![GE Corometrics 170 Series Скачать руководство пользователя страница 142](http://html.mh-extra.com/html/ge/corometrics-170-series/corometrics-170-series_service-manual_128350142.webp)
7-10
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Transducer Plug Replacement Kits
FECG Cable Preparation
Refer to
and
while following these steps:
1. Cut #26 AWG green wire (not supplied) to 2 inches long. Strip one end
3
⁄
4
inch.
2. Cut #26 AWG blue wire (not supplied) to 1-
1
⁄
2
inches long and strip each end
1
⁄
8
inch. This will be used as a jumper wire.
3. Route the transducer cable through the gripper. Slide the gripper, followed by
the retaining ring, down the cable until they are out of the way.
4. Score around cable jacket 1 inch. Peel and remove jacket. Trim filler flush to
jacket.
5. Pull back the braided shield and twist into pigtail.
6. Strip black conductive coating back to the outer jacket.
7. Solder the 2-inch prepared green wire to the shield, wrapping the
3
⁄
4
-inch strip onto the shield. Cut the wire flush with other conductors of cable.
8. Install shrink tubing over the solder joint of the shield.
9. Strip all conductors
1
⁄
8
inch. Twist and tin conductors.
CAUTIONS
CABLE IDENTIFICATION—Identify each cable as it is stripped
since colors will be difficult to determine once the sleeving
contracts.
HEAT—Avoid excess heat when tinning ends and soldering
braids since the polyethylene inner insulation melts very easily.
10. Slide shrink tubing over cable.
11. Locate the grounding contact pin (longer pin) in the kit. Solder this pin to one
end of the blue jumper wire.
12. Solder a short pin to the blue wire and the free end of the blue jumper wire.
Both wires share this pin.
13. Solder a short pin to each of the other conductors,
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......