![GE Baker Hughes oxy.IQ Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/ge/baker-hughes-oxy-iq/baker-hughes-oxy-iq_user-manual_18634071.webp)
Appendix D. Certifications
oxy.IQ User’s Manual
61
D.4 FM Certificate of Compliance
To verify the availability of the Approved product, please refer to
www.approvalguide.com
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail:
www.fmapprovals.com
F 347 (Mar 16)
Page 1 of 3
CERTIFICATE OF CONFORMITY
1.
HAZARDOUS (CLASSIFIED) LOCATION ELECTRICAL EQUIPMENT PER US REQUIREMENTS
2.
Certificate No:
FM17US0190X
3.
Equipment:
(Type Reference and Name)
Model oxy.IQ Oxygen Transmitter
4.
Name of Listing Company:
GE Infrastructure Sensing
5.
Address of Listing Company:
1100 Technology Park Drive
Billerica, MA 01821
USA
6.
The examination and test results are recorded in confidential report number:
3047174 dated 3
rd
March 2015
7.
FM Approvals LLC, certifies that the equipment described has been found to comply with the following Approval
standards and other documents:
FM Class 3600:2011, FM Class 3610:2010, FM Class 3810:2005,
ANSI/ISA 60079-0:2013, ANSI/UL 60079-11:2013, ANSI/ISA 61010-1:2004
8.
If the sign ‘X’ is placed after the certificate number, it indicates that the equipment is subject to specific
conditions of use specified in the schedule to this certificate.
9.
This certificate relates to the design, examination and testing of the products specified herein. The FM
Approvals surveillance audit program has further determined that the manufacturing processes and quality
control procedures in place are satisfactory to manufacture the product as examined, tested and Approved.
10. Equipment Ratings:
Intrinsically Safe (Entity) for use in Class I, Division 1, Groups A, B, C and D; Temperature Class T4
Tamb = -20˚C to +60˚C in accordance with Control Drawing No.752-347; Intrinsically safe (Entity) for use in
Class I, Zone 0, AEx ia IIC T4 Tamb = -20˚C to +60˚C; in accordance with Control Drawing No.
752-347; Nonincendive for us in Class I, Division 2, Groups A, B, C and D; Temperature Class T4
Tamb = -20˚C to +60˚C; Hazardous (Classified) Location.
Certificate issued by:
13 June 2017
J. E. Marquedant
VP, Manager, Electrical Systems
Date
Содержание Baker Hughes oxy.IQ
Страница 1: ...Oxygen 910 296 Rev E July 2018 bhge com oxy IQ Panametrics Oxygen Analyzer User s Manual ...
Страница 2: ......
Страница 4: ...ii oxy IQ User s Manual no content intended for this page ...
Страница 30: ...Chapter 3 Initial Setup Operation 20 oxy IQ User s Manual no content intended for this page ...
Страница 38: ...Chapter 4 User Programming 28 oxy IQ User s Manual no content intended for this page ...
Страница 46: ...Chapter 6 Specifications 36 oxy IQ User s Manual no content intended for this page ...
Страница 58: ...Appendix B Menu Maps 48 oxy IQ User s Manual no content intended for this page ...
Страница 60: ...Appendix C Order String 50 oxy IQ User s Manual no content intended for this page ...
Страница 66: ...Appendix D Certifications 56 oxy IQ User s Manual D 2 ATEX IECEx MAM Ex Certificate ...
Страница 67: ...Appendix D Certifications oxy IQ User s Manual 57 D 2 ATEX IECEx MAM Ex Certificate cont ...
Страница 74: ...Appendix D Certifications 64 oxy IQ User s Manual D 5 IECEx Certificate of Conformity ...
Страница 75: ...Appendix D Certifications oxy IQ User s Manual 65 D 5 IECEx Certificate of Conformity cont ...
Страница 76: ...Appendix D Certifications 66 oxy IQ User s Manual D 5 IECEx Certificate of Conformity cont ...
Страница 77: ...Appendix D Certifications oxy IQ User s Manual 67 D 5 IECEx Certificate of Conformity cont ...
Страница 87: ...Addendum A oxy IQ Safety Manual 77 oxy IQ User s Manual no content intended for this page ...
Страница 89: ... no content intended for this page ...
Страница 90: ......