Chapter 1: Installation
Page 6
HygroPro User’s Manual
1.4.1 Standard Wiring Connections
The
HygroPro
is a 4-20 mA loop-powered device that can use the
same two wires for its measurement signal lines and its power supply
lines. Follow the instructions in this section only for units not
requiring hazardous (classified) area protection.
IMPORTANT:
To install the HygroPro in a hazardous (classified) area,
see
Hazardous Area Wiring Connections
on page 15.
The transmitter must be wired using the factory-supplied, 2-meter
cable. If a different cable length is required, please contact the factory
for assistance.
Note:
If a longer cable is required, you may splice an extension onto
the standard cable. Be sure to connect leads of the same color
together. For example, connect blue lead to blue lead, brown
lead to brown lead, etc.
The
HygroPro
may be connected to either of the following types of
external systems:
•
an external device that can provide the loop power to the
HygroPro
and can also receive and display the 4-20 mA
measurement output from the
HygroPro
(proceed to next page)
•
an external power supply to provide power to the
HygroPro
plus a
personal computer (PC) running GE’s
PanaView™
interface
software (proceed to page 13)
Содержание HygroPro
Страница 1: ...GE Measurement Control Moisture 916 099 Rev D November 2012 HygroPro User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page proceed to next page ...
Страница 24: ...Chapter 1 Installation Page 18 HygroPro User s Manual ...
Страница 40: ...Chapter 2 Operation Page 34 HygroPro User s Manual ...
Страница 48: ...Chapter 3 Service Maintenance Page 42 HygroPro User s Manual ...
Страница 58: ......
Страница 59: ......