 2802 0 
F 49 212 2802 28

APAC
GE Healthcare Asia Pacific
 
4-7-127, Asahigaoka,  
Hino-shi, Tokyo
191-8503 Japan
T +81 42 585 5111

Содержание Invenia ABUS

Страница 1: ...1 Invenia ABUS Scan Station Positioning Guide ...

Страница 2: ...sion About this Guide This guide is for sonographers technologists and other trained clinical care providers using the Invenia ABUS Scan Station It describes techniques and procedures for obtaining optimal image quality during each breast exam The disposable membrane is intended for one time use and is changed for each patient Always check to make sure a membrane has been replaced on the scanner a...

Страница 3: ...he areola and nipple in a circular motion for 3 4 rotations using the three middle fingers to reduce the potential of air bubbles and associated nipple shadowing Avoid using a spoon or tongue depressor Lotion is re applied before each view to maintain coupling Note Ultrasound coupling lotion reduces the incidence of air bubbles which commonly cause artifact in conventional ultrasound gel Therefore...

Страница 4: ...ucer into the medial edge Transducer should be angled to follow the contour of the body Additional Views View Type Description Transducer Positioning Superior Superior tissue used when the AP view does not cover superior tissue Nipple will appear in the center at the inferior edge Lotion is applied from the clavicle to the nipple Shift the scanner assembly towards the patient s head so that the su...

Страница 5: ...illa Scanner assembly placement is shifted towards the axilla and laterally Bring the scanner assembly straight down on the breast flattening the tissue equally on all sides tilt superior then laterally making contact with the lateral edge of the breast Apply pressure snugging the transducer into the lateral edge The transducer should be angled to follow the contour of the body If breast shape and...

Страница 6: ...ue equally then apply pressure snugging the transducer into the medial edge Transducer should be angled to follow the contour of the body Superior View This view is used when the AP view does not cover superior tissue The nipple will appear in the center at the inferior edge Lotion is applied from the clavicle to the nipple Shift the scanner assembly towards the patient s head so that the superior...

Страница 7: ...ple will appear in the center at the superior edge of the active scan area Upper Outer Quadrant View This view includes superior and lateral tissue and also the axillary tail The nipple may not appear in the active scan area or will be at the lower inner edge A rolled towel can be placed behind the scapula to flatten out the armpit Lotion is applied to the upper outer quadrant of the breast extend...

Страница 8: ...t at each level Pressure can be released one level at a time using the decrease compression button on the right handle or red abort button on the left handle for quick release The breast should be compressed firmly flattening the tissue equally on all sides The transducer can be tilted medial or lateral to optimize contact Review the image on the touchscreen to make sure there is contact all acros...

Страница 9: ...e and tissue extends more than 15 cm from the nipple it will be necessary to obtain views that do not include the nipple in order to ensure coverage of all the tissue Additional views include SUP INF and UOQ or a second LAT When the breasts are large the operator obtains a view that includes as much tissue as possible with the nipple visible and then obtains a second volume of the same view farthe...

Отзывы: