Chapter 2: Operation
Page 32
HygroPro User’s Manual
2.3.2 Entering Sensor Calibration Data (cont.)
Select the desired units and press
enter
.
Use the arrow keys to enter a value and
press
enter
.
Repeat the previous steps until all of your
Hygro Curve
data points
have been entered.
To save the data you have entered, scroll
to the bottom of the
Hygro Curve
options list to the
SAVE HYGRO TABLE?
section. Then, select
Save
and press
enter
.
IMPORTANT:
Any data not saved at the above prompt will be lost.
Repeat the above process to enter any available
Press Curve
and
CS Table
data points.
Note:
The
CS Table
is required only if ppm
w
measurements will be
made. Consult GE for the table values to use for your
application.