RTX 1000H series user manual
K0297 Issue 2
4-6
4.4
Manual Configuration - LCD (Figure 4-5)
There are three items to set up for the optional LCD: the pressure units, the
decimal point position, and the display contrast.
Note:
The push buttons on the LCD module only work when it is
attached to the PCB.
1.
Set DIP switch 1 to OFF (write protect - OFF).
Figure 4-5: Selection sequence - LCD configuration
4.
UNITS: Use the S and Z buttons to step through the available units.
Wait 5 seconds or press S and Z together to go to the DECIMAL display.
5.
DECIMAL: Use the S and Z buttons to move the decimal position. There
is a maximum number of decimal places for each unit.
Wait 5 seconds or press S and Z together to go to the CONTRAST
display.
6.
CONTRAST: Use the S and Z buttons to adjust the contrast.
Wait 5 seconds or press S and Z together to return to normal operation.
7.
To prevent accidental changes to the new values, set DIP switch 1 to
ON.
2.
Press S and Z together:
3.
Press S: