10.9
Calibration Procedure (CH1): Resistance (measure)
(standard)
Perform the calibration as follows:
1.
Connect the applicable calibration equipment
specified in Table 10-1; for example see
Section 2.11
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref:
Section 10.2
) to do a
two-point calibration for each of the following ranges:
Range: 0-400Ω
•
Nominal zero ohms:
Make a 4-wire connection to the 0Ω resistor.
•
Nominal positive full-scale ohms:
Make a 4-wire connection to the 400 Ω resistor
Range: 400
Ω
-4k
Ω
•
Nominal 400Ω: Make a 4-wire connection to the
400Ω resistor
•
Nominal positive full-scale ohms: Make a 4-wire
connection to the 4k resistor
4.
Check the calibration is correct
•
Select the Resistance (measure) function (Ref
Section
).
•
Make a 4-wire connection to the applicable standard
resistor (Ref: Table 10-11) and measure the value.
•
Check the error is within the specified limits (Ref:
Table 10-11
.