2.2.5 ECG Device Room Configurations
The hardware delivered with the ECG acquisition kit will need to be installed in the Control
Room and in Exam Room depending on the type of ECG Device used on the site.
The ECG Device/Room configurations need to be checked as this can impact ON the way parts
and cables can be installed.
Basically, there will be two different Installation Configurations depending on the type of ECG
device to connect to the Innova system:
2.2.5.1 Installation Configuration #1:
For ECG devices installed in Control Room (such as GE ECG like MacLab, CardioLab or
ComboLab)
Both the Physio & Hubican modules are installed in the Control Room as shown below:
Illustration 2-61:
2.2.5.2 Installation Configuration #2:
For ECG devices installed in Exam Room (ECG device from the competitor)
Physio and/or Hubican modules will be installed in Exam Room as shown below:
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 2 Equipment Requirements
121