Monitoring Invasive Pressures
2000966-386D
Dash™ 3000/4000/5000
9-7
Performing PA Wedge Measurements
Automatic
To perform an automatic PA wedge measurement, follow this procedure.
1.
Select the
PA
parameter window label.
2.
From the
PA
menu, select
PA WEDGE
. A single beep sounds and the message
“
INFLATE BALLOON
” displays in the
PA
parameter window.
3.
Inflate the balloon according to your unit’s policy and procedure. The message
“
WEDGE PROCESSING
” displays.
4.
To add a cursor to the real-time wedged waveform, select
PA CURSOR
.
5.
Deflate the balloon after three or four respiratory cycles (no more than 20
seconds).
The
PA WEDGE REVIEW
information window displays 20 seconds of
compressed waveform data with horizontal and vertical cursors identifying the
end-exhalation point, PAW value and time of measurement.
705B
6.
To change the calculated PA wedge value, select
MOVE WEDGE CURSOR
.
Adjust the value and select main display to save the new cursor value.
7.
To save this PA wedge value in the cardiac calculations and
VITAL SIGNS
,
select
CARDIAC OUTPUT
.
NOTE
To change the displayed waveform sweep speed, select
SPEED
and select an
option.
Содержание Dash 3000
Страница 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Страница 12: ...x Dash 3000 4000 5000 2000966 386D ...
Страница 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Страница 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Страница 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Страница 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Страница 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Страница 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Страница 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Страница 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Страница 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Страница 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Страница 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Страница 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Страница 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Страница 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Страница 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Страница 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Страница 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Страница 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Страница 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Страница 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Страница 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Страница 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Страница 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Страница 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Страница 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Страница 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Страница 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Страница 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Страница 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Страница 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Страница 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Страница 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Страница 269: ......