GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 7 - Gantry
Page 227
7.5 PET Gantry Replacement Procedures
For repair procedures for the Advance PET gantry components:
Refer to the
Advance PET
Imaging Repair Procedures, Direction 2100323.
7.5.1 PET Gantry Drive Mechanism Replacement (P/N 2290248)
The PET Gantry drive mechanism is shown in
Figure 7-3:
PET Gantry Drive Mechanism
Note:
The sub-assemblies of the drive mechanism are shown in
and
.
To replace the PET gantry drive mechanism:
1.) Disassemble the old drive mechanism by loosening the 4x M6 screws on the gantry back
beam, and removing the back and front bearing bases (
).
Figure 7-4:
Driving Mechanism – Internal Assembly
Tracking trapezoidal
Shaft for Electric Drill
screw drives PET Gantry
to and from service
position.
Nut Bearing Housing
Retaining
Ring
Nut Rotation Stopper
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......