Chapter 3 Installation
3-83
•
Logging Level:
Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
•
Logging Type:
Set the types of logging used by the RA 600 application. (Default: File)
-
File:
If this check box is checked, RA 600 will write general information about its operation to
the RadWorks.log file, which you can view in the Log folder.
-
Pipe:
Used in combination with the listening tool in the X:\AMI_61\bin folder. Among other
things, it enables you to observe DICOM transmissions in a pipe. Consult the technical
reference manual or contact technical support for additional information.
-
Console:
If this check box is checked, LOGIQworks will display a window that continuously
shows the actions LOGIQworks is performing. This is useful when troubleshooting.
•
Log File:
Set the name of the general log file. Each Service has its own default log filename. See
the Database Service, Connection Service and Print Service properties for details.
(Default: D:\LOG\AMI_61\RadWorks.LOG; DO NOT CHANGE THE DEFAULT SETTINGS!)
•
Truncation Size (kB):
Set the truncation size of the logging file. The truncation size is the size the
log file will be shortened to when its maximum size has been exceeded.(Default: 16 kB)
•
Maximum Size (kB):
Set the maximum size of the log file.
Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for
troubleshooting. (Default: 64 kB)
•
Always Flush File after Write:
If checked, all logging actions will be written immediately to the log
file. This option will have a serious impact on the performance of LOGIQworks, so it should not
normally be checked. (Default: not checked)
•
Log Pipe:
Name of the pipe where logging information is written.
(Default: \\.\pipe\AMI2\RadWorks\Logging)
3-5-2-4
Popup Menu
To select/deselect popup menu:
1.) Select
Configuration
>
Generic...
from the Data Selector menu.
2.) Select the Popup Menu tab from the Generic Properties dialog.