K405 Issue 3
To start - Prepare the instrument
Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument,
and that there are no missing items.
• Remove the plastic film that protects the display. Use
the tag (
◗
) in the top right-hand corner.
• Install the batteries (refer to B1). Then re-attach the
cover.
To start - Power on or off
To set the power on or off, press
❍
(A1 - item 1). The
instrument does a self test and then shows the applicable
data.
When the power is off, the last set of configuration options
stays in memory. Refer to “Maintenance”.
To start - Set up the basic operation
Use the
Set Up
menu to set up the basic operation of the
instrument.
If there is additional data for a menu option, select
Settings
(
■
■
) to see the values that are set up. If
necessary, adjust the values.
Table 1:
Menu options - Set Up
To start - Select a task (Measure and/or supply)
When the instrument is set up (Table 1), use the task
selection menu to select the applicable task.
If you attach a Universal Measurement Module (UMM) to
the communications port (A1 - item 9), the task selection
menu shows the applicable IDOS options.
Make the necessary selections from each area (
➀
and
➁
).
One task is permitted in each area.
Note: Use the Dual Function area (
➁
) to do two operations
at the same time. If the area
➁
selection is not necessary,
set this area to off
(
■
)
. This saves the battery power.
Table 2:
Menu options - Task selections (Area
➀
)
1
2
Task selection
menu:
3
Menu:
Set Up
4
5
▲
▼
+
■
■
▲
▼
(Table 2)
[
✓
]/[ ]
Options
Description
... Scale To select the applicable international temperature
scale: IPTS 68 or ITS 90.
To add a 250
Ω
series resistor into the mA circuit.
You can then use this instrument together with a
HART
®
communicator to set up and calibrate
HART
®
devices.
To select and set up the backlight fa timer.
Additional data: Select Settings
(
■
■
)
To select and set up the power off fa timer.
Additional data: Select Settings
(
■
■
)
To show the battery level (%).
To set the display contrast (%).
▲
Increases %,
▼
decreases %
To set the time + date. The calibration facility uses
the date to give service and calibration messages.
To set the language option.
To calibrate the instrument.
Additional data: Refer to “Calibration”.
To select and show the applicable status data.
(Software Build, Calibration Due date, Serial
Number, IDOS Information).
1
Task selection
menu:
2
3
4
5
▲
▼
+
■
■
▲
▼
+
■
■
Task = mA output
Task selection
menu:
6
Display:
mA output
Sk1 = Edit
Sk2 = Settings
Options
(If applicable)
Description
Input measurement tasks:
mA Measure ±55 mA
V Measure ±30V
mV Measure ±120mV
Hz Measure the frequency (Units: Table 4)
RTD Measure RTD temperature
Ω
Measure RTD resistance or
Ω
TC Measure thermocouple temperature OR mV
Only when an IDOS UMM is attached. An IDOS
measurement task.
Output tasks:
mA Supply 0 to 24 mA
V Supply 0 to 12V
mV Supply 0 to 120mV
Hz Supply an output frequency (Units: Table 4)
RTD Simulate RTD temperature
Ω
Simulate RTD resistance or
Ω
TC Simulate thermocouple temperature OR mV