Chapter 5 - Environmental Conditions
Page 99
n
ta
l
Condit
ion
s
Section 2.0
Cooling Requirement
Use
to assist in cooling requirements planning. Over half the cooling used by your
scanner is required for gantry operation. Locate a wall air-conditioning vent at floor level beside and
behind gantry to meet both gantry cooling needs and provide patient comfort. Do not locate any
cooling vents directly above the gantry. Air returns above the gantry are recommended.
, for component air flow
requirements.
Section 3.0
Altitude
System operating altitude is from -492 ft. to 9842.5 ft. (-150 meters to 3000 meters).
SYSTEM COMPONENT
BTU/HR
WATT
1. Gantry recommended (See NOTE 1)
18,766
5,500
2. HPower Table / GT1700 Table
682/1023
200/300
3. Power distribution unit
5,118
1,500
Recommended Scan Room Subtotal (see notes):
24,566/24,907
7,200/7,300
4. Operator's console w/ 1 IG, 2 monitors & SCSI Tower
8,189
2,400
MOD Drive (Option)
682
200
IG Node, each (Option)
1,024
300
Remote Monitor (Option)
222
65
DASM (Option)
290
85
Recommended Control Room Subtotal
(w/o option)
:
8,189
2,400
System Total (Recommended)
(See NOTE 1)
32,755/33,096
9,600/9,700
Option
: Advantage Windows
256
75
Option
: UPS
5,200
1,200
ROOM TOTAL (SEE NOTE 2)
NOTE 1:
With 75 scan rotations per patient:
Recommended BTU/hr. provides for up to six patients per hour. It is also needed
during calibration of the system.
NOTE 2:
Cooling requirements do not include cooling for room lighting, personnel or non-CT
equipment.
Table 5-1 Cooling Requirements (Worksheet)