![GE Corometrics 170 Series Скачать руководство пользователя страница 145](http://html.mh-extra.com/html/ge/corometrics-170-series/corometrics-170-series_service-manual_128350145.webp)
Revision C
170 Series Monitor
7-13
2000947-004
Serviceable Assemblies:
Transducer Plug Replacement Kits
Plug Assembly
Refer to
while replacing a transducer plug. Note that the strain relief
clamp (3) orientation is different between tocotransducers and all other transducers
(ultrasound, FECG). In addition, the Oetiker clamp (11) is used for
Nautilus
Tocotransducers only.
Figure 7-4. Transducer Plug Construction
1. Remove the old plug.
2. Follow cable preparation procedures earlier in this document:
“Ultrasound Transducer Cable Preparation”
3. Insert pinned wires into corresponding positions in the pin housing according to
the wiring table in each of the above listed sections. Push in each pin until it
bottoms out and is retained by the pin spring fingers which prevent it from
backing out of the hole.
NOTE:
If a pin is removed from the sprocket housing, it
must
be discarded
and replaced.
4. Insert remaining pins into unused positions in the pin housing.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......