CARESCAPE Monitor B850
90
2106778-001
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Save
.
4.
In the
Save Settings
window, select the radio button next to the type of settings you want
to save.
5.
Select
Save
.
6.
According to your Webmin access:
a.
If you are using Webmin on a service PC, you can save the settings file to any mass
storage device connected to the service PC:
In the
File Download
dialog box, select
Save
.
In the
Save as
dialog box, select the destination drive and folder and select
Save.
NOTE: You may change the default filename, but do not change the file extension.
b.
If you are using Webmin locally through the integrated browser, you can save the settings
file to a USB flash drive that is connected to one of the patient monitor’s USB ports:
The
Save As
window will show you the name of the created settings file.
Select
Save
to save the settings file to the USB flash drive.
NOTE: You may change the default filename, but do not change the file extension.
NOTE: Do not disconnect the USB storage device until saving is complete.
7.17.2 Loading settings
Load the saved platform and/or clinical settings from a service PC or USB flash drive to the
patient monitor.
NOTE: The loaded settings will remain in an inactive state in the patient monitor until they are
purposely activated by the user (see section”
”).
1.
Log in to Webmin.
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Load
.
4.
According to your Webmin access:
a.
If you are using Webmin on a service PC, you can load the settings file from any mass
storage device connected to the service PC:
In the
Load Settings
window, enter the file name or select
Browse
to select a file
from the
Choose File to Upload
dialog box.
Select
Upload
to load the settings.
b.
If you are using Webmin locally through the integrated browser, you can load the settings
file from a USB flash drive that is connected to one of the patient monitor’s USB ports:
The
Load Settings
window will show you the available settings files. Select the
settings file to be loaded.
Select
Load Settings
to load the selected settings file from the USB flash drive.
NOTE: Do not disconnect the USB storage device until loading is complete.
7.17.3 Activating settings
NOTE: Settings activation takes place only when the patient monitor is in a case reset / patient
discharged state. If you are going to activate settings immediately, verify that the patient
monitor is in a case reset / patient discharged state. You can alternatively initiate the setting
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......