2.3.6
Release Hydraulic Pressure (1000 bar pump)
1
2
Open the refill valve
fully counter-
clockwise
Open the pressure
release valve
counter-clockwise
(1 turn)
2.3.7
Drain Excess Hydraulic Fluid
If you add more hydraulic fluid during the pressure procedure, drain
this fluid out of the device when the pressure procedure is complete.
Note:
If it is safe and there is no risk of contamination, you can leave
the hydraulic fluid inside the device.
Preparation
To drain the device, we recommend these items:
•
the applicable skin and eye protection
•
a container that is large enough to hold the hydraulic fluid and
prevent contamination of the work surface
•
applicable materials to make sure the instrument and the area
stay clean; see Chapter 6 (Maintenance procedures)
Procedure
1
Release the pressure (Section 2.3.6Release Hydraulic pressure
2
Remove the device (Section2.1.1) but do not let fluid spill onto the
DPI 612 hFlexPro.
3
If necessary, drain the hydraulic fluid from the device under test.