Programming the Calibration Factor for CF-LP Clamp-on Transducers
June 2012
XMT868i
1.
Press
[ESC], [ENTER], [ESC]
. The meter enters the
Keypad Program
.
2.
Press
[X]
until
PROG
appears and press
[ENTER]
.
3.
Press
[X]
until the desired channel appears. Press
[ENTER]
at CH1 or CH2.
4.
Press
[X]
until
PIPE
appears and press
[ENTER]
.
5.
Once in the
PIPE
submenu, the meter interface will guide the user through entering information about the
parameters listed below. Enter or confirm that the information entered is correct and press
[ENTER]
at each step
to walk through the interface:
a.
TRANSDUCER NUMBER
b.
PIPE OD
c.
PATH LENGTH P
d.
AXIAL LENGTH L
e.
TRACKING WINDOWS
f.
FLUID TYPE
g.
FLUID SOUNDSPEED
h.
REYNOLDS CORRECTION
6.
After selecting the option for
REYNOLDS CORRECTION
, the meter interface will be directed to the
CALIBRATION
FACTOR
screen. Enter the calibration factor engraved on the mounting rail. If there is no engraving, enter a
calibration factor of 1.000. Press
[ENTER]
.
7.
Press
[ESC]
until you are out of the
Keypad Program
.
Note:
For a visual representation of the steps above, please refer to the menu map (Figure 29) in the XMT868i
Programming Manual.