MASTR II ICOMs should be reset only when the fre-

quency shows deviation in excess of the following limits:

A.

±

 0.2 PPM, when the radio is at 27.5°C (81.5°F)

B.

±

 1 PPM at any other temperature within the range of -

30°C to + 85°C (-22°F to + 185°F)

If an adjustment is required, pry up the cover on the top

of the ICOM to expose the trimmer, and use one of the fol-
lowing procedures:

If the radio is at an ambient temperature of 27.5°C

(81.5°F), set the oscillator for the correct operating fre-
quency.

If the radio is not at an ambient temperature of 27.5°C,

setting errors can be minimized as follows:

A. To hold the setting error to  0.1 PPM (which is consid-

ered reasonable for 1 PPM ICOMs):

1.

Maintain the radio at 27.5°C (81.5°F) and set the
oscillator to desired frequency, or -

2.

Maintain the radio at 27.5°C ( 10°C) and offset the
oscillator, as a function of actual temperature, by
the amount shown in the correction factor chart
(Figure 2).

For example: Assume the ambient temperature of
the radio is 18.5°C (65.4°F). At that temperature,
the curve shows a correction factor of 0.44 PPM.
(At 851 MHz, 1 PPM is 851 Hz. At 875 MHz, 1
PPM is 875 Hz.)

With an operating frequency of 851 MHz, set the oscilla-

tor for a reading of 374 Hz (0.44 x 851 Hz) higher than the
licensed operating frequency. If a negative correction factor
is obtained (at temperatures above 27.5°C), set the oscillator
for the indicated PPM lower than the licensed operating fre-
quency.

TRANSMITTER ALIGNMENT

EQUIPMENT REQUIRED

1.

GE Test Set Model 4EX3A11 or Test kit 4EX8K12.

2.

A 50 ohm wattmeter capable of dissipating 150
Watts connected to the output terminal of the PA as-
sembly.

3.

A frequency counter.

PRELIMINARY CHECKS AND
ADJUSTMENTS

1.

Place ICOM on Exciter Board (crystal frequency =
operating frequency 

 48).

2.

For a large change in frequency or a badly mis-
aligned transmitter, preset all slugs to the top of the
coil form.

3.

Set Z101 and Z102 to top of coil form.

4.

Set both air variable capacitors to minimum capac-
ity (not meshed).

5.

Connect the red plug on the GE Test Set to the Sys-
tem Board metering jack, and the black plug to the
Exciter metering jack. Set the polarity to +, and set
the range to the Test 1 position (1 Volt position for
4EX8K12) for all adjustments.

All adjustments are made with the transmitter keyed.

Unkey the transmitter between steps to avoid unnecessary
heating.

Figure 2 - Correction Factor Chart

LBI-31985

LBI-31985

2

Содержание Ericsson MASTR II

Страница 1: ...multipliers to provide the station operating frequency The PA assembly uses 3 RF amplifier stages to drive a two transistor power output circuit Each stage employs matching networks consisting of discrete components and strip line technology The RF output power is fed to the an tenna through a circulator containing a low pass filter The PA assembly includes a power control module The power control...

Страница 2: ...or thereby preventing overmodu lation while preserving intelligibility TEST EQUIPMENT 1 An audio oscillator GE Model 4EX6A10 2 A frequency modulation monitor 3 Voltmeter 4 GE Test Set Model 4EX3A11 or 4EX8K12 5 Frequency Counter 6 Oscilloscope 7 50 ohm Wattmeter PROCEDURE 1 Connect the audio oscillator and the meter across audio input terminals J10 Green Hi and J11 Black Lo on GE Test Set and conn...

Страница 3: ...t 875 MHz 1 PPM is 875 Hz With an operating frequency of 851 MHz set the oscilla tor for a reading of 374 Hz 0 44 x 851 Hz higher than the licensed operating frequency If a negative correction factor is obtained at temperatures above 27 5 C set the oscillator for the indicated PPM lower than the licensed operating fre quency TRANSMITTER ALIGNMENT EQUIPMENT REQUIRED 1 GE Test Set Model 4EX3A11 or T...

Страница 4: ...er reading 2 C MULT 1 T102 and T103 See Procedure Tune T102 for maximum meter reading then tune T103 for a dip in meter reading 3 D MULT 2 T104 and T105 See Procedure Tune T104 for maximum meter reading and then tune T105 for a dip in meter reading 4 F MULT 3 T106 and C157 See Procedure Tune T106 for maximum meter reading and then tune C157 for a dip in meter reading 5 G MULT 4 C159 and Z101 See P...

Страница 5: ...103 T101 Q102 Q103 T101 ICOM Q102 Q103 T101 ICOM C MULT 1 Q104 T103 T101 T102 Q104 T101 T102 Q104 T103 D MULT 2 Q105 T105 T103 T104 Q105 T103 T104 Q105 T105 F MULT 3 Q106 C155 Q106 T105 T106 Q106 T105 T106 L105 G MULT 4 Q107 Z101 Z102 Q107 C159 C157 Q107 C159 C157 Z101 Z102 A REL OUTPUT Q107 Q107 C174 CR102 R154 R153 851 870 MHz 100 WATT TRANSMITTER Issue 1 LBI 31985 LBI 31985 4 ...

Страница 6: ...This page intentionally left blank LBI 31985 LBI 31985 5 ...

Отзывы: