Neuromuscular transmission
NMT cautions
CAUTION
Always stop the NMT measurement before handling the
stimulating electrodes.
CAUTION
Never subject a patient with an implanted electronic device to
electrical stimulation without consulting a medical specialist
first.
CAUTION
If used in close proximity to shortwave or microwave therapy
equipment, stimulator output may become unstable.
NMT measurement limitations
●
This measurement is not available in the NICU software package.
●
NMT measurement is not indicated for pediatric patients weighing less than 5
kg (11 lbs).
●
Pediatric MechanoSensor is validated for children weighing 5 to 20 kg (11 to 44 lbs).
●
Electrosurgery may cause incorrect measurement results.
●
NMT measurement is not indicated for patients with known abnormal function of
neuromuscular junction.
NMT points to note
●
Start monitoring before the administration of a muscle relaxant drug (but after the
induction of sleep in general anesthesia) to prevent voluntary muscle contraction
and tension from interfering with the reference search.
●
When placing the electrodes, make sure that they do not touch each other.
●
Do not place electrodes on areas with excessive body hair or lesions.
●
If the electrodes are placed incorrectly, wrong nerves are stimulated and this
causes wrong muscle response.
●
When multiple nerves are stimulated, the measured response may be affected
by electrical activity of other muscles.
●
If the stimulation electrodes are placed very close to the palm of the hand, the
muscles are stimulated directly by the stimulation pulses.
●
Do not apply stimulation across or through the head, directly on the eyes, covering
the mouth, on the front of neck (especially over the carotid sinus), or from electrodes
placed on the chest and the upper back or crossing over the heart.
●
If the current is too strong, it may stimulate the muscles too much.
●
Moving or touching the patient during measurement may cause incorrect results.
●
For safe extubation, the TOF% should be higher than 90. Also assess other clinical
signs.
Checking the NMT measurement
1.
Always check the electrode quality.
362
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...