Chapter 3. Initial Setup & Operation
oxy.IQ User’s Manual
19
3.3.4 Span Gas Calibration
Before beginning the span gas calibration, make sure the oxy.IQ is
indicating an O
2
level less than the span gas value, to ensure an
accurate calibration. Then, start the flow of the span gas to the
sensor. For accurate calibration, the span gas should have an oxygen
content of 70-90% of the range being calibrated.
To perform the span calibration, complete the following steps:
1.
Use the equation below to calculate the expected mA output that
corresponds to the known oxygen content of the span gas:
For example, if the span gas contains 80 ppm oxygen and the
0-100 ppm range is being calibrated, the analog output should equal
4 + 16 x (80/100) = 16.8 mA.
2.
If you have not done so already, start the flow of span gas to the
sensor and allow both the 4-20 mA output reading and the
display reading to stabilize.
3.
After the reading has stabilized, press the
Enter
key to enter
the
Main Menu
.
4.
Press the
Enter
key to enter the
Calibration
menu.
5.
Press the
key and then press the
Enter
key to select the
Span Gas
menu option.
6.
Press the
and
keys until the measurement agrees with the
span calibration gas value.
7.
Confirm that the reading on the display has stabilized, and press
the
Enter
key to save the calibration. Then, press the
Cancel
key twice to return to measurement mode.
4.0 16.0
Span Gas ppm
Full Range ppm
--------------------------------------
+
mA Output
=
Содержание Baker Hughes oxy.IQ
Страница 1: ...Oxygen 910 296 Rev E July 2018 bhge com oxy IQ Panametrics Oxygen Analyzer User s Manual ...
Страница 2: ......
Страница 4: ...ii oxy IQ User s Manual no content intended for this page ...
Страница 30: ...Chapter 3 Initial Setup Operation 20 oxy IQ User s Manual no content intended for this page ...
Страница 38: ...Chapter 4 User Programming 28 oxy IQ User s Manual no content intended for this page ...
Страница 46: ...Chapter 6 Specifications 36 oxy IQ User s Manual no content intended for this page ...
Страница 58: ...Appendix B Menu Maps 48 oxy IQ User s Manual no content intended for this page ...
Страница 60: ...Appendix C Order String 50 oxy IQ User s Manual no content intended for this page ...
Страница 66: ...Appendix D Certifications 56 oxy IQ User s Manual D 2 ATEX IECEx MAM Ex Certificate ...
Страница 67: ...Appendix D Certifications oxy IQ User s Manual 57 D 2 ATEX IECEx MAM Ex Certificate cont ...
Страница 74: ...Appendix D Certifications 64 oxy IQ User s Manual D 5 IECEx Certificate of Conformity ...
Страница 75: ...Appendix D Certifications oxy IQ User s Manual 65 D 5 IECEx Certificate of Conformity cont ...
Страница 76: ...Appendix D Certifications 66 oxy IQ User s Manual D 5 IECEx Certificate of Conformity cont ...
Страница 77: ...Appendix D Certifications oxy IQ User s Manual 67 D 5 IECEx Certificate of Conformity cont ...
Страница 87: ...Addendum A oxy IQ Safety Manual 77 oxy IQ User s Manual no content intended for this page ...
Страница 89: ... no content intended for this page ...
Страница 90: ......