Configuration
89
2106778-001
7.16.3 Testing the connectivity
NOTE: Connectivity test works only if the connection is enabled, that is the current state is
enabled/running.
1.
Log in to Webmin.
2.
Select
Configuration
>
Remote Service
.
3.
In the
Sub-Modules for Remote Service
menu, select
Control
.
4.
Below
Test Connectivity
in the
Remote Service Control
window, select
Test
to test the
connectivity to the remote service server.
5.
Verify that
Connection to Enterprise URL
and
Connection to Tunnel URL
both pass the
test.
6.
If one of the connection tests fail, check the failure description and fix the connectivity
issue. Re-test the connectivity.
7.17 Transferring settings from one patient monitor to
other patient monitors
You can transfer platform and/or clinical settings from one patient monitor to another, take
backup copies of the settings to an external device and restore the settings from an external
device.
1.
Complete the platform and/or clinical configuration in one patient monitor.
2.
Save the completed platform and/or clinical configuration settings to an external device
(section
3.
Load the saved platform and/or clinical configuration settings from an external device to
the destination patient monitors (section
4.
Activate the loaded platform and/or clinical configuration settings in the destination
patient monitors (section
5.
Some platform settings can't be transferred from one patient monitor to another. See
section
for details. Configure these unaffected settings
manually in the destination patient monitor following the instructions in sections
.
NOTE: Clinical settings are software package and profile specific. For example, you can transfer
clinical settings from a patient monitor with OR and PACU software only to other patient
monitors with OR and/or PACU software, not to patient monitors with ICU, NICU and ED
software.
7.17.1 Saving settings
NOTE: The patient monitor must not be configured to receive any alarm notifications from
other monitors when the settings are saved.
Before saving the settings to a patient monitor that is connected to the MC or S/5 Network,
make sure that the monitor is not receiving any alarm notifications from other monitors:
1.
Select
Data&Pages > Other Patients > Receive Alarms.
2.
Check that
Change All Notifications
is set to
Off
.
Save the completed platform and/or clinical settings to a service PC or USB flash drive.
1.
Log in to Webmin.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......