[EN] English -
7
K405 Issue 3
Operation - Measure/supply mA
To measure/supply a current:
1. Connect the instrument (Figure 1, 2 or 3) and, if
necessary, adjust the
Set Up
(Table 1).
2. Select the task from the task selection menu
Note: Use the Dual Function area (
➁
) to do two operations
at the same time. If the area
➁
selection is not necessary,
set this area to off
(
■
)
. This saves the battery power.
3. If necessary, adjust the
Settings
(Table 4/5) and/or the
output values to the system (Table 7).
Figure 1:
Example configuration - To measure/supply mA
with external loop power (Area
➀
)
Figure 2:
Example configuration - To supply mA with internal
loop power (Area
➀
)
Figure 3:
Example configuration - To measure mA (Dual
Function, area
➁
)
Edit
Settings
Maximum: 30V
a) Measure
b) Supply
Edit
Settings
a) with external loop
power
b) with internal loop
power
Maximum: 30V