User Manual
Druck DPI612
Page 89 of 105
[EN] English – 109M4017 Revision -
•
Apply the values that follow:
mA: -55, -25, -20, -10, -5, 0 (open circuit)
mA: 0, 5, 10, 20, 25, 55.
•
Check the error is in the specified limits (See
Table 5-2 Current (measure) error limits
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI612 error (mA)
±55
0.0030
0.0057
±25
0.0025
0.0043
±20
0.00063
0.0022
±10
0.00036
0.0016
±5
0.00025
0.0013
0 (open circuit)
0.0002
0.0010
PROCEDURES: CURRENT (source)
5.6
When recalibrating measure and source functions for this range, the
measure function must be adjusted before the source function.
Do the procedure as follows:
1.
Connect the applicable calibration equipment
(See
•
CH1 (24 mA range): (See Table 5-3).
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Use the calibration menu (
See Section 5.4
calibration (0.2 mA and FS):
•
CH1: 24 mA.
4.
Check the calibration is correct:
•
Select the applicable Current (source) function
•
Apply the values that follow:
mA:
0.2, 6, 12, 18, 24
•
Check the error is within limits (
See Table 5-3