5 Emergency
During an examination, any operator can encounter two main cases of failures.
5.1 Main power supply cut
In this case, refer to
.
5.2 System failure
In the event of a system failure with a patient on table during an examination, the operator can
utilize a Surgical Imaging mobile unit to finish the examination.
In this case a wall outlet single phase + ground is required to feed the mobile. It is also requires a
free space around the patient table to proceed with the mobile instead of Innova LC Positioner.
The table has to rotate to 90º. The minimum room width of 4400 mm (14.5 in) may not be sufficient
for this scenario.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 3 Physical Requirements of room
83