Monitoring Impedance Cardiography (ICG)
2000966-386D
Dash™ 3000/4000/5000
17-5
13. Verify that a good quality waveform displays.
14. Adjust the control settings as necessary.
Control Settings
To adjust the control settings for the patient admitted to the monitor, select the
ICG
parameter window label. Then select any of the following options and make the
desired changes.
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Option
Function
ICG PATIENT
INFORMATION
Select the required patient information. When monitoring the
following parameters additional information is required:
MAP
— Select
NBP
or
Arterial Line
.
CVP
— Select invasive CVP or enter a specific value
for SVR and SVRI calculations.
PAW
— Select invasive PAW or enter a specific value
for LVSWI and LCWI calculations.
HB
— Enter a specific value for eDO
2
I monitoring. This
value is retained for 12 hours and may be used in other
calculations (for example, pulmonary calculations).
FAST LOOK
Select one of the following options for viewing and printing
parameter data:
NORMAL RANGES
— Display data with normal
parameter ranges. A plus (+) or minus (-) indicates
values outside the normal range.
UNITS OF MEASURE
— Displays data with units of
measurement.
SECONDARY
PARAMETERS
Select three parameters from the available parameters to
display in the parameter window. The default selections are
CO, SVR and TFC.
LIMITS
Select the high and low parameter alarm limits. An alarm
sounds when a value occurs outside of these limits.
TRENDS
Display the
GRAPHIC TRENDS
information window. All ICG
parameters display with an “i” preceding the parameter label
(for example, iHR, iMAP, iCO2, etc.).
HELP
Display the
HELP
information window for help with patient
preparation, proper sensor placement and troubleshooting
issues.
WAVEFORMS
Select one of the following options for waveform display:
ICG
— Reflects beat-to-beat changes in the impedance
consistent with events of the cardiac cycle.
DZ/DT
— Reflects the rate of change at the first
derivative of the waveform.
Содержание Dash 3000
Страница 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Страница 12: ...x Dash 3000 4000 5000 2000966 386D ...
Страница 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Страница 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Страница 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Страница 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Страница 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Страница 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Страница 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Страница 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Страница 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Страница 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Страница 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Страница 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Страница 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Страница 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Страница 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Страница 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Страница 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Страница 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Страница 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Страница 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Страница 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Страница 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Страница 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Страница 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Страница 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Страница 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Страница 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Страница 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Страница 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Страница 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Страница 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Страница 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Страница 269: ......