Section 8-5 - Back End Parts Replacement
8-5-11
CDRW Drive Replacement Procedure
8-5-11-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter.
When installing a new CDRW Drive on systems that have software versions earlier than 4.1, it is
necessary to install the CD-RW patch (P/N 2413448) that is supplied as part of the CD-RW Kit, as
shown below.
CD-RW Kit part number 2413449 consists of the following items:
8-5-11-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
.
8-5-11-3
CDRW Drive Removal Procedure
1) Remove the Vivid™ 4 left and right side covers, and rear cover - see the
on page 8 - 7 and the
Side Covers Installation Procedure
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP left and right side covers, as described in the
4) Remove the BEP front cover as described in the
BEP2 - Front Cover Removal Procedure
5) Remove the cables (data and power) that connect to the rear of the CDRW Drive.
6.) Carefully slide the CDRW Drive out from the front of the BEP as shown in
ITEM PART NO.
DESCRIPTION
2346017
CD-RW Drive
2413448
CD-RW Patch
2413450-100
CD Patch CD-RW Installation Instructions
Figure 8-51 CDRW Drive Removal