3-10
Dash Port Docking Station
Revision B
2000966-037
Maintenance: Electrical Safety Tests
Ground (Earth) Wire Leakage Current Tests
Perform this test to measure current leakage through the ground (earth)
wire of the equipment during normal operation.
1.
Set the leakage tester switches as follows:
Selector knob - 1,
GND switch - OPEN,
Polarity switch - NORM,
Power switch - OFF.
2.
Connect the DMM to the METER jacks on the leakage tester. Set the
DMM to measure AC millivolts.
3.
Connect the power cord of the device under test to the power
receptacle on the rear of the leakage tester.
127(
The device under test is to be tested at its normal operating
voltage.
4.
Set the leakage tester power switch to ON.
5.
Set the power switch of the device under test to ON.
6.
Read the current leakage indicated on DMM. If the reading is greater
than the appropriate specification below, the device under test fails
and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under
test is powered from 100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is
powered from a centered-tapped 200-240 V/50-60 Hz, single
phase circuit
500 µA (0.5 volts on the DMM), and the device under test is
powered from a non-center-tapped, 200-240 V/50-60 Hz, single-
phase circuit.
127(
Center-tapped and non-center-tapped circuits produce different
leakage currents and the UL and IEC limits are different.
7.
Set the polarity switch on the leakage tester to RVS (reverse).
8.
Read the current leakage indicated on DMM. If the reading is greater
than the appropriate specification below, the device under test fails
and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under
test is powered from 100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is
powered from a centered-tapped 200-240 V/50-60 Hz, single
phase circuit
500 µA (0.5 volts on the DMM), and the device under test is
powered from a non-center-tapped, 200-240 V/50-60 Hz, single-
phase circuit
Содержание Dash D 0 0005 G Series
Страница 1: ...Dash Port Docking Station Service Manual 2000966 037 Revision B ...
Страница 5: ...Revision B Dash Port Docking Station 1 1 2000966 037 1 Introduction ...
Страница 6: ...1 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 11: ...Revision B Dash Port Docking Station 2 1 2000966 037 2 Equipment Overview ...
Страница 12: ...2 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 17: ...Revision B Dash Port Docking Station 3 1 2000966 037 3 Maintenance ...
Страница 18: ...3 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 36: ...3 20 Dash Port Docking Station Revision B 2000966 037 Maintenance Checkout Procedures For your notes ...
Страница 37: ...Revision B Dash Port Docking Station 4 1 2000966 037 4 Troubleshooting ...
Страница 38: ...4 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 46: ...4 10 Dash Port Docking Station Revision B 2000966 037 Troubleshooting General For your notes ...
Страница 47: ...Revision B Dash Port Docking Station 5 1 2000966 037 5 Field Replaceable Units ...
Страница 48: ...5 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 53: ...Revision B Dash Port Docking Station 6 1 2000966 037 6 Drawing and Part List ...
Страница 54: ...6 2 Dash Port Docking Station Revision B 2000966 037 For your notes ...
Страница 59: ......