This chapter describes the Documenting functions
available with the DPI620G-IS calibrator and are as
follows:
•
ANALYSIS
•
RUN PROCEDURE
5.1
Analysis
The Analysis function takes readings from two or more
DPI620G-IS channels to calibrate the transfer
characteristic of the device being tested. One channel is
the Reference channel and is used as follows:
•
Provides a measure of the input signal to the device.
•
If calibrating a temperature transmitter the
Reference channel could be CH1 in either RTD or TC
source mode.
•
If the device is a pressure transmitter the Reference
channel would be P1 or P2 measuring the input
pressure to the device.
The other channel is the Input channel and is used as
follows:
•
Measures the output signal from the device
•
If calibrating a process transmitter it could be CH2 in
Current Measure mode.