4 Product Storage and Handling Requirements
4.1 Product Storage and Handling limits
NOTICE
Avoid extremes in temperatures
To obtain Temperature, Humidity or Pressure requirements for components not specified in
, refer to the appropriate component Pre–Installation Manual listed in
Innova LC System Compatibility
.
Table 1-4: Storage and Transport
TEMPERATURE
HUMIDITY (see Note (2))
PRESSURE
Component
MIN
MAX
MIN
MAX
MIN
MAX
)
-20°C (-4°F)
+40°C (+104°F)
10%
90%
79.4 kPa
106 kPa
All Monitors
-20°C (-4°F)
+55°C (+131°F)
10%
80%
79.4 kPa
106 kPa
All Injectors
-20°C (-4°F)
+60°C (+140°F)
10%
80%
79.4 kPa
106 kPa
All Footswitches
-20°C (-4°F)
+60°C (+140°F)
10%
90%
79.4 kPa
106 kPa
Vivid E9 Option
-40°C (-40°F)
+70°C (+158°F)
30%
60%
79.4 kPa
106 kPa
Cardiac Collimator
-20°C (-4°F)
+70°C (+158°F)
10%
95%
79.4 kPa
106 kPa
Diamentor
-20°C (-4°F)
+60°C (+140°F)
0%
85%
79.4 kPa
106 kPa
All other components
-40°C (-40°F)
+70°C (+158°F)
10%
95%
79.4 kPa
106 kPa
NOTE:
(1) The detector should be stored at 10 to 40 °C (50 to 104 °F) and less than or equal
to 90% RH in the plastic wrapped shipping box. (This should include two bags of
desiccant as well). The lowest temperature (e.g. 10 ºC (50 °F)) and humidity is
preferable. If they are to be stored outside of their shipping box or in the inner
shipping box without plastic wrapping they should be stored at 20 °C (68 °F) or less
and 30% RH or less. In terms of transportation, do not expose to temperatures below
–20 ºC (–4 °F) in its shipping box for more than 15 hours. The detector will reach the
ambient temperature after 20 to 25 hours. The detector should not be allowed to
reach temperatures less than –10 ºC (14 °F) or irreparable damage to the detectors
scintillator will occur. Care must be taken when removing a detector from a shipping
box. If the detector has been subject to cold temperatures for an extended period the
detector in the box should be allowed to sit in the plastic wrapped box to reach room
temperature. This will prevent condensation from occurring. Condensation on the
detector can cause irreparable damage to the electronics. Storage 10 to 40 ºC (50 to
104 °F); 10 to 90 % RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 °F)
and 10 to 80% RH. The Detector chiller is shipped within GEMS packaging.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 1 General Requirements
43