10
Nebulizer
Warning
• Do not use a filter, heat-moisture exchanger or heat-moisture
exchanger filter between the nebulizer and the patient airway.
• Use of a heat-moisture exchanger or nebulizer in the breathing
circuit can substantially increase flow resistance when a
nebulizer is active. Monitor the breathing system filter
frequently for increased resistance and blockage.
• Use of an external pneumatic nebulizer may significantly
impact volume delivery and monitoring, decrease trigger
sensitivity, and cause alarms if external flow is introduced
and Pneumatic Nebulizer Flow Compensation is not used.
Caution
• It is strongly recommended to use an expiratory filter when a
nebulizer is used to help protect the expiratory flow sensor.
Nebulizer Treatment
The ventilator supports the Aerogen
®
Pro In-line Nebulizer
and Aerogen
®
Solo Disposable In-line Nebulizer. The
nebulizer time can be set for specific duration delivery
times or for the volume of medication that will be delivered
to the patient. A general message shows the nebulizer
treatment time remaining.
1. Select Menu.
2. Select Nebulizer.
3. Select Aerogen.
4. Select Time or Continuous.
Note:
Time and approximate volume of nebulized
medication are shown in the table.
5. Select Start.
Note:
To end nebulizer treatment before the set time,
select Stop.
Time
(min)
7
8
11
16
21
26
32
Volume
(ml)
2.5
3.0
4.0
6.0
8.0
10.0
12.0
1
2
3
5
4