Programming
9 programs each with up to 9 phases
Regulation
Maximum voltage, current and power with automatic
crossover at preset limits
Output range
Voltage: 0–3 500 V DC,
Current: 0–400 mA
Power: 0–200 W
Programming range
Voltage: 35–3 500 V DC
Current: 1–400 mA
Power: 1–200 W
Time: 00:01–500 h
Volthour: 1–500 000 Vh
Amperehour: 1–25 000 mAh
Output resolution
Voltage: 1 V
Current:
1 μA, 0–399 μA
10 μA, 0.4–3.99 mA
100 μA, 4.0–19.9 mA
1 mA, 20–400 mA
Power: 1 W
Programming resolution
Voltage: 5 V
Current: 1 mA
Power: 1 W
Time:
1 min, 00:01–99:59,
1 h, 100–500 h
Volthour:
1 Vh, 1–9999 Vh,
100 Vh, 10.0–99.9 kVh
1 kVh, 100–500 kVh
Amperehour: 1 mAh, 1–999 mAh
10 mAh, 1.00–9.99 Ah
100 mAh, 10.0–25.0 Ah
Accuracy
Voltage: 1%, ± 5 V
Current: 1%, ± 1 mA
Power: 2%, ± 1 W
Timer: 0.1% ± 1 min, 00:01–99:59 h
0.1%, ± 10 min, 100–500 h
Line regulation
< 0.2%
Load regulation
< 1% at load change 10–90% of maximum load
Ripple
< 1% at 3 500 V
Short term stability
< 0.2% /10 h after warm up
Long term stability
< 1% /year
Start current check
Resistance not greater than 5 M at 40 V (Current less than
8 μA). Can be disabled
Sudden load change check
Ground leakage check
Leakage not greater than 500 μA
Output protection
Fully protected against any overload conditions
Recovery after power failure
Duration < 8 s,
The program continues automatically
Duration
≥
8 s, The program continues after manual restart
8 Technical specifications