Programming the Calibration Factor for CF-LP Clamp-on Transducers
June 2012
PT878
1.
Press
ESC
until you are at the home screen (where pressing
ESC
no longer changes the screen).
2.
Press the
Menu
key.
3.
Using the arrow keys, scroll to
Program
. Press
ENTER
.
4.
Using the arrow keys, scroll to
Correction Factors.
Press
ENTER
.
5.
Using the arrow keys, scroll to the
Calibration Factor
tab. Press
ENTER
.
6.
Using the arrow keys, scroll to the first set of radio buttons to turn the Calibration Factor
On
.
7.
Using the arrow keys, scroll to the next set of radio buttons and choose
Single
K-Factor.
8.
Using the arrow keys, scroll to the
Meter K-Factor
entry. Press
ENTER
.
9.
In the dialog box, enter the
Calibration Factor
engraved on the mounting rail. If there is no engraving, enter
1.000. Press
ENTER
.
10.
Press the
F3 (OK)
button to accept the changes.
Note:
For a visual representation of the steps above, please refer to the menu map (Figure 152) in the PT878 User
Manual.