DPI 740 User Manual
31
K0200 Issue No. 1
Entering a new PIN
The instrument calibration is protected by a PIN, which must
be entered before accessing the calibration menu. The default
PIN set at manufacture is 000.
Changing the PIN
Proceed as follows:
1. Make sure the instrument is in measurement mode and
displaying pressure.
2. Press SET until the display shows PIN on the status line.
3. Press F2 to change the PIN.
4. Using F1
, F2
and SET to enter the current PIN.
5. Using F1
, F2
and SET to enter the new PIN.
6. Using F1
, F2
and SET to enter the new PIN again. The
instrument compares the first new PIN with the second
new PIN to verify the entry.
7. If verified successfully the display briefly shows New PIN
Accepted on the status line.
Note:
If not verified the display briefly shows Verify Failure on
the status line.