Calibration - Procedures: TC (mV) input/output
1. Connect the instrument to the calibration equipment:
• TC (mV) input = Figure 9b
• TC (mV) output = Figure 9d
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do the calibration:
• TC (mV) input = three-point calibration
(-FS, Zero and +FS).
• TC (mV) output = two-point calibration (Zero and +FS).
The display shows the applicable instructions to
complete the calibration.
4. To make sure that the calibration is correct, select the
applicable TC (mV) input or output task (Table 2) and
apply the necessary values:
• TC (mV) input: -10, 0 (short circuit)
Then TC (mV): 25, 50, 75
• TC (mV) output: -10, 0, 25, 50, 75
5. Make sure that the error is in the specified limits
Table 20:
TC (mV) input/output error limits
Calibration - Procedures: IDOS UMM
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Specification
All accuracy statements include one year stability.
Specification - General
Specification - Electrical (A1 - Item 10)
Specification - Electrical connectors (A2)
Input
or output
Calibrator error
TC (mV)
Permitted DPI 880 error
TC (mV)
TC (mV)
mV
mV
mV
mV
-10
0.000 5
0.000 18
0.008
0.008
0
-
0.000 05
0.006
0.006
25
0.000 6
0.000 36
0.010
0.010
50
0.000 8
0.000 68
0.014
0.014
75
0.001 0
0.000 99
0.018
0.018
Languages
English [Default]
Operating
temperature
14 to 122°F (-10 to 50°C)
Storage
temperature
-4 to 158°F (-20 to 70°C)
Humidity
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration
BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
EMC
BS EN 61326-1:1998 + A2:2001
Safety
Electrical - BS EN 61010:2001; CE Marked
Size (L: W: H)
7.1 x 3.3 x 2.0 in
(180 x 85 x 50 mm)
Weight
15 oz (425 g)
Power supply
3 x AA alkaline batteries
Duration
Measure functions (area
➀
):
≈
60 hours
Dual Function, mA measure (area
➁
):
≈
7 hours (24 V Source at 12 mA)
Range (Measure):
0 to ±55 mA
0 to 4000
Ω
*
*
Refer to “Specification - Resistance ranges (Ohms/RTD)”
0 to ±120 mV
0 to ±30 V
Accuracy: Measure mA
0.02% of r 3 counts
Accuracy: Measure mV
0.02% of r 2 counts
Accuracy: Measure V
0.03% of r 2 counts
Range (Supply):
0 to 24 mA
0 to 4000
Ω
*
0 to 120 mV
0 to 12 V
Accuracy (Supply):
mA, mV, V
0.02% of r 2 counts
Temperature coefficient
(Measure or supply)
14 to 50°F, 86 to 122°F
(-10 to 10°C, 30 to 50°C)
0.0017% FS / °F
(0.003% FS / °C)
Connectors (A1 - Item 10)
Four 0.16 in (4 mm) sockets
One TC mini-connector socket
Range (Measure)
0 to ±55 mA
0 to ±30 V
Accuracy: Measure mA
0.02% of r 3 counts
Accuracy: Measure V
0.03% of r 2 counts
Temperature coefficient
14 to 50°F, 86 to 122°F
(-10 to 10°C, 30 to 50°C)
0.0017% FS / °F
(0.003% FS / °C)
Switch detection
Open and closed. 2 mA current.
Loop power output
24 V ± 10% (Maximum: 35 mA)
HART
®
resistor
250
Ω
Connectors (A2)
Three 0.16 in (4 mm) sockets