March 2008
1-8
Installation
Wiring the Line Power
(cont.)
1.
Remove the plastic shroud that covers the terminal blocks. Be sure
to reinstall the shroud after all of the wiring has been completed.
2.
Strip ¼” of insulation from the end of the power and neutral or line
leads (or the positive and negative DC power leads), and ½” from
the end of the ground lead.
3.
Connect the ground lead to the internal ground connection located
on the side panel of the inclosure (see Figure 1-3 below).
IMPORTANT:
The incoming ground lead must be connected to the
internal ground connection.
4.
Connect the neutral or line lead (or the negative – DC power lead)
to
TB1-2
and the line power lead (or the po DC power lead)
to
TB1-3
as shown in Figure 1-3 below.
IMPORTANT:
Do not remove the existing PC board ground wire or
the cover ground wire.
Figure 1-3: Wiring the Line Power
1
TB1
N/L2
G
ro
u
n
d
2
3
L1
G
Internal
Ground
Connection
PC Board
Ground Wire
Cover
Ground Wire
G – +
AC
DC
AC Line Power
or DC +
AC Neutral/Line
or DC –
Содержание DigitalFlow GS868
Страница 6: ...Chapter 1...
Страница 28: ...Chapter 2...
Страница 40: ...Chapter 3...
Страница 41: ...Operation Introduction 3 1 Powering Up 3 2 Using the Display 3 3 Taking Measurements 3 5...
Страница 49: ...Chapter 4...
Страница 50: ...Specifications General 4 1 Electrical 4 2 Operational 4 4 Transducer 4 4 Flowcell 4 5...
Страница 56: ...Appendix A...
Страница 57: ...CE Mark Compliance Introduction A 1 Wiring A 1 External Grounding A 1...
Страница 59: ...Appendix B...
Страница 60: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...
Страница 64: ...Appendix C...
Страница 65: ...Optional Enclosures Introduction C 1 Rack Mount Enclosure C 1 Rack Mount Wiring C 1 Rack Mount Front Panel C 2...
Страница 71: ...Appendix D...
Страница 72: ...Measuring P and L Dimensions Introduction D 1 Measuring P and L D 1...