Specification
Dimensions
Weight: (nominal).................................................................. ..................0.5 kg
Size:
height = 190 mm width = 90 mm, depth = 36 mm
Environmental
Temperature:
Operating .......................... ........
......... ......... -10 to +50
C
Storage ......................................................... ......... -40 to +70
C
Pressure medium: ................ Any gas compatible with pyrex,
.................. .................. silicon, stainless steel and epoxy resin
Sealing standard................................................................... ..................IP54
Pressure
Connector...................................................... ......... 6 mm o/d
or.................................................... ......... 4 mm i/d hose fitting
Accuracy
Non-linearity, hysteresis and repeatability:
over the range of 10
°
to 30
°
C
........................................................................... ......... ±0.02 % FS
over the range of 0° to 40°C
........................................................................... ......... ±0.03 % FS
over the range of -10° to 50°C
........................................................................... .........±0.045 % FS
Stability .................................................................. ........ <100 ppm/year
Range (barometric)............................................ ........ 750 to 1150 mbar a
alternative ranges:
.................................................................. ........ 35 to 1300 mbar a
.................................................................. ........ 35 to 2600 mbar a
.................................................................. ........ 35 to 3500 mbar a
Maximum safe working pressure........................ ......... 4375 mbar a
Display
Read-out................................... .........999999 LCD digits, 13.6mm high
.................. .................. with additional 16 text characters
Overload indication
Error code flashing................. .........at nominal 110% FS
Response............................................ ........ nominal 2 readings/second
Resolution
0.01 mbar (e.g., 1013.25 mbar a)
Содержание Druck DPI 740
Страница 1: ...GE Sensing Druck DPI 740 Precision Pressure Indicator User Manual K0200 ...
Страница 2: ... The General Electric Company All rights reserved ...
Страница 16: ...K0200 Issue No 1 8 DPI 740 UserManual Figure 2 2 Electrical connections ...
Страница 32: ...K0200 Issue No 1 24 DPI 740 User Manual intentionally left blank ...
Страница 46: ...K0200 Issue No 1 38 DPI740 UserManual ...
Страница 72: ...K0200 Issue No 1 64 DPI 740 UserManual Approved Service Agents www gesensing com ...