Keypad Display
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 73
9.
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the
Pt remains unchanged
*
.
10.
SETUP
, [UNITS] can be used to change the units of display, if setup mode is not
switched off.
11. Before entering a new CAS or Qc aim, the rate of change should be set using the
RATE
key.
12. In Control Mode (AIM display), a new CAS or Qc can be entered using the numeric
keys. Use
LEAK MEASURE/CONTROL
to change between control and measure
mode.
7.2.4 MACH/Pt
1.
In aeronautical units (ft/kts or m/km/h), this key selects a Mach display.
2.
In pressure units (mbar, inHg etc.), this selects a total pressure (Pt) display.
3.
If the altitude aim is changed, after the entry of CAS, the Mach and TAS aim values
change while the CAS value remains unchanged*. If the altitude aim is changed, after
the entry of TAS or Mach, the CAS aim value changes while the TAS and Mach values
remain unchanged
*
.
4.
If the Ps aim is changed, after the entry of Qc, then the Pt aim value changes while the
Qc remains unchanged
*
.
5.
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the
Pt remains unchanged
*
.
6.
Before entering a new Mach or Pt aim, set the rate of change of speed using the
SPEED/Qc
key.
7.
Press the
RATE
key.
8.
In Control Mode (AIM display), a new Mach or Pt aim can be entered using the numeric
keys, Use
LEAK MEASURE/CONTROL
to change between control and measure
mode.
7.2.5 EPR
The ADTS 405 Mk2 can be used to check EPR sensors and indicators.
1.
Engine Pressure Ratio (EPR) is only available in pressure units (mbar, inHg etc.).
2.
For EPR testing, use Ps as INLET pressure and EPR to set the ratio of OUTLET pressure
(Pt) to INLET pressure.
3.
Before entering a new EPR aim, the rate of change should be set using the
RATE
key.
4.
In Control Mode (Aim display), a new EPR aim can be entered using the numeric keys.
Use
LEAK MEASURE/CONTROL
to enter measure mode and make the appropriate
Ps and Pt control state changes.
*
The display shows = symbol next to the parameter that remains unchanged.
Содержание Druck ADTS 405 F Mk2
Страница 2: ......
Страница 16: ...xiv ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 18: ...xvi ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 118: ...Chapter 7 Reference 100 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 134: ...Appendix A Menus 116 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 139: ......